Keithley 2601B Quick Start Manual - Connect The Instrument

Other manuals for 2601B: Reference Manual, Quick Start Manual
Manual is about: 2600B series System SourceMeter Instrument

Summary of 2601B

  • Page 1

    Safety introduction models 2634b, 2635b, and 2636b system sourcemeter instruments quick start guide.

  • Page 2: Safety Precautions

    Safety introduction safety precautions the following safety precautions should be observed before using this product and any associated instrumentation. Although some instruments and accessories would normally be used with nonhazardous voltages, there are situations where hazardous conditions may be...

  • Page 3

    Safety introduction the warning heading in the user documentation explains dangers that might result in personal injury or death. Always read the associated information very carefully before performing the indicated procedure. The caution heading in the user documentation explains hazards that could...

  • Page 4

    *see full specifications for other pulse regions. Carefully consider and configure the appropriate output-off state, source levels, and compliance levels before connecting the instrument to a device that can deliver energy. Failure to consider the output-off state, source levels, and compliance leve...

  • Page 5: Introduction

    Safety introduction software for the series 2600b instruments is available for download from the keithley web page. Go to www.Tek.Com/keithley and search for the specific software you need. • test script builder (tsb) integrated development environment: provides an environment to develop a test prog...

  • Page 6

    Unpack and inspect the instrument to unpack and inspect the instrument: 1. Inspect the box for damage. 2. Open the top of the box. 3. Remove the documentation and accessories. 4. Carefully lift the instrument out of the box. 5. Remove the packaging insert. 6. Inspect the instrument for any obvious s...

  • Page 7

    Safety introduction in addition to the series 2600b system sourcemeter ® instrument, you should have received: 1. Keithley usb flash drive containing the model 2400 software emulation personality script. 2. Interlock db-25 male connector kit hardware. 3. One rj-45 lan crossover cable for the 2636b. ...

  • Page 8: Connect The Instrument

    • make sure any test fixture connected to the system protects the operator from contact with hazardous voltages, hot surfaces, and sharp objects. Use shields, barriers, insulation, and safety interlocks to accomplish this. • cover the device under test (dut) to protect the operator from flying debri...

  • Page 9: Install The Instrument

    Safety introduction install the instrument the 2600b can be used on a bench or in a rack. Please see the instructions that came with your rack-mount kit if you are installing the 2600b in a rack. Connect line power the series 2600b operates from a line voltage of 100 v to 240 v at a frequency of 50 ...

  • Page 10: Turn On The Instrument

    To connect line power: 1. Make sure the front panel power switch is in the off (0) position. 2. Connect the socket of the supplied power cord to the power connection on the rear panel. 3. Connect the plug of the power cord to a grounded ac outlet. Operating the instrument on an incorrect line voltag...

  • Page 11: Power-Up Sequence

    Safety introduction power-up sequence when the instrument is turned on, you should see: • a series of dots. • all segments of the display light. • a brief display showing the instrument model. For example, if the instrument is a 2635b, the display will show “keithley model 2635b.” • line frequency d...

  • Page 12: Test The Instrument

    Test the instrument the following test verifies basic operation of the 2634b, 2635b, and 2636b. In this test, you will use the front-panel controls shown below to source voltage and measure the voltage output. You do not need to connect a device under test (dut) for this test. Step 1: set source fun...

  • Page 13

    Safety introduction the main display screen reappears: 3. Press the cursor keys to move the cursor to the 10s digit. 4. Press the navigation wheel to enter edit mode. The edit indicator appears in the upper left corner of the display. 5. Turn the navigation wheel to set the source value to 20.0000 v...

  • Page 14

    Step 3: set measurement function and range 1. Press the meas key as many times as needed to select the v (voltage) measurement function. In the following figure, the measurement function has been set to “v.” 2. Press the auto key as many times as needed to select the auto range function. When auto i...

  • Page 15

    Step 6: turn the output off when you finish making measurements, turn the output off by pressing the output on/off control or control. The output indicator led turns off. For a basic understanding of the of the series 2600b functionality, we strongly recommend that first-time users become familiar w...

  • Page 16: Faqs

    Faqs what should i do if i see an error message when i turn the instrument on? If an error message is displayed, press the exit (local) key. The series 2600b will return to the default display screen. For detailed information about error messages, see “errors and status messages” in the series 2600b...

  • Page 17: Next Steps

    Safety introduction next steps the series 2600b reference manual refer to the series 2600b reference manual for detailed information about all features of the instrument. Keithley instruments website see: www.Tek.Com/keithley for support and additional information about the instrument. Connect unpac...

  • Page 18

    Contact information: australia* 1 800 709 465 austria 00800 2255 4835 balkans, israel, south africa and other ise countries +41 52 675 3777 belgium* 00800 2255 4835 brazil +55 (11) 3759 7627 canada 1 800 833 9200 central east europe / baltics +41 52 675 3777 central europe / greece +41 52 675 3777 d...