Kemper Figure 138 00 002 Installation And Operating Instructions Manual

Manual is about: ‘Control-plus‘ Hand-held measuring instrument for sensors

Summary of Figure 138 00 002

  • Page 1

    - 1 - installation and operating instructions kemper ‘control-plus‘ hand-held measuring instrument for sensors figure 138 00 002 table of contents 1. General........................................................................................................................ P.2 2. Safety instruct...

  • Page 2

    - 2 - 1. General the kemper ´control-plus’ hand-held measuring instrument in combination with the kemper ´control-plus’ flow and temperature measurement valve figure 138 4g and figure 638 4g is used for precisely determining volume rates and temperatures and for documenting the measurements in potab...

  • Page 3

    - 3 - 3.Shipping and intermediate storage 3.1 transport and shipping the device must be properly shipped. The device was checked for compliance with all stated data before dispatching. The device should therefore be in a flawless electrical and mechanical condition on arrival. To convince yourself o...

  • Page 4

    - 4 - 6. Menu control / "display" "menu" function 6.1 control panel the hand-held measuring instrument is set and controlled using a four key control panel. Key 1 esc: exit the menu – switch between display / main menu key 2 ok: acknowledgement key - switch between display / units menu key 3 : scrol...

  • Page 5

    - 5 - measurement display, example 2: measurements with measurement module and figure 138 4g/638 4g measurement display, example 3: measurements with measurement module and various sensors general sensors 0 (4) – 20 ma 0- 10v temperature sensor valve pt1000 figure plug-in temperature sensor pt1000 f...

  • Page 6

    - 6 - 6.5 operating main menu additional controls in the menu: e.G. To set times, dates "scroll back/forward": to select the submenuc arrows indicate the selected 1 x "ok" changes into the submenu "scroll back / forward": select in the menu „ok“ confirm point / change „esc“ back arrow points to numb...

  • Page 7

    - 7 - 6.6 operating units menu 6.7 password protection password protection factory setting is 0000 and can be changed in the system settings menu. The following functions are password protected: - system settings/load factory settings (resets all settings) -data logger/erase memory (the entire memor...

  • Page 8

    - 8 - 7.1 connection with 2m, m12x1 connection cable the 2m, m12x1 ´control-plus’ connection cable is required to connect the hand-held measuring instrument to the flow sensor (included in the scope of delivery). 7.2 sensor rating plate / measurement range / code letter the correct sensor type 138 4...

  • Page 9

    - 9 - figure 138 4g 025: figure 638 4g 020: measurement range 5-85l/min. ►d◄ measurement range 5-85l/min. ►d◄ figure 138 4g 032: figure 638 4g 025: measurement range 9-150l/min. ►e◄ messbereich 9-150l/min. ► e◄ figure 138 4g 040: measurement range 12-216l/min. ►f◄ figure 138 4g 050: measurement rang...

  • Page 10

    - 10 - 8. Connecting ´control-plus’ pressure sensor figure 138 00 006 8.1 connection with 2m, m12x1 connection cable the 2m, m12x1 ´control-plus’ connection cable is required to connect the hand-held measuring instrument to the flow sensor (optionally available, both ends are marked yellow). 8.2 mak...

  • Page 11

    - 11 - 9. Measuring with the sensor measuring module figure 138 00 011 9.1 application the kemper sensor measurement module is used to read out measurement data, in combination with the hand-held measuring instrument, which originate in a sensor in, e.G., an inaccessible installation area. The senso...

  • Page 12

    - 12 - - kemper ´control-plus´ flow and temperature sensor valve, figure 138 4g - khs vortex flow sensor, figure 638 4g - khs pt1000 temperature sensor valve, figures 628/629 - temperature sensor pt100, figure t5100 140 00 003 - any sensor with 4-20 ma optional 0-20 ma output - any sensor with 0-10v...

  • Page 13

    - 13 - 9.4 measurement module configuration for flow and temperature sensor valve, figure 138 4g, figure 638 4g 1. Use esc in the main menu to switch (if not yet displayed in the main menu) 2. Select measuring module submenu 3. Enter module name, e.G. Room -3.0 4. Sensor selection: sensor 138 4g/638...

  • Page 14

    - 14 - 9.5 measurement module configuration for temperature measurement fitting pt1000 figure 628 0g and 629 0g and temperature sensor pt1000 pt1000, figure 628/629 0g pt1000, part no. T5100 140 00 003 1. Use esc in the main menu to switch (if not yet displayed in the main menu) 2. Select measuring ...

  • Page 15

    - 15 - 9.7 measurement module configuration for 0-20ma and 4-20ma 1. Use esc in the main menu to switch (if not yet displayed in the main menu) 2. Select measuring module submenu 3. Enter module name, e.G. Room -3.0 4. Sensor selection 0-20ma or 4-20ma 5. Stipulate value at (0)4 ma, 20ma and unit of...

  • Page 16

    - 16 - 9.8 measurement module configuration for 0 – 10v 1. Use esc in the main menu to switch (if not yet displayed in the main menu) 2. Select measuring module submenu 3. Enter module name, e.G. Room -3.0 4. Sensor selection 0-10v 5. Stipulate value at 0v, 1 0v and unit of measurement e. G. 0v / 0....

  • Page 17

    - 17 - 9.9 measurement module configuration for 0-500hz frequency signal, square wave signal min. 3v – max. 24v 1. Use esc in the main menu to switch (if not yet displayed in the main menu) 2. Select measuring module submenu 3. Enter module name, e.G. Room -3.0 4. Sensor selection 5. Stipulate value...

  • Page 18

    - 18 - 10.3 display saved values 1. Use esc in the main menu to switch (if not yet displayed in the main menu) 2. Select datalogger menu 3. Select display saved values menu 4. View measurements: ok key switches the displayed measurements, and keys view permanent 5. Use esc to exit the menu again 10....

  • Page 19

    - 19 - 11.1 software installation pc system requirements: windows xp or higher, monitor resolution min. 1024 x 600, one free usb port 1. Insert cd rom in a pc-cd rom drive 2. Start start.Exe 3. Confirm welcome message with ok 4. Start installation with pc button 5. Confirm kemper program group with ...

  • Page 20

    - 20 - 9. Install usb driver with yes 10. In the status window, the software indicates in red letters that no connection has been established to the -master-. 11.2 connecting hand-held measuring instrument to pc usb cable hand-held measuring instrument scope of delivery.

  • Page 21

    - 21 - 11.3 general system settings the following general system settings can be made in the middle box. 1. Language selection: select the language for the pc software desktop 2. Time and date: apply from pc to the controller, automatic daylight savings time / standard time switchover enabled or ina...

  • Page 22

    - 22 - 11.4 read and save measurement formatted data from the hand-held measuring instrument 1. Select units for test record 2. Read the measurements from the device 3. Wait until all data have been completely read. Acknowledge message 5. Save measurements in csv file 2 3 1 4 5. Csv file open with e...

  • Page 23

    - 23 - 11.5 test record in the csv format test record measuring point 2.04 sensor 1384g dn20 measurement range =>c pipe dn 19.6 in mm glycol proportion 0 in % measurement no. Date time flowrate speed pt1000 [l/min] [m/sec] [°c] 1 1 15.12.2009 09:39:06 10,08 0,56 56 test record measuring point 2.04 s...