    Model 7708 40-channel differential multiplexer with automatic cjc user’s guide pa-744 rev. C / 8-03 a g r e a t e r m e a s u r e o f c o n f i d e n c e.

  • Page 2: Warranty

    Warranty keithley instruments, inc. Warrants this product to be free from defects in material and workmanship for a period of 1 year from date of shipment. Keithley instruments, inc. Warrants the following items for 90 days from the date of shipment: probes, cables, rechargeable batteries, diskettes...

  • Page 3: Safety Precautions

    Safety precautions the following safety precautions should be observed before using this product and any associated instrumentation. Although some in- struments and accessories would normally be used with non-haz- ardous voltages, there are situations where hazardous conditions may be present. This ...

  • Page 4

    If a screw is present, connect it to safety earth ground using the wire recommended in the user documentation. The symbol on an instrument indicates that the user should re- fer to the operating instructions located in the manual. The symbol on an instrument shows that it can source or mea- sure 100...

  • Page 5: Model 7708

    Pa-744 rev. C / 8-03 model 7708 40-channel differential multiplexer with automatic cjc user’s guide keithley instruments, inc. 28775 aurora road cleveland, ohio 44139 (440) 248-0400 fax: (440) 248-6168 www.Keithley.Com introduction the model 7708 is a 40-channel differential multiplexer card with th...

  • Page 6

    2 specifications the specifications for the model 7708 switching card are located at the end of this user’s guide. Operating instructions switching card operation is covered in the model 27xx user’s manual. The model 7708 operates the same as the model 7700 switching card with the following differen...

  • Page 7: Card Configuration

    3 card configuration description figure 1 shows a simplified schematic diagram of the model 7708 module. As shown, the model 7708 has channels that are grouped into two banks of twenty channels (forty channels total). Backplane isolation is provided for each bank. Each bank also includes separate co...

  • Page 8

    4 channel 41 (2w/4w configuration), channel 42 (sense isolation), and channel 43 (input isolation) are normally automatically configured by the model 27xx. However, by using the :rout:mult: commands (refer to section 2 of the model 27xx user’s manual), they can be manually configured. Note connect 4...

  • Page 9

    5 figure 3 model 7708 screw terminal channel designations h l input ch15 h l ch16 h l ch17 h l ch18 h l ch19 h l ch20 h l ch35 h l ch36 h l ch37 h l ch38 h l ch39 h l ch40 h l ch21 h l ch22 h l ch23 h l ch24 h l ch25 h l ch26 h l ch27 h l ch28 h l ch29 h l ch30 h l ch31 h l ch32 h l ch33 h l ch34 h ...

  • Page 10

    6 wiring procedure warning the information in this section is intended for qualified service personnel. Do not attempt to perform this procedure unless qualified to do so. Use the following procedure to wire the model 7708 module. Make all connections using the correct wire size (up to 20 awg). Also...

  • Page 11

    7 typical connections the following examples show typical wiring connections for the following types of measurements: • thermocouple connections, see figure 5 • Ω 2-wire and thermistor connections, see figure 6 • Ω 4-wire and rtd connections, see figure 7 • voltage (ac or dc), see figure 8 figure 5 ...

  • Page 12

    8 figure 7 Ω 4-wire and rtd connections figure 8 voltage connections (dc or ac) channel 1 hi lo channel 20 hi lo (channels 2–19) channel 21 hi lo channel 40 hi lo (channels 22–39) resistor or 4-wire rtd resistor or 4-wire rtd channel 1 hi lo channel 40 hi lo (channels 2–39) + + dc voltage ac voltage.

  • Page 13

    9 connection log make a copy of table 1 and affix it to the cover of the model 7708. Use this to record connection information and channel descriptions as needed. Table 1 connection log model 7708 channel color description description color channel input h l sense h l ch1 h h ch21 l l ch2 h h ch22 l...

  • Page 14: Performance Verification

    10 performance verification the performance of the model 7708 card is tested by verifying measurement accuracy through the card. If verification limits are met through the front panel terminals of the model 27xx, they should also be met through the card. Warning the information in this section is in...

  • Page 15

    11 line power the model 27xx requires a line voltage of 100v/120v/220v/240v, ±10% and a line frequency of 47.5hz to 63hz. Note that the line frequency is automatically sensed at power-up, but the line voltage must be manually set to either 100v/120v or 220v/240v as described in section 3 of the mode...

  • Page 16

    12 example reading limit calculation the following is an example of how reading limits have been calculated. Assume you are testing the 10v dc range using a 10v input value. Using the model 27xx one-year accuracy specification for 10v dc of ± (30ppm of reading + 5ppm of range), the calculated limits...

  • Page 17

    13 performance verification procedures note the following procedures check one channel (ch1) or one channel pair (ch1 and ch21) of the model 7708 card. To check other channels (or channel pairs), modify the procedures by connect- ing the verification equipment to the appropriate channel (or channel ...

  • Page 18

    14 3. Select the dc volts function by pressing the dcv key, and set the model 27xx to the 100mv range. Close channel 1 by pressing the close key and then keying in 101. 4. Set the calibrator output to 0.00000mv dc, and allow the reading to settle. 5. Enable the model 27xx rel mode. Leave rel enabled...

  • Page 19

    15 figure 10 connections for ac volts verification 4. Set the model 27xx for the 100mv range; make sure that rel is disabled. 5. Source 1khz and 50khz ac voltages for each of the ranges summarized in table 4 , and make sure that the respective model 27xx readings fall within stated limits. 6. Press ...

  • Page 20

    16 verifying resistance check resistance by connecting accurate resistance values to the model 7708 and verifying that its resistance readings are within the specified limits. Caution do not apply more than 300v between plug-in module input or sense h and l terminal, or between any adjacent channels...

  • Page 21

    17 5. Set the model 27xx for the 100 Ω range, and make sure the filter is on. Enable ocomp (offset-compensated ohms) for the 100 Ω range test. (press shift then ocomp.) 6. Recalculate reading limits based on actual calibrator resistance values. 7. Source the nominal full-scale resistance values for ...

  • Page 22

    18 8. Connect the model 7708 ch1 and ch21 terminals to the calibrator as shown in figure 12 . 9. Disable external sense on the calibrator. 10. Set the model 27xx for the 100m Ω range. 11. Source a nominal 100m Ω resistance value, and verify that the reading is within calculated limits for the 100mΩ ...

  • Page 23

    19 verifying temperature thermocouple, thermistor, and rtd temperature readings are derived from dc volts and resistance measurements, respectively. For that reason, it is not necessary to independently verify the accuracy of temperature measurements. As long as the dc volts and resistance functions...

  • Page 24

    20 4. Configure the model 27xx for °c units, type k temperature sensor, and internal reference junction as follows: a. Press shift then sensor, and note the unit displays the temperature units: units: c. (if necessary, use the cursor and range keys to select °c units.) b. Press enter. The unit then ...

  • Page 25

    21 verifying frequency follow the steps below to verify the model 27xx frequency function: 1. Connect the function generator to the model 7708 ch1 h and l input terminals. (see figure 14 .) 2. Install the model 7708 in slot 1 of the model 27xx, then turn on the power, and allow the unit to warm up f...

  • Page 26

    22 verifying ratio and average follow the procedure below to verify ratio and average. Caution exceeding 300v between plug-in module input or sense h and l terminals may cause instrument damage. 1. Connect the model 7708 ch1 and ch21 h and l terminals to the dc calibrator, as shown in figure 15 . 2....

  • Page 27: Calibration

    23 calibration the following procedures calibrate the temperature sensors on the model 7708 plug-in modules. Warning the information in this section is intended only for qualified service personnel. Do not attempt these procedures unless you are qualified to do so. Recommended test equipment in orde...

  • Page 28: Replaceable Parts

    24 remote model 7708 calibration 1. Connect the model 7708 to the 7797 calibration/extender board (see “extender board connections” above). 2. With the power off, install the model 7708/7797 combination in slot 1, and select the rear inputs with the inputs switch. Allow three minutes for thermal equ...

  • Page 29

    25 table 9 model 7708 parts list circuit designation description keithley part no. C1, c3, c4, c7, c9-c11, c14, c200, c201, c203-c213 cap, 0.1µf, 20%, 50v, ceramic c-418-.1 c2, c6, c17-c20, c23-c25, c202 cap, 47pf, 5%, 100v, ceramic c-465-47p c5 cap, 4.7µf, 10%, 35v, tantalum c-476-4.7 c16 cap, 220µ...

  • Page 30

    26 ordering information to place an order, or to obtain information concerning replacement parts, contact your keithley representative or the factory (see inside front cover for addresses). When ordering parts, be sure to include the following information: • card model number (model 7708) • card ser...

  • Page 31: 7708

    Sbg 11/7/00 rev. B general 40 channels: 40 channels of 2-pole relay input. All channels configur- able to 4-pole. Relay type: latching electromechanical. Actuation time: specified for model 2700, firmware revision bo2 or later. Capabilities channels 1–40: multiplex one of 40 2-pole or one of 20 4-po...

  • Page 34

    Specifications are subject to change without notice. All keithley trademarks and trade names are the property of keithley instruments, inc. All other trademarks and trade names are the property of their respective companies. Keithley instruments, inc. 28775 aurora road • cleveland, ohio 44139 • 440-...