Icare TA01 Quick Manual - Quick Guide
QUICK GUIDE
MEASUREMENT
Ask the patient to relax and look
1.
straight
ahead at a specific point
to keep the eye
steady during the measurement.
Bring the tonometer near the patient’s eye.
2.
Check that the
central groove is in a horizon-
tal position
. It’s not necessary to look at the
display during the measurement since the
device gives sound signals about errors etc.
Keep the distance from the tip of the probe
3.
to the patient’s cornea at
4 to 8 millimeters
(1/6-1/3 inch)
. If necessary, adjust the
forehead support.
Press the measuring button lightly. The tip of
4.
the
probe should make contact with the
central cornea
.
Six measurements are made. Once the six
5.
correct measurements are done, the IOP will
be shown on display.
Please refer to the manual for additional in-
6.
formation.
ONE SHORT BEEP
–
Indicates a correct single measurement.
DOUBLE BEEP
––
Indicates an incorrect measurement.
(See error codes from manual).
LONG BEEP
Indicates that six succesfull measure-
ments are done and IOP result is available
at the display.
Press either selector button until the dis-
play shows ‘End‘. Press the measuring but-
ton for two seconds - the display will show
‘byE’ and the tonometer will switch off.
If the tonometer is not used for two
minutes, it will automatically switch off.