7.
ROOM CO TESTING
Select “Room CO” for CO investigations. Please refer to user manual.
ROOM CO display
ROOM CO
CO
P
00
TEST 14
LOG 01
-
Duration of this test is variable from 0 to 30 minutes as per BS7967
-
Real time CO reading, (ppm)
-
Test 00 = start. To stop the Room CO test press the PUMP button
-
The complete Room CO test is automatically stored as a log number
8.
OTHER DISPLAY CODES
-PO- = Pump Off
-OC- = Open Circuit on temperature input
9.
FOR BEST PRESSURE SENSOR ACCURACY
1)
Switch the analyser on for 5 minutes to let the temperature stabilise.
2)
Zero the pressure sensor when the analyser in the exact position that it will be
used.
10. TO FULLY CHARGE NiMH RECHARGEABLE BATTERIES
1)
The analyser must be switched on.
2)
Connect the charger and switched it on; charging indicator illuminates.
3)
Switch the analyser off; the display will show “BATTERY CHARGING”.
4)
The BAT status number of fully charged NiMH batteries is typically 70+