Kanomax OMX-ADM Operation Manual

Manual is about: Handheld Odor Meter

Summary of OMX-ADM

  • Page 1

    Ver. 2 20160719 handheld odor meter omx-adm operation manual www. Globaltestsupply .Com find quality products online at: sales@globaltestsupply.Com.

  • Page 2

    2 table of contents 1. Introduction............................................................................................................... 4 1-1. Supplied components list ..................................................................................... 4 1-2. Warning .......................

  • Page 3

    3 4-4. Open software ...................................................................................................... 18 5. Specification ............................................................................................................. 19 6. Maintenance and servicing .................

  • Page 4

    4 1. Introduction 1-1. Supplied components list thank you for purchasing of shinyei handheld odor meter. Handheld odor meter is packaged carefully and includes all the components necessary for full operation. Upon receipt, check the contents and find you received the items as below in good condition...

  • Page 5

    5 1-2. Warning warning do not use any adaptors except supplied original ac adaptor. Do not use in explosive atmosphere. Do not use the device where combustible and/or explosive gasses exist. The device is not designed for explosion proof. Do not dismantle the base unit. Www. Globaltestsupply .Com fi...

  • Page 6

    6 1-3. Cautions cautions do not install, use, or store the unit at a place where: - ambient temperature is not in the range of -10 to 40 degree in c. (14 to 104 degree in f) - ambient humidity exceeds 90%r.H. - condensation occurs due to rapid temperature change. - corrosive and/or explosive gases e...

  • Page 7

    7 1-4. Unit dimension unit: mm www. Globaltestsupply .Com find quality products online at: sales@globaltestsupply.Com.

  • Page 8

    8 2. General description handheld odor meter is the most popular simplified tool for odor analysis which enables to show the relative odor strength numerically by comparing odor gases and purified air. The reading shown on the display is calculated based on our original principle. With the installat...

  • Page 9

    9 2-1. Features 2-1-1. Lcd display no. Name function ① ready blinks for about 10min for initializing when the unit is turned on. Start measuring when it stops blinking. ② odor strength displays odor strength in numeric value from 0 to 999. Start blinking when overflowed. ③ calendar function upper: y...

  • Page 10

    10 2-1-2. Base unit no. Name function ① hold 3 seconds for “peak-hold function”. (see 3-3.) hold and turn on the device at the same time to set calendar. ② select sampling rate. (see 3-2.) set calendar. ③ use for “zero-point function". (see 3-4.) set calendar. ④ confirm sampling rate and/or “zero-po...

  • Page 11

    11 2-2. Accessories name function ac adaptor connect to the unit during the measurement if necessary. Air purifying unit attach to the unit for “zero-point function”. (see 3-4.) activated carbon exchange activated carbon when deodorization using air purifying unit is not effective. Replacement filte...

  • Page 12

    12 3. Measurement 3-1. Real-time sampling mode 3-1-1. Measurement preparation 1. Place filter to suction nozzle. 2. Attach suction nozzle to the unit before measurement. 3. Use “aa” alkaline batteries or ac adaptor. Or 4. Turn on the unit at clean ambient. It takes about 10 min for the unit initiali...

  • Page 13

    13 3-1-2. Measurement 1. Turn on the unit and wait for about 10 minutes until “ready” stop blinking and the initializing is done. 2. Start measuring. 3. Use data filing software if necessary. 4. Turn off the unit when odor strength shows close to “0”. 3-2. Memory sampling mode 3-2-1. Measurement pre...

  • Page 14

    14 3-2-3. Memory overflow memory overflow occurs when the internal memory is full. If there is no space remained, calendar on lcd display starts blinking and data number becomes “0”. To delete saved memory, press and at the same time while the unit is turned off. Keep pressing and and turn on the de...

  • Page 15

    15 3-3. Peak-hold function press for over 3 seconds during real-time sampling. By using this function, the highest value of odor strength keeps being displayed until the unit measures higher value. Note press for over 3 seconds again to cancel the function. Note the function is not displayed on the ...

  • Page 16

    17 how to install usb drivers software install usb drivers software before connecting usb cable to pc. 1. Download the latest software from the link as below. Https://www.Silabs.Com/products/mcu/pages/usbtouartbridgevcpdrivers.Aspx 2. Install the software. Note if pc cpu is 32bit, select “cp210xvcpi...

  • Page 17

    18 4-3. Preparation connect the unit to pc by usb connection cable. 4-4. Open software 1. Double-click shortcut icon on desktop, or open the file from program files. 2. Confirm main screen appears. 3. Read “handheld odor meter data filing software manual” for details. Www. Globaltestsupply .Com find...

  • Page 18

    19 5. Specification detection method semiconductor gas sensors object gas hydrogen sulfide, methyl mercaptan, ammonia, etc. Sampling method continuous sampling by built-in pump detection range 0 ppm to 25 ppm approx. (ammonia concentration) display digital lcd display odor strength (0 to 999) odor i...

  • Page 19

    20 6. Maintenance and servicing 6-1. Replacement of “aa” alkaline batteries replace “aa” alkaline batteries when battery status on lcd display becomes “low”. 6-2. Replacement of button battery the unit uses a button battery (model: cr2032) for memory back up. When to replace: - pasts 5 years approx....

  • Page 20

    21 6-3. Maintenance of air purifying unit replace activated carbon and filters in air purifying unit to prevent gases absorbed in the unit. 1. Twist the air purifying unit to open. 2. Take out top wire gauze, top filter, activated carbon, and bottom filter. Note do not take out bottom wire gauze. 3....

  • Page 21

    22 6-5. Troubleshooting if you find: check: ① the unit does not operate. Check battery status or connect ac adaptor. Check battery polarity. ② odor strength does not change. Get the unit closer to object. Connect air purifying unit to clean the absorbed gas in the unit. Clean nozzle with alcohol and...