Hoffrichter Trend 300 User Manual

Manual is about: Respiration Therapy Device

Summary of Trend 300

  • Page 1

    User`s manual trend 300 tren d 300

  • Page 2

    Serial number each h o ffrichter device is provided with a serial num- ber which guarantees the product can be traced back. The serial number is on the type identification plate to be found on the reverse of the device. In case of any query or complaint, please always state the serial number. Confor...

  • Page 3

    3 respiration therapy device trend 300 legend of symbols 04 safety instructions 05 intended use 08 technology of trend 300 09 hints to be observed before starting 10 description of the device 11 proper use of the device 14 use of oxygen 16 cleaning instruction, changing of filter 17 alarm functions ...

  • Page 4

    Information on the type identification plate: pay attention to user´s manual applied part type bf protection class ii (insulation - enclosed) ce - conformity statement do not dispose of the device with normal domestic waste. For information on the proper disposal of the device, pleasa contact your r...

  • Page 5

    5 respiration therapy device trend 300 safety instructions before first use of your cpap-device, please read this manual thoroughly. Keep the manual in a safe place, so that you will be able to consult it whenever necessary. Power supply trend 300 has a switched mode power pack with a wide power sup...

  • Page 6

    6 respiration therapy device trend 300 please observe the following points in order to avoid dangerous situations which may occur: 1. Use the device exclusively for your own respiration therapy and only according to what has been directed by your physician. 2. Only use accessory recommended and offe...

  • Page 7

    7 respiration therapy device trend 300 safety instructions in order to maintain electric safety please observe the following points: ■ do not switch on the device if casing or cable are damaged. ■ do not open the casing of trend 300 . In case of failures always contact your authorized dealer. ■ the ...

  • Page 8

    Intended use the trend 300 is for treatment of nighttime sleep apnea in patient weighing at least 30 kg. This device is not intended for use with ventilator- dependent patients. The trend 300 provides positive airway pressure in two pressure levels. This device is designed to be used with nasal brea...

  • Page 9

    Trend 300 is a respiration bilevel therapy unit for treatment of nighttime sleep apnea. An air humidifier can be obtained as an accessory. For the pressure generation trend 300 has an electronically controlled blower. In order to keep the patient's burdening as low as possible the turbine has been e...

  • Page 10

    Before operating the device, please pay attention to the following: 1. Never switch on the device if it has recently been in a very cold environment. Wait approximately one hour for temperature compensation. 2. Check the filter cassette at regular intervals and change it regular (see pages 17, 26). ...

  • Page 11

    11 respiration therapy device trend 300 description of the device measuring tube connection air outlet display program keys control key input key on/off connectors for aquatrend (accessory) release button (top) for accessories hoffrichter trend 300 front view.

  • Page 12

    12 respiration therapy device trend 300 description of the device fernbedienung remote control dc in 24 v trend 300 ac power plug ac power switch type identification plate dc-input rs232-interface filter cassette (top) air input back of the device.

  • Page 13

    13 respiration therapy device trend 300 description of the device breathing-mask (additional accessory) therapy tube (1,80 m) carrier bag measuring tube stoppers filter cassette cable for motor car exhalation port (additional accessory) power supply cable accessory and spare parts.

  • Page 14

    1. As soon as the device is placed properly please connect it with the power supply cable to an earthing contact socket. 2. Switch on the ac power switch on the back of the device. The device welcomes you with a little text in the display. Afterwards the display switches over to indication of curren...

  • Page 15

    7.Place the therapy tube in such a way that there is no tension onto the mask when you lay down. 8.Breathe deeply and calmly and only through the nose. If the softstart-function is activated the device will after leakage test automatically reduce the therapeutic pressure first and then gradually inc...

  • Page 16

    16 respiration therapy device trend 300 use of oxygen please observe the following advices if you must additionally inhale oxygen during the respiration-therapy: if you want to lead the oxygen directly into the breathing-mask please use a non-buckling tube made from material approved for medical app...

  • Page 17

    17 respiration therapy device trend 300 cleaning instructions, changing of filter daily cleaning of the breathing-mask ■ disconnect the mask and exhalation port from the therapy tube. ■ clean the breathing mask in accordance with the instructions of the manufacturer of the mask. Weekly cleaning of t...

  • Page 18

    18 respiration therapy device trend 300 alarm functions of trend 300 power failure probably you won't notice a power failure at night-time. As a consequence you would permanently inhale the exhausted air from the therapy tube. Therefore trend 300 has an alarm mechanism responding if a power fai- lur...

  • Page 19

    19 respiration therapy device trend 300 for programming the functions of trend 300 three keys are available: = input key = program key = program key by holding the input key pressed, you reach the input mode after 1 second, and then you see the menu of trend 300 in the display. The following 2 diffe...

  • Page 20

    20 respiration therapy device trend 300 functions of trend 300 wake time.......... 06:00 exit menu! Press input key to leave the programming mode. During standard operation the display indicates date and time: s a 2 8 . F e b 1 2 : 4 2 the alarm clock can be activated or deactivated by the keys and ...

  • Page 21

    21 respiration therapy device trend 300 functions of trend 300 date, time date and time of the internal clock can be adjusted. The internal clock has a built-in battery with reserve power for 8 years without mains connection. You can adjust the weekdays (mo thru su), day (1 till 31), month (1 till 1...

  • Page 22

    22 respiration therapy device trend 300 automatic operation in case of automatic operation you have only to put on the mask. As soon as you breathe the turbine of trend 300 will be switched on. If you take off the mask the turbine will be switched off automatically after a delay time of 5 seconds. D...

  • Page 23

    23 respiration therapy device trend 300 trouble shooting pain in nose, nasal inflammation stop therapy and contact cavity or ears your doctor unpleasant feeling of too problem with contact your doctor high pressure prescribed high pressure you are not yet used to the relax and use softstart function...

  • Page 24

    24 respiration therapy device trend 300 trouble shooting irriteted or dry eyes leakage between mask check correct position and skin of mask new mask if defective wet nose reaction to the air flow contact your doctor about use of a humidifier head cold, sinus drainage or normal cold or sinus infectio...

  • Page 25

    25 respiration therapy device trend 300 trouble shooting air too warm dirty filter exchange filter air input is blocked make sure air input is clear heating too close to check position of trend 300 trend 300 no air flow device is defective ask your authorized dealer water in measuring tube make sure...

  • Page 26

    26 respiration therapy device trend 300 display indications in the following table the most important display indications are shown. The sense off all other indications can be taken from context. Mask test. . .55s you still have 55 s for fitting the mask correctly. After that time the softstart func...

  • Page 27

    27 respiration therapy device trend 300 inspection and check although all parts of trend 300 are proved to have a long life the device should be examined from time to time. A device check is necessary if the message "service!" appears in the display. In this case we urgently recommend you to have th...

  • Page 28

    Specification hoffrichter order-no. Filter cassette 00007870 ac power supply cable 31100001 therapy tube 00007116 stoppers 44000012 (for closing measuring tube during tube cleaning) carrier bag 00007860 user´s manual 00007877 cable for motor car 00009212 plug-in-humidifier aquatrend iii-h0 00007992 ...

  • Page 29

    29 respiration therapy device trend 300 technical data dimensions (b x t x h) 230 x 225 x 110 mm mass ca. 2,5 kg power supply 100 v~ … 240 v~, 50 ... 60 hz 24 v dc pressure range ipap, epap: 3 … 18 hpa (mbar) in steps of 0,5 hpa power consumption without humidifier 22 w (average value) power consump...

  • Page 30

    30 respiration therapy device trend 300 trend 300 - patient pass name first name street postal zip code/place telephon birth. On health insurance company hospital-stamp family doctor-stamp customer service-stamp serial-no. Software-version type of mask mask size therapy process date weight high bloo...

  • Page 31

    31 respiration therapy device trend 300 notes.

  • Page 32

    H o ffrichter type names: trend 110 monolevel typ 5cp 102 trend 210 monolevel typ 5cp 210 trend 300 bilevel typ 5cp 301 trend 400 bilevel st20 typ 5cp 401 trend 500 bilevel st33 typ 5cp 501 autotrend auto cpap typ 5cp 253 aquatrend iii-h0 humidifier typ 5cp 931 aquatrend iii-h1 humidifier typ 5cp 94...