    Www.Cangiaphat.Com hotline: 0909 899 833 1 contents 1. Introduction ………………………………….…………………………………………. 1 2. Precautions …………………………………….……………………………………….. 1 3. Before using the product 3-1 unpacking and checking………………………………………………………………….. 3-2 installing components ………………………………………..…………………………… 3-3 levelin...

  • Page 3

    Www.Cangiaphat.Com 1. Introduction this manual contains installation and operation instructions for the jwl series weighing scale. Please read the manual completely before installation and operation. 2. Precautions ◎ place the scale on a flat and stable surface( refer to section 3-3). ◎ verify that ...

  • Page 4

    Www.Cangiaphat.Com specifications & general features specifications ) general features ·easy to clean smooth abs plastic housing with included in-use cover and stainless steel weighing pan ·large bright backlit 6 digit lcd with prominent 31mm high digits ·rs-232 serial communication interface ·overl...

  • Page 5

    Low battery indication tare or preset tare indication center of zero indication, the zeroing range is auxiliary display (parameter, accumulated number of weighments) tare symbol “ ” points at “tare” preset tare. Symbol “ ” points at “net” net weight--gross weight minus tare or preset are actions are...

  • Page 6

    Www.Cangiaphat.Com hotline: 0909 899 833 5 print/hi key * short press sends data a rs232 port; long press lower limit initials lower limit setting * shift key (shift rightwards) g/n key * displays gross and net weight by turns * stores current condition mr key * memory recall * switches on /off chec...

  • Page 7

    Vetek weighing ab low battery warning when appears in the upper left corner of the weight window, the battery requires recharging. The charge lamp turns green from red when the recharging is completed (which takes about 8 hours).Disconnect the scale from power supply when it is fully charged. Weighi...

  • Page 8

    3) press key g/n to save and return reverting to preset tare 2) put the load on the container, the scale will automatically deduct the value of the container from the total value. Note: press key g/n to display gross and net weight by turns. Checkweighing use this mode to compare the weight of an it...

  • Page 9

    500、1000( pieces) . 4) put the corresponding samples on the weighing pan, and then press is displayed momentarily before the display reverts to the sample size. → 5)remove the samples and put the load on, the scale begins to count. Note: 1. The larger the sample size, the more accurate unit 2. Symbo...

  • Page 10

    Vetek weighing ab press key m+ or key m- to choose setting item p11,(there are 12 available setting items ---p00-p11, indicated by auxiliary display at the lower left corner.) then press key unit/lo to start the unit setting menu. 2) how to set the init weight unit? Press key unit/lo to set init wei...

  • Page 11

    4) wait till appears, put weights jwl-6k, 2/3 of full load is 4 kg) 5) wait till appears, put weights of full load on the weighing pan (e.G. For jwl load is 6kg) 6) the calibration procedure ends up with a symbol 7) remove all the weights, and then press key single point calibration 1) press and hol...

  • Page 12

    3) press key unit/lo or key print/hi 4) press key mode to save the settings and go back to the weighing description of parameter values 1) p00 set the period of inactivity before the scale automatically turns off. Options are aut.No=non power-off, aut.5, aut.10, 2) p01 set check weighing buzzer beep...

  • Page 13

    Vetek weighing ab =enable =disable 7) p06 set the activation mode of backlight. Options are onoff (auto on with items greater than 9d placed on the weighing pan or any key is pressed), on (backlight on) and off (no backlight) 8) p07 set print mode. = manual print = stable print = continuous print 9)...

  • Page 14

    Hotline: 0909 899 833 13 vetek weighing ab 9. Troubleshooting and error message error status possible causes basic inspection & treatment e02 no.Samp pcs no sampling press down the key n/g to begin sampling. E04 ee.Err eeprom access error re-solder eeprom or contact service . E05 out.Ran overload co...