KENDALL KANGAROO ePump Operation And Service Manual

Summary of KANGAROO ePump

  • Page 1

    Kendall kangaroo e pump ™ enteral feeding pump tyco/ healthcare enteral feeding pump — operation and service manual nutripompe — manuel d’utilisation et d’entretien pumpe für die enterale ernährung — betriebs- und serviceanleitung pompa di alimentazione enterale — manuale operativo e di manutenzione...

  • Page 2: Table Of Contents

    Thank you for purchasing the kangaroo e pump™ enteral feeding pump. With proper care, this device will provide you with years of precision service. Table of contents page section i — general information ..........................................................................................5 secti...

  • Page 3: Table Of Contents

    3 enteral feeding pump english english table of contents page continuous mode (adjust feed) .................................................................................................................16 continuous mode feeding .......................................................................

  • Page 4: Pump

    4 kangaroo r r e pump ™ table of contents page section ix — error/warning/informational screens and troubleshooting system error ..............................................................................................................................................29 hold error ..................

  • Page 5

    5 enteral feeding pump english english section i – general information the kangaroo epump enteral feeding pump is a simple-to-use, precision enteral feeding pump. It is easily programmed to provide patients with either continuous or intermittent feeding, and can provide automatic flushing capability...

  • Page 6: Pump

    6 kangaroo r r e pump ™ english english figure 1a. Kangaroo epump features, front and rear views. ...

  • Page 7

    7 enteral feeding pump english english section ii — safety and warnings caution: u.S. Federal law restricts the sale of this device to physicians or to their direct representatives. 1. Read this booklet thoroughly before using the kangaroo epump enteral feeding pump. 2. Do not use this device near f...

  • Page 8: Pump

    8 kangaroo r r e pump ™ english english 11. See icon descriptions in section iii-icon identification and section x - specifications and symbols for additional safety information. 12. This device is designed to minimize the effects of uncontrolled electromagnetic interference and other types of inter...

  • Page 9: Battery Setup

    9 enteral feeding pump english english error indicator pump operations are stopped until the error is resolved. Information indicator this indicates information only, and does not require any immediate action. Section iv — initial setup attaching the a/c adapter power cord plug the power cord into t...

  • Page 10: Pump

    10 kangaroo r r e pump ™ english english attaching pole clamp the kangaroo epump enteral feeding pump can be attached to a vertical pole using the pole clamp, included with the pump. To attach the pole clamp to the kangaroo epump enteral feeding pump, align the hole on the pole clamp with the mounti...

  • Page 11: Quick Start

    11 enteral feeding pump english english section v — instructions for use quick start 1. Press power on lower, right-hand corner of control panel. 2. Open blue door enclosing pump set loading area. 3. To load pump set (refer to figure 4): • grasp finger tab on valve and insert into left pocket (1). •...

  • Page 12: Pump

    12 kangaroo r r e pump ™ english english general startup placement/mounting there are two recommended placement methods for the kangaroo epump enteral feeding pump. • attached to a vertical iv pole via the pole clamp peripheral device, included with the pump. • placed feet down on any horizontal sur...

  • Page 13: Loading Pump Sets

    13 enteral feeding pump english english loading pump sets the kangaroo epump enteral feeding pump will indicate load a set in the screen’s status line if a pump set is not load a set load a set installed. A blinking icon of a pump set will also appear in the upper left corner of the screen. Figure 5...

  • Page 14: Pump

    14 kangaroo r r e pump ™ english english the display should read set loaded and will show one of the following opening menus, depending on the type of set loaded set loaded pump set detected, and whether continuous or intermittent feeding mode has been selected. Prime pump the kangaroo epump enteral...

  • Page 15

    15 enteral feeding pump english english auto priming the auto prime option will be available if the pump senses that the currently installed pump set was not previously primed, and that there is no fluid in the line near the rotor. If auto-priming is desired, press and release4auto prime to automati...

  • Page 16: Pump

    16 kangaroo r r e pump ™ english english continuous mode (adjust feed) continuous mode feeding for continuous feeding mode, the adjust feed option is available in the opening menus (figure 7a for feed- only sets or figure 7b for feed & flush sets). Select this option to program the feeding parameter...

  • Page 17

    17 enteral feeding pump english english figure 10a. Adjust feed menu. Adjust feed adjust feed figure 10b. Setting the feed rate. Figure 10c. Setting the feed volume to be delivered. Continuous mode flushing for continuous feeding mode with a feed & flush set loaded, the opening menu will appear as s...

  • Page 18: Pump

    18 kangaroo r r e pump ™ english english intermittent mode (adjust bolus) intermittent mode feeding for intermittent feeding mode, the adjust bolus option is available in the opening menus (figure 7c for a feed-only set or figure 7d for a feed & flush set). Select this option to program the feeding ...

  • Page 19

    19 enteral feeding pump english english figure 12a. Adjust bolus menu. Figure 12b. Setting the bolus rate. Figure 12c. Setting the number of boluses. Figure 12d. Setting the volume per bolus. Figure 12e. Setting the time interval between starts of bolus. Intermittent mode feeding – bolus max the bol...

  • Page 20: Pump

    20 kangaroo r r e pump ™ english english note: the nature of two intermittent processes, periodic bolus feeding and periodic flushing, is such that occasional conflicts can arise where each process needs to be running at the same time. The flushing process is assigned precedence over the feeding pro...

  • Page 21

    21 enteral feeding pump english english figure 14. Running mode screen for running running continuous feed mode. Hold pump operation can be paused by pressing4hold while it is running, such as in figure 14. The status line will indicate holding and the yellow led indicator will be lit up to give a q...

  • Page 22: Pump

    22 kangaroo r r e pump ™ english english clear volume fed press4clear volume to re-zero the output feed totals “ml fed,” “boluses fed,” and “ml flush.” adjust settings press 4adjust settings to return to one of the opening menus shown in figure 7, thus allowing any setting to be modified. If the pum...

  • Page 23: More Options

    23 enteral feeding pump english english figure 16. More options menu. Options options figure 17. Feed and flush totals for up to 72 hours of previous history, excluding data from the current hour. More options press4more in one of the opening menus, figures 5 or 7, to access the more options screen,...

  • Page 24: Pump

    24 kangaroo r r e pump ™ english english biotech screen data sn is the serial number of the pump. Sw shows the software version number in this pump. Flash shows the version number of the flash data installed in this pump. Cert intv indicates the number of times the pump has been powered on since the...

  • Page 25: Lock Screen Feature

    25 enteral feeding pump english english lock screen feature the kangaroo epump enteral feeding pump can be set to run and then locked out from unauthorized use. While the lock settings feature prevents only changes to the input parameters, the lock screen feature will lock out any button presses, in...

  • Page 26: Pump

    26 kangaroo r r e pump ™ english english load the re-certification pump set. The pump will automatically recognize the re-certification pump set and give brief instructions to initiate the self-diagnostic procedure. Follow the instructions and review the pass/fail information at the end of the test....

  • Page 27

    27 enteral feeding pump english english general cleaning directions cleaning of kangaroo epump enteral feeding pumps must be performed as follows: caution: disconnect pump from ac power source before cleaning. After cleaning, do not connect to ac power source until pump and power cord are thoroughly...

  • Page 28: Pump

    28 kangaroo r r e pump ™ english english preventative maintenance this pump must be periodically re-certified to assure proper functioning and safety. The recommended service interval is at least once per year. Re-certification may be done at the user’s biomedical engineering department, an outside ...

  • Page 29: System Error

    29 enteral feeding pump english english section ix — error/warning/informational screens and troubleshooting for operational errors, the kangaroo epump enteral feeding pump display will show the error indicator icon, with specific information regarding the error condition, and will sound the alarm. ...

  • Page 30: Pump

    30 kangaroo r r e pump ™ english english error # related sub-system 14 time out during downstream occlusion check, never got rotoron3=0. May blow fuse. 20 could not find stop cock flush position during prime routine. 21 could not find stop cock feed position during prime routine. 22 motor error duri...

  • Page 31: Rotor Error

    31 enteral feeding pump english english rotor error the rotor error screen appears during running or running running priming, when the pump has detected an unusual operating condition for the rotor. See figure 24. The rotor error generally results from a problem with the pump set tubing around the r...

  • Page 32: Pump

    32 kangaroo r r e pump ™ english english if the error still cannot be resolved, press power to stop operation of the pump and put a different pump into service. If the error still cannot be resolved, press if the error still cannot be resolved, press flow error the flow error screen appears when eit...

  • Page 33: Battery Low

    33 enteral feeding pump english english battery low the battery low screen appears and the alarm beeps continuously when the battery needs to be recharged. There is approximately 15 minutes of battery life remaining when this screen appears. Plug the power cord in to begin recharging. The pump will ...

  • Page 34: Pump

    34 kangaroo r r e pump ™ english english section x — specifications and symbols specifications ul listing type infusion device volumetric pumping mechanism rotary peristaltic pump sets kangaroo epump mistic feed-only set or feed & flush set feeding formula delivery rate 1-400 ml/hr in 1 ml increment...

  • Page 35

    35 enteral feeding pump english english flushing solution dose range 10-500 ml in 1 ml increments flushing solution interval range 1-24 hr in 1 hr increments accuracy ±10% or 0.5 ml/hr, whichever is larger, with bag at 46 cm (18”) above pump, at room temperature 22° c ± 2° c (72° f ± 3° f), using a ...

  • Page 36: Pump

    36 kangaroo r r e pump ™ english english symbols on pump and accessories the following symbols are found on the pump or on the accesories, such as the a/c adapter. Refer to accompanying documents type bf protection (degree of protection against electrical shock - there is no conductive connection to...

  • Page 37: Section Xii — Maintenance

    37 enteral feeding pump english english section xi — customer service the circuitry of the kangaroo epump enteral feeding pump is not customer serviceable. In particular, electronic assembly rework by non-authorized kangaroo epump enteral feeding pump technicians will likely affect accuracy. Certain...

  • Page 38: Pump

    38 kangaroo r r e pump ™ english english battery pack see section viii for full instructions on replacing battery pack. During long-term non-usage or storage (greater than 9 months) of the pump, battery discharge and deactivation may tend to occur. It is recommended that the battery be disconnected ...

  • Page 39

    Section xv – electromagnetic conformity declaration the kangaroo epump enteral feeding pump has been built and tested according to ul60601-1, e181931 csa-c22.2 no.601.1-m90, and en60601-1-2 standards. The kangaroo epump enteral feeding pump is intended for use in the electromagnetic environment spec...

  • Page 40: Pump

    40 kangaroo r r e pump ™ english section xvi – accusystem√ the kangaroo epump is a reliable and accurate enteral feeding device. The reason for the accuracy of the kangaroo epump enteral feeding pump is the design, controlled calibration methods, and stringent manufacturing controls. One of the most...

  • Page 41: Performance Certifi Cation

    ™ trademark of tyco healthcare group lp or its affi liate. Tyco healthcare group lp • mansfield, ma 02048 • made in u.S.A. • © 2005 tyco healthcare group lp • all rights reserved • 050310 • product info in u.S. 1-800-962-9888 • www.Tycohealthcare.Com ec rep: tyco healthcare u.K. Ltd • gosport, po13 0a...

  • Page 42

    Kendall tm kendall kangaroo e pump ™ enteral feeding pump tyco / healthcare enteral feeding pump — operation and service manual nutripompe — manuel d’utilisation et d’entretien pumpe für die enterale ernährung — betriebs- und serviceanleitung pompa di alimentazione enterale — manuale operativo e di ...