J-KEM Scientific Gemini User Manual - page 31
31
15. Press
the
▼
key until the word “LEVL” appears in the display.
16.
First hold in the ‘*’ key, then while holding in the ‘*’ key press the
▲
key until “2” appears in the display.
Let go of all the keys.
17. Press
the
▲
key until “SP2.A” appears in the display. Next, hold in the ‘*’ key, then while holding in the
‘*’ key press the
▲
key until the word “nonE” appears in the display. Let go of all the keys.
18. Press
the
▲
key until “diSP” appears in the display. Next, hold in the ‘*’ key, then while holding in the ‘*’
key press the
▼
or
▲
key until the value “0.1o” appears in the display. Let go of all the keys.
19. Press
the
▲
key until “Lo.SC” appears in the display. Next, hold in the ‘*’ key, then while holding in the
‘*’ key hold in the
▼
key until the number in the display stops changing (this will be “0” or “-50” or “-199.9”
depending on thermocouple type). Let go of all the keys.
20. Press
the
▼
key until the word “LEVL” appears in the display.
21.
First hold in the ‘*’ key, then while holding in the ‘*’ key press the
▲
key until “3” appears in the display.
Let go of all the keys.
22. Press
the
▲
key until “rEU.d” appears in the display. Next, hold in the ‘*’ key, then while holding in the ‘*’
key press the
▲
key until the value “1r.2d” appears in the display. Let go of all the keys.
23. Press
the
▲
key until “ZEro” appears in the display. Next, hold in the ‘*’ key, then while holding in the ‘*’
key press the
▼
or
▲
key
until the correct value (see below) appears in the display. Let go of all the keys.
If the TOP meter is being reset, enter the value: ___________. If the BOTTOM meter is being reset, enter
the value: ___________.
24. Press
the
▼
key until the word “LEVL” appears in the display.
If your controller does not have a USB port on the back, skip to step 29.
25.
First hold in the ‘*’ key, then while holding in the ‘*’ key press the
▼
key until “C” appears in the display.
Let go of all the keys.
26. Press
the
▲
key and “Addr” will appear in the display. Next, hold in the ‘*’ key, then while holding in the
‘*’ key press the
▼
or
▲
key until the value “1” appears in the display. Let go of all the keys.
27. Press
the
▲
key and “bAud” will appear in the display. Next, hold in the ‘*’ key, then while holding in the
‘*’ key press the
▼
or
▲
key until the value “9600” appears in the display. Let go of all the keys.
28. Press
the
▲
key and “dAtA” will appear in the display. Next, hold in the ‘*’ key, then while holding in the
‘*’ key press the
▼
or
▲
key until the value “18n1” appears in the display. Let go of all the keys.
29.
Press and hold in both the
▼
and
▲
keys until the temperature appears in the display, then release both keys.
30.
If the alarm is sounding, reset it by pressing the Output Power Level button for the channel being reset.