    Digital dual display ac/dc clamp meter model- km 2782 / 2782-t off hz v w cap °c 200a 1000a 400a 1000a auto power off dh>2sec ~ ~ ~ select range rel hzduty m v auto -004.0 com t- t+vw hz cat-ii max1000v max 750v ac true rms dc/ac clamp ~ ! Km 2782 g 17,bharat industrial estate, t. J. Road, sewree (w...

  • Page 2: Table of Contents

    Nearly every electrical engineer has a hand held digital clamp meter (tongtester). We sometimes take them for granted, until we damage them or “burn them out”. If you incorrectly connect your clamp meter to a circuit, or if you have the clamp meter on wrong setting, you damage the meter and possibly...

  • Page 3

    1. Safety information the following safety information must be observed to insure maximum personal safety during the operation of this meter. 1. When measuring voltage ensure that instrument is not switched to the current range, resistance range, diode and continuity range, capacitance range of temp...

  • Page 4

    2. General specifications ? Display : 3¾ digit lcd with a max. Reading of 3999 (4000 counts) ? Auto / manual range selection ? Polarity : automatic negative polarity indication ? Zero adjustment : automatic ? Over range indication : “ol” is displayed. ? Low battery : the “ ” is display when the batt...

  • Page 5

    05 ac current (km 2782) 400a 0~800 800~1000 1000a ±(3.0%rdg + 10dgts) accuracy range 0.1a 1a resolution ±(3.5%rdg + 10dgts) ±(6.5%rdg + 10dgts) average sensing, calibrated to rms of sine wave overload protection : 1000arms within 60 seconds frequency : 50~400hz 2.2.5 ac voltage (true rms)(km 2782-t)...

  • Page 6

    07 temperature (k-type) range diode and audible continuity test range description resolution 1°c 1°c 1°c accuracy* -20~150°c 150~300°c 300~1000°c overload protection : 250v dc / 250vrms ac * thermocouple sensor accuracy not included. 1)temperature probe supplied can measure upto 250°c display read a...

  • Page 7

    09 5. Hz / duty key : in hz range, push the key, you can measure the duty, push again, go back to hz measurement. In voltage range, push it, you can measure hz and duty, but the measurement range will be smaller, and the auto range mode will change to manual range mode. 6. Rel key: press the key, th...

  • Page 8

    11 1) connect the black test lead to “com” socket and red test lead to the “vÙ” socket. Ac voltage measurement 3) read the result from the lcd panel. 2) set the selector switch to desired “v~” position and connect the probes across the source or load under measurement off hz v w cap °c 200a 1000a 40...

  • Page 9

    13 correct load off hz v w cap °c 200a 1000a 400a 1000a auto power off dh>2sec ~ ~ ~ select range rel hzduty v 1000 com t- t+vw hz cat-ii max1000v max 750v ac true rms dc/ac clamp ~ ! ~ + - + -ve incorrect load off hz v w cap °c 200a 1000a 400a 1000a auto power off dh>2sec ~ ~ ~ select range rel hzd...

  • Page 10

    15 load off hz v w cap °c 200a 1000a 400a 1000a auto power off dh>2sec ~ ~ ~ select range rel hzduty v 1000 com t- t+vw hz cat-ii max1000v max 750v ac true rms dc/ac clamp ~ ! ~ ac 3phase source incorrect resistance measurement 1) connect the black test lead to “com” socket and red test lead to the ...

  • Page 11

    17 caution : a) capacitors should be discharged before being tested. B) this device adopts charging mode to measure capacitance, so when testing large capacitance, it will take longer time before the final indication, and larger the capacitor, the longer the time. C) when testing small capacitance, ...

  • Page 12

    19 temperature measurement 4) read the result from the lcd panel. A. The temperature function shows random number when temperature probe is not connected. Insert the thermocouple in temperature terminal while examining temperature. B. The temperature probe supplied along with the meter can measure t...

  • Page 13

    21 relative measurement press “rel” button you can measure the relative value and “rel” appears on the display, the auto range mode will change to manual range mode. Press again to return to the previous condition & “rel” disappears, but meter will you can not go back to auto range mode. In “hz” mea...

  • Page 14: Test Certificate

    23 4.2 battery replacement 1) ensure the instrument is not connected to any external circuit. Set the selector switch to “off” position and remove the test leads from the terminals. 2) open the cover of the battery cabinet using a screwdriver. 3) replace the old batteries with the same type new batt...

  • Page 15

    Warranty each “kusam-meco” product is warranted to be free from defects in material and workmanship under normal use & service. The warranty period is one year (12 months) and begins from the date of despatch of goods. In case any defect occurs in functioning of the instrument, under proper use, wit...