LabRepCo LABL-27-HG Manual

Summary of LABL-27-HG

  • Page 1

    Important information for this cooler has passed the quality control inspection and meets high standards. This inspection includes complete refrigeration system, cabinet construction & finish. Important please retain for your records labl-27-hg, labl-38-sg, labl-45-sg, labl-49-hg, and labl-74-hg 101...

  • Page 2

    Installation and operating instructions 1. Installation a. Receiving inspection: upon receipt, check all packages and accessories or optional components including legs, casters, and shelves. B. Legs & casters: (see fig. 1) legs screw directly into the base. Casters are screwed directly into the base...

  • Page 3

    2. Operation a. Electrical supply and connections: plug all standard models into 115 volts a.C. 60 cycle outlet. Low line voltage in often the cause of service complaints. Check to see that the line voltage is 110 volts or more with the unit running. Other motors or heavy appliances should not be us...

  • Page 6

    Shelf support location illustration ilb-1621 all mt's rear front 1/2" tab 1/4" tab 1/2" tab 1/4" tab shelf support - rear with 1/2" tab ( full scale ) shelf support - front with 1/4" tab ( full scale ) -- notes & instructions -- 1. Place shelf supports with 1/4" tabs into pilaster at desired positio...

  • Page 7

    Bi-pin fluorescent lamp installation /verification prior to applying power to the unit, verify that all lamps are properly installed and fully engaged in the lamp holders. Proper lamp engagement is achieved by rotating the lamp 90° from its insertion position until it "snaps" or "clicks" into place....

  • Page 9

    Methods for cleaning stainless steel method of cleaning agent* application** effect on finish routine cleaning soap, ammonia or sponge with cloth, then satisfactory for use on detergent and water. Rinse with clear water and all finishes. Wipe dry. Stubborn spots and stains, revere ware cleaner, appl...

  • Page 10: Lab

    Service and analysis refrigeration chart system malfunction possible cause solution compressor will not start - no hum 1. Line cord not plugged in. 1. Plug in the cord. 2. Fuse removed or blown. 2. Replace fuse. 3. Overload protector tripped. 3. Refer to electrical section. 4. Temp control stuck in ...