Icare ic100 Quick Manual - page 3
Green probe base light indication
ONE SHORT BEEP
Indicates a correct single measurement.
DOUBLE BEEP
Indicates an incorrect single measurement. (See error messages from full manual).
LONG BEEP
Indicates that six succesful measurements are done and IOP result is available
at the display.
ENGLISH
Icare ic100 tonometer Quick Guide
Step 2.
Keeptheprobehorizontalandpointingperpendicularlytothecenterofthecornea.Thedistance
fromthetipoftheprobetothepatient´scorneashouldbe4-8mm(5/32-5/16inch).Theprobe
baselightmustbegreen.
Step 3.
Press the Measure button to measure IOP.
Probewillcontacttheeyeduringmeasurements:Pressandholdthebuttontocollect6measurements
inarowORPressthebuttonbriefly6timestocollect1measurementatthetime.
DISPLAY DURING MEASUREMENTS
Before
Afterthesecond
measurement
Afterthesixth
measurement
Afterthesixth
measurementbutwith
toomuchvariation
Afterthesixth
measurementbutwith
acceptablevariation
REPEAT