KENDALL FILAC 3000 AD Operating Manual - 3000
10
Electronic Thermometer
3000
AD/ADA
X. TroubleshooTinG
1. If device fails to function properly after washing & reconnecting probe, rinse all contacts in de-ionized water and blow dry.
2. Icons indicate all other possible errors and remedies.
3. Device Error Codes:
E01 - System error during synchronization
E02 - System error during patient thermistor calibration
E03 - System error during heater thermistor calibration
E04 - System timing error
E05 - Heater error
E06 - LCD communication error
E07 - Shut off error
P01 - Probe configuration (or no probe connected) error
P02 - Direct Mode patient thermistor unstable or out of range
P03 - Direct Mode heater thermistor unstable or out of range
P04 - Predict Mode patient thermistor unstable or out of range
P05 - Predict Mode heater thermistor unstable or out of range
P06 - Unable to pre-heat probe tip
4. No user-serviceable parts inside.
5. Refer to a factory or qualified Biomedical Technician for service.
Xi. sPeCifiCaTions
UL LISTING
Medical Equipment
FILAC 3000 electronic thermometer
47DA
Classified with respect to electrical shock, fire, and mechanical hazards only in accordance with UL60601-1
Classified with respect to electrical shock, fire, and mechanical hazards only in accordance with
CAN/CSA C22.2 NO.601.1
Dimensions (Approximate, without base):
in.- 3.2 W x 6.0 H x 2.1 D ; mm - 84 W x 152 H x 56 D
Dimensions (Approximate, with base):
in.- 4.1 W x 7.5 H x 3.1 D ; mm - 105 W x 190 H x 79 D
Weight (Approximate, without base):
13.5 Ounces ; 385 Grams
Temperature Measurement Range:
30°C to 43°C; 86°F to 109°F
Ambient Operating Temperature Range:
10°C to 40°C; 50°F to 104°F, at 10% - 95% RH, non-condensing.
Average prediction times
(after insertion into measurement site):
Quick mode: less than 4 seconds
Standard mode: less than 6 seconds
Typical times for switch to Direct Mode:
If no measurement site detected:
60 seconds (after pulling probe from probe well)
If temperature does not stabilize:
70 seconds (after measurement site insertion)
Pulse Timer:
60 Second count with a “beep” at 15 seconds, 2 “beeps” at 30 seconds, 1 “beep” at 45 seconds
and 2 “beeps” at 60 seconds
Water Bath Accuracy (35.5°C to 42.0°C):
Direct Mode:
±0.1°C
Standard Prediction Mode**:
±0.1°C
Quick Prediction Mode (Oral Only)**:
±0.3°C
**Greater than 95% of the prediction mode readings will be within the specified accuracy.