Kanomax 3888 User Manual

Manual is about: Handheld Particle Counter

Summary of 3888

  • Page 1

    I handheld particle counter model 3888/3889 user manual.

  • Page 3

    I list of components ■ standard items model functions body 3888 body (or 3889) ac adaptor 3888-10 operates the instrument with ac power and charge the internal rechargeable battery usb cable 3888-20 communicates with pc or printer zero count filter with tubing 3888-60 cleans the air flow path inside...

  • Page 4: Warning

    Ii important safety information the symbols for the warnings used in this manual are defined below: classifications warning warnings in this classification indicate risks that may result in serious injury or death if not observed. Caution warnings in this classification indicate risks that may resul...

  • Page 5: Caution

    Iii instructions provided in this manual. Failure to do so may cause electric shock, fire, or sensor damage. (caution)if abnormal noise, smell, or smoke is observed, or if liquid has entered the instrument, turn off the power immediately, remove the batteries or pull out the plug. Failure to observe...

  • Page 6

    Iv table of contents part names and functions ................................................................................................ 5 §1 getting started ................................................................................................................... 6 §2 charging batter...

  • Page 7

    5 part names and functions §1 the cradle is an optional accessary. Names functions 1. Inlet 2.83l/min suction volume outer diameter 6.4mm 2. Power/home button turns the power on/off to return to the main screen in operation 3. Usb port (for pc) connects with your pc by using the usb cable 4. Usb por...

  • Page 8

    6 getting started §2 charging batteries this instrument operates on the internal rechargeable battery or the provided ac adaptor. Connect the provided ac adaptor to start charging. It takes approximately 5 hours to charge the battery completely. If the battery’s run-time becomes shorter, the battery...

  • Page 9

    7 measurement use the general setup and measurement setup sub-menus to configure the necessary settings. Set the measurement conditions from the measurement mode. You can also select the measurement conditions from the preset.  return to the main screen. Press the start button to perform a measurem...

  • Page 10

    8 user interface §4 main screen  there are four configurations for the main screen that can be selected to best suit your purposes. (the figures below show the user interfaces of 6 channel model 3889.) from each main screen, you can move to sub-menu screens. Table cumulative table differential bar ...

  • Page 11

    9 icons names functions general setting opens the menus to set the unit, calendar, data, sound, language and to display instrument information measurement setting opens the menu to set the mode, alarm, save, remote and temperature/humidity preset opens the menu to set or select up to 18 presets hist...

  • Page 12

    10 icons names functions mode display displays the selected measurement modes mode display main repeat repeat single single continuous continuous calculation calculation iso iso gb gb displays the selected preset modes mode display preset 1 preset 2 preset 3 repeat p1(repeat) p2(repeat) p3(repeat) s...

  • Page 13

    11 icons names functions data confirms the information and perform operation on the entire measured data memory remaining: displays the ratio of the free space to the whole memory capacity. Number of records: displays the number of the stored records. Copy: copies all files to the usb flash memory. ...

  • Page 14

    12 measurement setting select [measurement setup] in the main screen to set the mode, alarm, save, remote, and temperature/humidity. Icons names functions mode selects 6 options of the measurement modes. Measurements will be performed under the latest mode setting. Repeat mode: repeats the measureme...

  • Page 15

    13 items description range location sets measurement locations by number. From 1 to 99 sampling time sets the sampling time of 1 cycle. From 6 sec. To 99 min. 59 sec. Repeat count sets the number of repeat. From 2 to 999 interval time sets the interval between the starting time of a measurement and ...

  • Page 16

    14 preset select [preset] in the main screen to set and select measurement conditions. Three options of preset number provide 6 measurement modes each; therefore 18 measurement conditions in total can be preset. When setting the preset, select one of the 18 measurement modes in the step 3, then set ...

  • Page 17

    15 print example the following figures show print examples of 6 channel model. (1) repeat mode (2) single, continuous mode (3) calculation mode 2017/3/30 11: 02 e= - - - repeat number 1 location 1 s-time 00:21 i -time 00:21 0.3um 16 cnt 0.5um 14 cnt 1.0um 12 cnt 3.0um 2 cnt 5.0um 1 cnt 10.0um 1 cnt ...

  • Page 18

    16 (4) iso mode (5) gb mode 2017/3/30 11: 02 e= - - - gb number from 1 to 6 s-time 00:21 i -time 00:21 - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - piont=001 times=001 0.5um 43 /m3 5.0um 0 /m3 piont=001 times=002 0.5um 7 /m3 5.0um 0 /m3 piont=002 times=001 0.5um 7 /m3 5.0um 0 /m3 ...

  • Page 19

    17 screen transition the screens are structured by 6 layers and the operations are mainly performed from the layer 2 and higher. The main screen for ready to measure is displayed when the instrument is not in the measuring process. The main screen for measurement is displayed when performing measure...

  • Page 20

    18 specifications §5 product handheld particle counter model 3888 (3-channel model) 3889 (6-channel model) particle measurement measuring method light scattering size distribution 6 channels (0.3, 0.5, 1.0, 3.0, 5.0, 10.0µm) 3 channels (0.3, 0.5, 5.0µm) flow rate 0.1cfm (2.83l/min) accuracy: ±5% (co...

  • Page 21

    19 communication standard usb (host: for printer and usb flash memory, device: for pc) cradle (option) ethernet, wi-fi, and rs485 (9600, 19200, 38400 baud) recording media media internal memory number and format up to 10,000 records in csv format language english, japanese, chinese, and spanish powe...

  • Page 22

    20 troubleshooting §6 symptoms possible causes(s) → solution(s) reference the display does not appear when the power is turned on. The ac adapter is not connected properly. → confirm the ac adapter and power cable. Low battery → replace the batteries. → recharge the batteries. 6 the battery drains f...

  • Page 23

    21 warranty and after service §7 the limited warranty set below is given by kanomax usa, inc. (hereafter referred to as “kui”) with respect to this instrument, its attachment parts including standard accessories (hereafter referred to as “product”) that you have purchased. Product you have purchased...

  • Page 24

    22 contact information §8 u.S.A. Kanomax usa, inc. 219 us hwy 206, andover, new jersey 07821 u.S.A. Tel: (800)-247-8887 / (973)-786-6386 fax: (973)-786-7586 url: http://www.Kanomax-usa.Com/ e-mail: info@kanomax-usa.Com japan kanomax japan inc. 2-1 shimizu, suita city, osaka 565-0805, japan tel: 81-6...

  • Page 25

    23.