32
Chapter 5: Using Your Filtered Enclosure and Appropriate Application
•
Arrange for minimal disruptions, such as room traffic or
entry into the room while the enclosure is in use.
Start-up
•
Turn on exhaust system and accessory light. Turn off the
UV light, if equipped.
•
Only raise the sash for loading and cleaning.
•
Check the baffle air slots for obstructions.
•
Allow the enclosure to operate unobstructed for 1 minute.
•
Wear a long sleeved lab coat and rubber gloves. Use
protective eyewear. Wear a protective mask if appropriate.
Consult your Safety Officer for additional personal
protective equipment recommendations.
Loading Materials and Equipment
•
Load only the materials required for the procedure. Do not
overload the enclosure.
•
Do not obstruct the air foil, or rear baffle slots.
•
Large objects should not be placed close together and
should be elevated above the work surface to permit
airflow to sweep under the equipment.
•
After loading, wait one minute to purge airborne
contaminants from the work area.
Work Techniques
•
Keep all materials inside the lower air foil, and perform all
contaminated operations as far to the rear of the work area
as possible.
•
Segregate all clean and contaminated materials in the work
area.
•
Avoid using techniques or procedures that disrupt the
airflow patterns of the enclosure.
Final Purging
•
Upon completion of work, the enclosure should be allowed
to operate for two to three minutes undisturbed, to purge
airborne contaminants from the work area before shutting
down the blower.
•
If equipped, turn on the UV light to decontaminate the
work area as determined by your Safety Officer.