Summary of KATflow 150

  • Page 1

    Operating instructions ka tflow 150 ultrasonic flowmeter.

  • Page 2

    Operating instructions katflow 150 version v21en160420 copyright © 2016 all rights reserved..

  • Page 3

    Katflow 150 operating instructions table of contents page 1 safety instructions, legal requirements, warranty, return policy.................5 1.1 symbols used in these operating instructions........................................5 1.2 safety instructions...............................................

  • Page 4

    5.4.3 hart compatible output...........................................................35 5.4.4 analogue current output 0/4 ... 20 ma......................................35 5.4.5 analogue voltage output 0-10 v................................................35 5.4.6 analogue frequency output 2...10k...

  • Page 5: •

    Katflow 150 1 safety instructions, legal requirements, warranty, return policy 1 safety instructions, legal requirements, war- ranty, return policy 1.1 symbols used in these operating instructions danger this symbol represents an immediate hazardous situation which could result in serious injury, de...

  • Page 6

    Katflow 150 1 safety instructions, legal requirements, warranty, return policy 1.3 warranty any product purchased from katronic is warranted in accordance with the relevant product documentation and as specified in the sales contract provided it has been used for the purpose for which it has been de...

  • Page 7

    Katflow 150 2 introduction 2 introduction the katflow 150 is an ultrasonic flowmeter employing clamp-on sensors for the measurement of liquids in full, enclosed pipes. Flow measurements can be under- taken without interruption of the process or interference with the integrity of the pipeline. The cl...

  • Page 8

    Katflow 150 2 introduction system configuration a maximum of 2 sensor pairs can be installed - if two pairs are installed these can be configured either in a 1-pipe dual-path or a 2-pipe single path configuration. 8 illustration 4: kf150 2-pipe 1-path configuration using optional junction boxes illu...

  • Page 9

    Katflow 150 3 installation 3 installation 3.1 unpacking and storage 3.1.1 unpacking care should be taken when opening the box containing the flowmeter, any mark- ings or warnings shown on the packaging should be observed prior to opening. The following steps should then be taken: unpack the flowmete...

  • Page 10

    Katflow 150 3 installation 3.2 clamp-on sensor installation the correct selection of the sensor location is crucial for achieving reliable meas- urements and high accuracy. Measurement must take place on a pipe in which sound can propagate (see acoustic propagation) and in which a rotationally sym- ...

  • Page 11

    Katflow 150 3 installation for a free inlet or outlet pipe section: select the measuring point at a location where the pipe cannot run empty. Correct disadvantageous correct disadvantageous for a vertical pipe: select the measuring point at a location where the liquid flows upward to ensure that the...

  • Page 12

    Katflow 150 3 installation disturbance source: 2 x 90°-elbows in different planes inlet outlet l ≥ 40 d l ≥ 5 d disturbance source: t-section inlet outlet l ≥ 50 d l ≥ 10 d disturbance source: diffuser inlet outlet l ≥ 30 d l ≥ 5 d disturbance source: reducer inlet outlet l ≥ 10 d l ≥ 5 d disturbanc...

  • Page 13

    Katflow 150 3 installation 3.4 pipe preparation clean dirt and dust from around the area of the pipework where the sensors are to be placed. Remove loose paint and rust with a wire brush or file. Firmly bonded paint does not necessarily need to be removed provided the flow- meter diagnostics indicat...

  • Page 14

    Katflow 150 3 installation 3.6 flowmeter installation 3.6.1 wall mounting the katflow 150 is a wall mounted device and can be installed using suitable screws and wall plugs according to the following drawings. 14 drawing 1: outline dimensions katflow 150 ultrasonic flowmeter flowmeter outline di- me...

  • Page 15

    Katflow 150 3 installation make sure that the ambient temperature is within the -10 ... 60 °c operating tem- perature range specified for the flowmeter unit. 15 drilling aid for wall mounting 243.5 1 2 0 c r s 9 0 c r s !.

  • Page 16

    Katflow 150 3 installation 3.6.2 electrical connections please note that in order to supply the unit with mains power, the equipment must be protected by suitably sized switches and circuit breakers. 100 ... 240 v ac, 50/60 hz 10 w 9 ... 36 v dc 10 w 16 ! Electrical wiring !.

  • Page 17

    Katflow 150 3 installation 3.7 clamp-on sensor mounting before the sensors can be mounted the installation location should have been determined, a sensor mounting method should be chosen, the flowmeter must be mechanically and electrically installed, the sensors must be connected to the flowmeter. D...

  • Page 18

    Katflow 150 3 installation 3.7.3 correct positioning of the sensors always mount the transducer pair so that the free front edges of the sensors face each other. There is a different engraving on the top of each transducer. The transducers are mounted correctly if the engravings on the two transduce...

  • Page 19

    Katflow 150 3 installation press the sensors firmly onto the pipe. There should be no air pockets between the transducer surface and the pipe wall. Using a measuring tape, adjust the sensor separation distance as sugges- ted by the flowmeter. When the sensor positioning screen (section 3.3) is displ...

  • Page 20

    Katflow 150 4 operation 4 operation 4.1 switching on/off the flowmeter is switched on by connecting the power supply to the instrument. Disconnecting the external supply switches off the flowmeter. 4.2 keypad and display 4.2.1 keypad key functions key main function secondary function character entry...

  • Page 21

    Katflow 150 4 operation character entry: j k l 5 > character entry: m n o 6 $ q + = reset positive total value character entry: p q r s 7 toggle multiplexer (where mul- ti-channel functions are provided) character entry: t u v 8 * q off = stop totaliser function character entry: w x y z 9 move menu/...

  • Page 22

    Katflow 150 4 operation 4.2.2 display functions display icon function on not used off on datalogger recording off datalogger switched off on not used off on backlight switched on off backlight switched off on i/o processor error off i/o processor functioning correctly on without strike-through: spea...

  • Page 23

    Katflow 150 4 operation on time/date set off clock error on error recorded in error log off no error detected on serial communication on (rs232 and/or rs485) off serial communication off "l", "lt" or "t" displays whether flow is laminar, turbulent or mixed 4.3 quick setup wizard the quick setup wiza...

  • Page 24

    Katflow 150 4 operation enter outside pipe diameter using alphanumer- ical keys and confirm by pressing enter>. Use key up> as character backspace clear to correct for data entry errors. If 0 is entered, an additional screen appears that allows entering the pipe circumference. Enter pipe wall thickn...

  • Page 25

    Katflow 150 4 operation select transducer configuration (number of passes) using cursor keys. Auto automatically 1 1 pass, diagonal mode 2 2 passes, reflection mode 3 3 passes, diagonal mode 4 4 passes, reflection mode..Etc. Confirm by pressing enter>. Use cursor keys to select start measurement. Co...

  • Page 26

    Katflow 150 4 operation 4.4.2 3-line display format display screen operation the three-line display screen is configure- able to show flow, totalizers and diagnostic functions. Change to diagnostic displays by pressing and to totalizer screens by press- ing . Cycle through display screens using . 4....

  • Page 27

    Katflow 150 4 operation 4.4.5 dual pv display (multi-channel meters) display screen operation line 1 shows the pv on the selected channel. Line 2 shows the selected units. Line 3 shows the pv on the other channel (in its selected units) change to diagnostic displays by pressing and to totalizer and ...

  • Page 28

    Katflow 150 5 commissioning 5 commissioning 5.1 menu structure main menu menu level 1 menu level 2 description/settings quick start setup wizard ch1 sensor type indication of sensor type and serial number if automatically detected, otherwise select from list ↑↓ k1n,k1l,k1e,k1ex,k1p, k4n,k4l,k4e,k4ex...

  • Page 29

    Katflow 150 5 commissioning temperature -30 ... 300 °c liner material select from list ↑↓ none, epoxy, rubber, pvdf, pp, glass, cement, user (liner c-speed) liner c-speed only if lining material selected 600 ... 6553.0 m/s liner thickness only if lining material selected 1.0 ... 99.0 mm passes selec...

  • Page 30

    Katflow 150 5 commissioning lining material select from material list ↑↓ thickness 1 ... 99 mm c-speed 600 ... 6553.0 m/s passes passes select from list ↑↓ display select channel channel 1, channel 2 select line of display (top, middle, bottom) select from unit list ↑↓ damping reduces fluctuations i...

  • Page 31

    Katflow 150 5 commissioning error defines output behaviour in the event of error select from list ↑↓ pulse out source select from list ↑↓ units select from unit list ↑↓ mode select from list ↑↓ alarm (select on point, off point) pulse (select value, width) linear (select min value, max value, dampin...

  • Page 32

    Katflow 150 5 commissioning zero cal zero calibration settings zero perform auto zero calibration yes no track track zero offset yes no delta zero flow delta time offset in ns, read from sensor prom or entered directly for special sensors timeup upstream transit-time offset in µs, allows for fixed d...

  • Page 33

    Katflow 150 5 commissioning defaults reload factory default settings, except for date and time yes no key lock locks the keypad until password is entered (four number keys followed by “enter”). See also “password” above. Diagnostics temperature shows control unit temperature log memory percentage of...

  • Page 34

    Katflow 150 5 commissioning 5.2 diagnostics diagnostic screens can be viewed directly during measurement using the 3/disp and 1/next keys, or through the menu structure. 5.3 display settings customer specific settings for data to be displayed can be achieved using the ap- propriate menu items to sel...

  • Page 35

    Katflow 150 5 commissioning 5.4.3 hart compatible output the kf100 can also be configured with an optional module which responds to out- put commands conforming to the hart protocol. Please refer to customer support for further information. Hart® is a registered trademark of the hart communication f...

  • Page 36

    Katflow 150 5 commissioning 5.4.6 analogue frequency output (passive) frequency outputs may be assigned to process values in the “mode” section of the output menu. The outputs can be programmed and scaled within the menu struc- ture. Wiring electrical characterist- ics galvanically isolated from mai...

  • Page 37

    Katflow 150 5 commissioning 5.5 input configuration 5.5.1 pt100 inputs wiring electrical characterist- ics 3 and 4 wire options. Galvanically isolated from main electronics and from other i/o's. Temperature: range -50 ... 400 °c. Resolution: 0.01 k. Accuracy: ±0.1 k. 5.5.2 analogue current input 0/4...

  • Page 38

    Katflow 150 5 commissioning 5.7 sound velocity measurement (svm) the measured sound velocity (sos) is available as a process value and a dia- gnostic function (where specified) during measurement and may be applied to a process output by selecting “c m/s” from the appropriate menu. 5.8 dual-channel ...

  • Page 39

    Katflow 150 6 maintenance 6 maintenance katflow flowmeters are maintenance free concerning the flow measurement func- tions. Within the scope of periodic inspections, regular inspection for signs of dam - age or corrosion is recommended for the transducers, the junction box if installed, and the flo...

  • Page 40

    Katflow 150 7 troubleshooting 7 troubleshooting should there be the need to call customer service, please let us know the following details: model code serial number sw, hw revision error log list possible error messages may include the following: error message group description error handling usb i...

  • Page 41

    Katflow 150 7 troubleshooting comm hs1 err hardware internal communication error power on/off, otherwise call customer support comm read ave err hardware internal communication error power on/off, otherwise call customer support comm read raw err hardware internal communication error power on/off, o...

  • Page 42

    Katflow 150 8 technical data 8 technical data sound speed* shear wave (at 25 ºc) material m/s ft/s steel, 1% carbon, hardened 3,150 10,335 carbon steel 3,230 10,598 mild steel 3,235 10,614 steel, 1% carbon 3,220 10,565 302 stainless steel 3,120 10,236 303 stainless steel 3,120 10,236 304 stainless s...

  • Page 43

    Katflow 150 8 technical data all data given at 25 ºc (77 ºf) unless otherwise stated sound speed change v/ºc kinematic viscosity substance chemical formula specific gravity m/s ft/s m/s/ºc mm 2 /s x10-6 ft 2 /s acetic anhydride (ch3co)2o 1.082 (20 ºc) 1,180 3,871.4 2.5 0.769 8.274 acetic acid, anhyd...

  • Page 44

    Katflow 150 8 technical data methane ch4 0.162 (-89 ºc) 405 (-89 ºc) 1,328.7 (-128 ºf) 17.5 methanol ch4o 0.791 (20 ºc) 1,076 3,530.2 292 0.695 7.478 methyl acetate c3h6o2 0.934 1,211 3,973.1 0.407 4.379 methyl alcohol ch4o 0.791 1,076 3,530.2 292 0.695 7.478 methyl benzene c7h8 0.867 1,328 (20 ºc) ...

  • Page 45

    Katflow 150 8 technical data tetrachloroethane c2h2cl4 1553 (20 ºc) 1,170 (20 ºc) 3,838.6 (68 ºf) 1.19 12.804 tetrachloro-ethene c2cl4 1.632 1,036 3,399 tetrachloro-methane ccl4 1.595 (20 ºc) 926 3,038.1 0.607 6.531 tetrafluoro-methane (freon 14) cf4 1.75 (-150 ºc) 875.24 (- 150 ºc) 2,871.5 (-283 ºf...

  • Page 46

    Katflow 150 8 technical data 34 93.2 1517 4977 35 95.0 1519 4984 36 96.8 1521 4984 37 98.6 1523 4990 38 100.4 1525 4997 39 102.2 1527 5010 40 104.0 1528 5013 41 105.8 1530 5020 42 107.6 1532 5026 43 109.4 1534 5033 44 111.2 1535 5036 45 113.0 1536 5040 46 114.8 1538 5046 47 116.6 1538 5049 48 118.4 ...

  • Page 47

    Katflow 150 8 technical data 88 190.4 1551 5089 89 192.2 1551 5089 90 194.0 1550 5086 91 195.8 1549 5082 92 197.6 1549 5082 93 199.4 1548 5079 94 201.2 1547 5076 95 203.0 1547 5076 96 204.8 1546 5072 97 206.6 1545 5069 98 208.4 1544 5066 99 210.2 1543 5063 100 212.0 1543 5063 104 220.0 1538 5046 110...

  • Page 48

    Katflow 150 9 specification 9 specification general measuring principle : ultrasonic time difference correlation principle flow velocity range : 0.01 ... 25 m/s resolution : 0.25 mm/s repeatibility : 0.15 % of measured value ±0.015 m/s accuracy : volume flow ± 1 ... 3 % of measured value depending o...

  • Page 49

    Katflow 150 9 specification process inputs / process outputs (maximum of ten per instrument) inputs temperature : pt 100, three or four-wire circuit, measuring range - 50 ... 400 °c, resolution 0.1k, accuracy ±0.2 k current : 0 ... 20 ma active or 4 ... 20 ma passive, u = 30 v, r i = 50 ohm, accurac...

  • Page 50

    Katflow 150 10 index 10 index 50 acoustic coupling gel 17 negative separation distance 13 analogue current input 37 outer diameter 24 analogue current output 35 output configuration 34 analogue frequency output 36 output settings 34 analogue voltage output 35 packaging 9 certificate of conformity 51...

  • Page 51

    Katflow 150 appendix a appendix a certificate of conformity 51.

  • Page 52

    Katflow 150 appendix b appendix b customer return note (crn) company address name tel. No. E-mail instrument model katronic contract no. (if known) serial number sensor type(s) sensor serial number(s) the enclosed instrument has been used in the following environment (please √): nuclear radiation wa...