  • Page 1

    Kendall genius ™ 2 infrared tympanic electronic thermometer operation and service manual thermomètre électronique infrarouge auriculaire – manuel d’utilisation et d’entretien ’’ elektronisches infrarot-ohrthermometer – gebrauchs- und wartungsanweisung termometro elettronico timpanico a infrarossi – ...

  • Page 2: Table Of Contents

    Table of contents page section i — introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1 initial set up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ....

  • Page 3

    Genius ™ 2 1 infrared tympanic electronic thermometer section i — introduction this operator’s manual was prepared for the operator of the genius 2 infrared tympanic electronic thermometer. This manual contains instructions for use, precautions, and available maintenance and service information. To ...

  • Page 4

    Genius ™ 2 2 infrared tympanic electronic thermometer • always install a new probe cover prior to taking a temperature. The probe cover membrane should be smooth with no holes, tears, or wrinkles. • the thermometer was designed to be used only with genius 2 thermometer probe covers. Use of probe cov...

  • Page 5

    Genius ™ 2 3 infrared tympanic electronic thermometer the genius 2 infrared tympanic electronic thermometer is a reliable and accurate temperature-taking device. The reason for the accuracy of the genius 2 thermometer is the design, controlled calibration methods and stringent manufacturing contro...

  • Page 6

    Genius ™ 2 4 infrared tympanic electronic thermometer the following equivalence modes are available on the genius 2 infrared tympanic electronic thermometer. Data is available from tyco healthcare/kendall on request. Ear: in ear (ear) mode, the display will indicate the absolute temperature without ...

  • Page 7

    Genius ™ 2 5 infrared tympanic electronic thermometer 5. Place the probe in the ear canal and seal the opening with the probe tip. For consistent results, ensure that the probe shaft is aligned with the ear canal. 6. Press and release the scan button. 7. Remove the probe from the ear as soon as the ...

  • Page 8

    Genius ™ 2 6 infrared tympanic electronic thermometer pulse timer mode 1. Press and hold the timer button to enter timer mode. Press again to start the timer. The timer will run from 0 to 60 seconds. 2. The thermometer will issue a single beep at 15 seconds, a dual beep at 30 seconds, a triple beep ...

  • Page 9

    Genius ™ 2 7 infrared tympanic electronic thermometer if the display shows a system error 4 and the system error icon, the ambient conditions are changing too rapidly for the device to be used. Allow the device to equilibrate for 20 minutes before using. If the display shows any other system error, ...

  • Page 10

    Genius ™ 2 8 infrared tympanic electronic thermometer temperature mode site mode .

  • Page 11

    Genius ™ 2 9 infrared tympanic electronic thermometer site text pressing the °c/°f button when in this mode turns the body site text labels on or off. The labels will remain on when an “x” appears inside the box icon, and the text will remain off when the box is empty. section vi...

  • Page 12

    Genius ™ 2 10 infrared tympanic electronic thermometer section x — troubleshooting if the genius 2 infrared tympanic electronic thermometer is not functioning properly, check the following items: symptom action temperature reading unusually high check the probe cover for tears or gaps temperature re...

  • Page 13

    Genius ™ 2 11 infrared tympanic electronic thermometer displayed temperature measurement range: temperature range depends on site mode as follows: mode range °c range °f ear 33.0 to 42.0 91.4 to 107.6 oral 33.6 to 42.0 92.5 to 107.6 axillary 33.0 to 42.0 91.4 to 107.6 core 34.0 to 42.0 93.2 to 107.6...

  • Page 14

    Genius ™ 2 12 infrared tympanic electronic thermometer regulatory and safety standards: the genius 2 thermometer follows international regulatory standards • en 12470-5:2003 (e) clinical thermometers – part 5: performance of infrared ear thermometers where sections 6.3.4 and 6.5.5 were met based on ...

  • Page 15: Section Xiii — Warranty

    Genius ™ 2 13 infrared tympanic electronic thermometer parts listing to order repair parts, please contact your local customer service center or sales representative for the parts listed below description order part number genius 2 thermometer with base 303000 genius 2 probe cover 303030 genius 2 fi...

  • Page 16

    Kendall norwegian genius ™ 2 english job no.................................................................................................................15125 artwork no............................................................................................... 182500-04 package type.............