  • Page 2

    5.00 • rm626728.

  • Page 3: Table of Contents

    Table of contents general information packaging and delivery . . . . . . . . . . . . . . . . . . . . . . . . . . 4 general precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 safety precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 limits of liability . . . . . . ...

  • Page 4: General Information

    General information packaging & delivery experienced packaging personnel at lamotte company assure adequate protection against normal hazards encountered in transportation of shipments. After the product leaves the manufacturer, all responsibility for safe delivery is assured by the transportation c...

  • Page 5: Specifications

    Specifications instrument type single wavelength, direct reading colorimeter measurement wavelengths: 420 nm, 460 nm, 510 nm, 530 nm, 570 nm or 605 nm readable resolution: determined by test factor photometric precision: ±0.001 absorbance unit range: 0 - 2.00 absorbance units display: 3 ½ digits res...

  • Page 6: Parts & Accessories

    Parts & accessories included in the model 1200 colorimeter kit : 1200 colorimeter colorimeter tubes, set of 6 water sample collecting bottle reagent system optional accessories ac adapter, 110v/60hz code 1773 ac adapter, 220v/50hz code 1774 colorimeter tubes, set of 6 code 0290-6 6.

  • Page 7: Epa Compliance

    Epa compliance the 1200 chlorine colorimeter is an epa accepted instrument. Epa accepted means that the instrument meets requirements for colorimeters as found in test procedures that are approved for the national primary drinking water regulations (npdwr) or national pollutant discharge elimination...

  • Page 8: Chemical Testing

    Chemical testing water sampling for chemical analysis taking representative samples the underlying factor to be considered for any type of water sampling is whether or not the sample is truly representative of the source. To properly collect a representative sample: • sample as frequently as possibl...

  • Page 9: Sampling of Closed System

    For dissolved oxygen studies, or for tests requiring small sample sizes, a water sample bottle (lamotte code 1060) will serve as a subsurface or in-depth sampler. This weighted device is lowered to the sampling depth and allowed to rest at this depth for a few minutes. The water percolates into the ...

  • Page 10: An Introduction to

    An introduction to colorimetric analysis most test substances in water are colorless and undetectable to the human eye. To test for their presence we must find a way to “see” them. The lamotte colorimeter can be used to measure any test substance that is itself colored or can be reacted to produce a...

  • Page 11: Reagent Blank

    The 1200 colorimeter has been specially calibrated to read test results directly in parts per million (ppm) of the test factor. The sensitivity of the response has been maximized by using a specific, complementary light source for each test. Reagent blank some tests will provide greater accuracy if ...

  • Page 12: Sample Dilution Techniques

    Sample dilution techniques & volumetric measurements if a test result using the 1200 colorimeter gives an er2 (over range) message, the sample must be diluted. Then the test should be repeated to obtain a reading which is in the concentration range for the test. (note: this is not true for colorimet...

  • Page 13: Interferences

    Interferences lamotte reagent systems are designed to minimize most common interferences. Each individual test discusses interferences unique to that test. Be aware of possible interferences in the water being tested. The reagent systems also contain buffers to adjust the water sample to the ideal p...

  • Page 14: Overview

    General operating information overview the model 1200 colorimeter is a portable, microprocessor controlled, direct reading colorimeter. The microprocessor enables factory programmed calibrations to optimally match non-linear curves. The auto zero feature eliminates the need to dial in the zero manua...

  • Page 15

    15.

  • Page 16: The Keypad and Display

    The keypad and display 16.

  • Page 17: Testing Tips

    Testing tips 17.

  • Page 18: Calibration Procedure

    Calibration procedure the 1200 has been pre-calibratede. Recalibration of the 1200 by the user is not required. However, a procedure to standardize the calibration (shown below) should be performed to obtain the most accurate readings. 18.

  • Page 19

    U the 1200 can be calibrated with two prepared standards. The concentrations of the standards should be chosen from the low and high ends of the range of the meter. The low standard must be less than 5% of over range. To calibrate the meter with two standards, follow steps 1-11 of the calibration pr...

  • Page 20: Analysis Procedure

    Analysis procedure 20

  • Page 21

    21.

  • Page 22: Available Test Kits

    Available test kits test factor order code/model range (ppm) detectio n limit test method (# of reagents) ammonia nitrogen 3680-01 dc 1200-nh 0-5.00 0.05 nessler (2) bromine 3672-01 dc 1200-br 0-7.00 0.05 dpd tablets (1) chlorine (free & total) 3670-01 dc1200-cl 0-4.00 0.05 dpd tablets (2) chlorine ...

  • Page 23: Troubleshooting

    Troubleshooting problem check action meter won’t turn on battery replace ac adapter plug in ac wall outlet verify power source contact lamotte for return authorization return to lamotte for repair suspect calibration check calibration with standards use new standards verify standard concentration ru...

  • Page 24: Rs232 Port

    Rs232 port the 1200 colorimeter may be interfaced with any ibm compatible computer using an interface cable (code 27359-04). The meter may also be interfaced with an rs-232 serial printer, using an appropriate cable and setting the printer configuration to the output below. Output: rs232 compatible,...

  • Page 25: Maintenance

    Maintenance replacing the battery the lamotte 1200 uses a standard 9-volt alkaline battery that is available worldwide. The battery compartment is located on the bottom of the case. To replace the battery: 1. Open the battery compartment lid. 2. Remove the battery and disconnect the battery from the...