LabRepCo LABH?5?URB Owner's Manual

Manual is about: PRIME UNDER COUNTER BI SERIES REFRIGERATORS

Summary of LABH?5?URB

  • Page 1

    Prime under counter bi series refrigerators labh ‐ 5 ‐ urb, labh ‐ 5 ‐ urbss owner’s instructions this manual describes how to operate and care for your appliance to get the best, most efficient, performance. Read this book! Note to customer: this merchandise was carefully packed and thoroughly insp...

  • Page 2: Table

    2 table of contents receiving .......................................................................................... 3 installation .................................................................................... 4 temperature control ................ Error! Bookmark not defined. Refrigerato...

  • Page 3

    3 receiving your unit was built, packaged, and inspected with extreme care. We shipped it to you using carriers we trust with a proven track record of careful handling, good customer service, and on time delivery. Unfortunately, regardless of all of these efforts sometimes accidents happen and occas...

  • Page 4

    4 installation uncrating – move your refrigerator as close to the final location as possible before unpacking. Remove the carton and the styrofoam strips. The location should be as close as possible to the power outlet so no extension cord is needed. This unit requires a minimum of 3 inches on the s...

  • Page 5

    5 appliance, allow the cabinet to operate a minimum of four hours or until it has started cycling normally before placing product in the cabinet. The motor compressor may start and stop several times when the unit is first started or after defrosting, especially if the weather is very hot. This is o...

  • Page 6

    6 refrigerator operation after the unit is properly installed and power is applied, it will take some time before the system is cooled down to temperature and cycling normally. You should wait 8 hours on the first startup before beginning to add product to the unit. This ensures that the unit is ins...

  • Page 7

    7 temptech ‐ 2000 electronic temperature control description and functions the t ‐ 1000 electronic temperature control is a digital microprocessor based temperature control, indication, and alarm system. It continuously monitors the temperature of the refrigerator/freezer through its sensor assembly...

  • Page 8

    8 programming mode note: the temptech ‐ 2000 programming mode gives access to several functions that are not intended for user adjustment. The procedures below provide direction to the functions that you can adjust. Please do not attempt to change any of the other functions in the programming area o...

  • Page 9

    9 adjusting other user adjustable settings: ‐ to change between celsius and fahrenheit scales, press set and down keys for 3 seconds to enter programming mode. “hy” will display. Press the up or down keys several times to cycle through the parameters until “cf” is displayed. Press set to display the...

  • Page 10

    10 quick troubleshooting guide check these items before calling for service problem: possible cause / soultions: unit does not run electrical circuit is not 110 ‐ 120v 60hz. The power cord is not plugged in. No power at electrical outlet. Check to make sure breaker is not tripped or fuse is not blow...

  • Page 11

    11 problem: possible cause / soultions: moisture collects inside door gasket is not sealing properly. Check for debris, cracks, and items passing through door at the gasket. The refrigerator or freezer is facing a doorway or is underneath of air conditioning vent. Relocate the unit or redirect air v...

  • Page 12

    12 maintenance and cleaning these units require very little maintenance. They should be cleaned periodically to keep them running efficiently. Use the cleaning agents and suggestions below for best results. If ¼ to ½ inch of frost accumulates around the condenser tubes or if the unit will not mainta...

  • Page 13

    13 treat your water ‐ though this is not always practical, softening hard water can do much to reduce deposits. There are certain filters that can be installed to remove corrosive elements. If you are not sure of the proper water treatment, call a treatment specialist. Keep your equipment clean ‐ us...

  • Page 14

    14 defrosting it is important to defrost and clean the unit when ¼ to ½ inch of frost has accumulated around the condenser tubes or when the unit will not maintain proper temperature. When defrosting becomes necessary, press the def key for 2 seconds. The compressor will stop. Remove the product and...

  • Page 15

    15 product warranty information horizon scientific, inc. Product warranty horizon scientific, inc. Warrants to the original purchaser every new horizon scientific, inc. Refrigerated unit, the cabinet and all parts thereof, to be free from defects in material or workmanship, when such unit is install...

  • Page 16

    16 horizon scientific, inc.’s maximum cumulative liability relative to all claims and liabilities, including obligations under any indemnity, whether or not insured, shall not exceed the cost of the product(s) giving rise to the claim or liability. Warranty is not transferable. This warranty is not ...