Labomed Lx 500 User Manual - Assembly
12
This filter modifies the color of observation light into a natural (daylight
color).
• Fit the filter (1) into the bottom of the condenser (2) until it clicks into
place. (Figure 11).
1
2
Installing or Replacing the Lamp Bulb
Mounting the Daylight (Blue) Filter
Each standard set can be assembled by simply attaching the bulb, filter and power cord.
Fig. 10
ASSEMBLY
9
Lx 500
Before attaching the lamp bulb, remove the parts that may drop such as the
filter and specimen from the microscope frame, and place the microscope
on its back so that the bottom plate is exposed.
1. Pull out the lamp house cover (1) to detach it.
2. Pull up the lamp from the lamp holder & replace with lamp.
3. Replace the lamp house cover.
Always use the designated bulb. Using a bulb other than those
specified by LABOMED may lead to a fire hazard and decentering of
light. Fingerprints or stains on the lamp bulb reduce its life. If
contamination occurs, wipe bulb surface with a cloth slightly
moistened with alcohol.
Caution: For Bulb Replacement During Use or Right After Use
The bulb, lamp socket and areas near these will be extremely hot during and right after use. Set the main
switch to" O" (OFF), disconnect the power cord from the wall outlet, and allow the bulb and lamp socket to
cool before replacing the bulb with a new bulb of the designated type. Cooling time may vary to users
discretion.