IBP MEDIPRO 100f Instruction Manual

Manual is about: fully automatic upper arm blood pressure monitor

Summary of MEDIPRO 100f

  • Page 1

    Closing date: august 2003 © by ibp innovative business promotion gmbh, germany m edi p ro 100f gebrauchsanweisung instruction manual mode d’emploi gebruiksaanwijzing fully automatic upper arm blood pressure monitor medipro 100f englisch 05.08.2003 9:11 uhr seite 1.

  • Page 2

    Medipro 100f englisch 05.08.2003 9:11 uhr seite 2.

  • Page 3

    1 table of contents 1 introduction ....................................................................................................................... 2 2 preliminary remarks ........................................................................................................... 3 3 about blo...

  • Page 4

    2 dear customer, we are glad that you have decided for this blood pressure monitor. The product you have bought is state of the art, of highest quality and has many advanced features. The device measures the upper arm blood pressure both rapidly and extremely accurately, and avoids the need to const...

  • Page 5

    3 this blood pressure monitor complies with the european regulations and bears the ce mark “ce 0366“. This blood pressure monitor bears the ce (conformity) label for notinvasive blood pressure monitors to show that it conforms with the eu council directive (en 1060-1 / en 1060-3) for notinvasive blo...

  • Page 6

    4 more than half the number of people suffering from high blood pressure are unaware of it. The most common cause of death in today's society is from diseases that are directly related to the problem of high blood pressure. This blood pressure monitor allows you to measure your blood pressure at hom...

  • Page 7

    5 about blood pressure 3 remember that only the physician can interpret your blood pressure values. Never venture self- diagnosis or self-treatment. We recommend you to discuss your blood pressure values with your physician. He can advise you on any measures that may be good to take. The blood press...

  • Page 8

    6 3.2 what is “hypertension“ (high blood pressure) and why is it harmful? If parts of the blood transport system are constricted, diseased or in any way not functioning normally, the blood pressure can be higher than normal. High blood pressure can damage vital organs, such as the brain or the heart...

  • Page 9

    7 3.4 how to obtain accurate readings the blood pressure is constantly changing throughout a day. The changes are caused by psychic and physical factors, such as anxiety, exercise, smoking, caffeine and stress. Because it is changing, the blood pressure should always be measured at the same time of ...

  • Page 10

    8 4 the unit / accessories 60 auto-memory identification of the unit accessories included are instruction manual with warranty, batteries and keeping box (without ill.) 4 “aa” size 1,5 v alkaline batteries arm cuff with connector lcd display on/off/start - key memory key air tube medipro 100f englis...

  • Page 11

    9 the display / icons 5 display icons systolic pressure diastolic pressure pulse rate no. Of memory sequence memory icon the memory icon is displayed during save or memory display. Battery icon when this icon appears the batteries should be replaced. Pulse icon the pulse icon appears when the pulse ...

  • Page 12

    Battery installation / battery replacement the enclosed batteries must be fitted before use. The battery cover is on the bottom side of the unit. Push it open and remove it. Insert the 4 supplied “aa“ size, 1.5 v alkaline batteries as shown here. Ensure correct polarity according to the indication i...

  • Page 13

    11 plug in the cuff connecting tube into the left side of the unit. The monitor is used as a desktop unit. Unwrap the arm cuff, leaving the “index“ end of the cuff through the d-ring of the cuff, so that the cuff forms a tunnel shape (a). Put your left arm through the cuff loop. Position the red mar...

  • Page 14

    12 7 making a blood pressure measurement have yourself on a seat in a comfortable position, breathe deeply several times and relax. To get a correct picture of the blood pressure changes during the day, make the measurements at home daily at the same time. To make sure that you obtain a correct meas...

  • Page 15

    13 making a blood pressure measurement 7 place the cuff around your arm as described on pages 11 and 12. Start the monitor by pressing the on /off/start key. When the on/off/start key is pressed, all digits will light up (see fig. 1). This is a check of the display functions which is com- plete afte...

  • Page 16: ...

    14 7 making a blood pressure measurement the unit is ready for use. The unit will automatically start to inflate slowly to measure your blood pressure. During inflation, a changing measure of the cuff pressure will be shown on the display. If the cuff pressure is low, the unit interrupts the measuri...

  • Page 17

    15 making a blood pressure measurement 7 the cuff will deflate when the unit completes the measuring. The lcd then shows the systolic and diastolic pressures, and the pulse rate. The memory symbol blinks briefly. Then it will automatically display and save one’s systolic, diastolic, and pulse in the...

  • Page 18

    16 8 the memory function the data in the memory will not be lost even when the blood pressure monitor is switched off. The data in the memory is erased when the batteries are replaced. Up to 60 measurements can be saved in the memory automatically. The values (blood pressure and pulse rate) can be s...

  • Page 19

    An error is displayed 9 error display possible reason remedy the cuff position is not correct. Put on the cuff correctly. The unit is displaying very do not move while the high or very low values. Measuring continues. You have talked or moved don’t speak. During measurement. The batteries are empty....

  • Page 20

    18 10 other important notes the unit contains high-precision assemblies. Therefore avoid extreme temperatures, humidity, direct sunlight. Do not give strong shock. Do not try to open the blood pressure monitor. This will destroy the calibrated factory settings. Protect from ingress of dust. Do not b...

  • Page 21

    19 specifications 11 model : ibp medipro 100f measuring principle : oscillometric method measuring method : real fuzzy technologie indicator : lcd measuring range : 20 - 280 mmhg (blood pressure) : 40 - 180 pulse/min. (pulse rate) measuring resolution : ± 3 mm hg (blood pressure) : ± 5 % (pulse rate...

  • Page 22

    Guarantee your blood pressure monitor was checked to ensure its faultless quality before despatch. Please send the appliance for checking or repair togehter with the purchase document or receipt and the description of the faults and/or of the desired check to the adress below. From the date of purch...