7. USING THE TONOMETER
7.1 CHOOSING THE MEASUREMENT MODE
The tonometer can operate in two modes:
Series mode
The series mode is especially
useful in self-tonometry. In
the series mode, keeping
the button pressed down
(see figure left) initiates the
measurement function, and
the tonometer takes six rapid
measurements one after the
other to obtain the final IOP
reading.
Single mode
You can use the single mode to
take individual measurements
one at a time. The single mode
is especially useful for those
patients who tend to blink
heavily. Here you press the
measurement button briefly
(1 second) for each of the six
measurements to obtain the
final IOP reading (see figure
left).
Series mode, press the measurement button constantly.
Single mode, press the measurement button briefly
(1 second) for each of the six measurements.
Icare HOME tonometer Instruction manual for health care professionals
Icare HOME tonometer Instruction manual for health care professionals
English
English
LOAD
MEASURE
REPEAT
DONE
SERVICE
BATTERY
LOA
EASU
LO
EAS
1. Unwrap the probe.
2. Remove the lid of the
probe container as
shown in the figure
above. Point the
tonometer upward.
3. Drop the probe into
the probe base (1)
by turning the probe
container upside down.
4. Press the measurement
button (13) briefly (1
second) to activate the
probe.
5. The probe moves rapidly
back and forth.
6. See that the Measure
light (15) flashes. If so,
the probe is loaded
correctly and ready for
measurement.
To load the probe:
1X