HoMedics BPA-3000-EU Instruction manual

Manual is about: Automatic Arm Blood Pressure Monitor

Summary of BPA-3000-EU

  • Page 1

    1 instruction manual bpa-3000-eu deluxe automatic arm blood pressure monitor ib-bpa3000-0912-01:layout 1 09/01/2013 17:43 page 1.

  • Page 2

    2 g b homedics blood pressure monitor before use before the blood pressure monitor is taken into use, a number of basic steps should be observed. Please read this manual thoroughly before the blood pressure monitor is used for the first time and keep the manual for future use. The blood pressure mon...

  • Page 3

    3 g b what are systolic pressure and diastolic pressure? When ventricles contract and pump blood out of the heart, blood pressure reaches its maximum value, the highest pressure in the cycle is known as systolic pressure. When the heart relaxes between heartbeats, the lowest blood pressure is diasto...

  • Page 4

    4 g b caution: consult accompanying documents type b applied part ce mark: conforms to essential requirements of the medical device directive 93/42/eec. Disposal: do not dispose of this product as unsorted municipal waste. Collection of such waste separately for special treatment is necessary. Manuf...

  • Page 5

    5 g b lcd display signal symbol description explanation systolic blood pressure high pressure result diastolic blood pressure low pressure result pulse pulse/minute deflating cuff is deflating time (hour:minute) current time memory if “m” shows, the displayed measurement values is from the memory. M...

  • Page 6

    6 g b mem mem list inserting and changing batteries monitor components cuff air hose air connector plug lcd display user select switch dc power socket mem/up touch key start/stop button set/down touch key battery compartment 1. Blood pressure monitor 2. Cuff 1. Slide off the battery cover. 2. Instal...

  • Page 7

    7 g b setting date, time and measurement unit it is important to set the clock before using your blood pressure monitor, so that a time stamp can be assigned to each record that is stored in the memory. (year: 2000—2050, time: 24 h) 67$57 6723 67$57 6723 67$57 6723 1. When the unit if off, hold “set...

  • Page 8

    8 g b fitting the cuff correct cuff size is critical to accurate measurement. This monitor comes with a standard size 22-32cm arm cuff. A large size cuff 32-42cm is available from the homedics website www.Homedics.Co.Uk 1. If the cuff is not assembled at first use, put the end of the cuff which is furthe...

  • Page 9

    9 g b worth knowing about blood pressure measurement • blood pressure measurements should be performed in quiet and comfortable surroundings as measurements can be affected by high or low temperatures. The blood pressure should be measured at normal body temperatures. • do not move or speak during me...

  • Page 10

    10 g b up down 2. Press the “mem” or “set” to get the record you want. The order of the record, date, time will display alternately. The current record and total number of records. The month and day the corresponding time will show and alternate with date. Caution the most recent record (1) is shown...

  • Page 11

    11 g b error search this section includes a list of error messages and frequently asked questions for problems you may encounter with your blood pressure monitor. If irregularities occur during use, please check the following items. If the device still does not function, please contact the store whe...

  • Page 12

    12 g b power supply battery powered mode: 6vdc 4*aaa alkaline batteries ac adaptor powered mode: 100-240v~, 50-60hz, 400ma (you can purchase the ac adaptor pp-adpr2k3k-eu from the homedics website). Display mode digital lcd v.A.128*50mm measurement mode oscillographic testing mode measurement range ...

  • Page 13

    Battery directive this symbol indicates that batteries must not be disposed of in the domestic waste as they contain substances which can be damaging to the environment and health. Please dispose of batteries in designated collection points. Weee explanation this marking indicates that this product ...

  • Page 14

    14 g b guidance and manufacture’s declaration – electromagnetic emission the bpa-3000-eu is intended for use in the electromagnetic environment specified below. The customer of the user of the bpa-3000-eu should assure that it is used in such and environment. Emission test compliance electromagnetic...

  • Page 15

    15 g b guidance and manufacture’s declaration – electromagnetic immunity the bpa-3000-eu is intended for use in the electromagnetic environment specified below. The customer of the user of the bpa-3000-eu should assure that it is used in such an environment. Immunity test iec 60601 test level compli...

  • Page 16

    16 g b guidance and manufacture’s declaration – electromagnetic immunity the bpa-3000-eu is intended for use in the electromagnetic environment specified below. The customer or the user should assure that it is used in such an environment. Immunity test iec 60601 test level compliance level electrom...

  • Page 17

    17 g b 16 recommended separation distances between portable and mobile rf communications equipment and the ele007839v1 fitness equipment. The bpa-3000-eu is intended for use in an electromagnetic environment in which radiated rf disturbances are controlled. The customer or the user of the bpa-3000-e...

  • Page 18

    Ib-bpa-3000eu-0912-01 register your product today at: http://www.Homedicsgroup.Com/register 0120 homedics group ltd, homedics house somerhill business park, five oak green road tonbridge, kent tn11 0gp, uk ib-bpa3000-0912-01:layout 1 09/01/2013 17:47 page 244.