LaMotte 1742 PockeTester Instruction Manual

Manual is about: ORP Tracer

Summary of 1742 PockeTester

  • Page 1

    Tracer pocketester ä orp code 1742.

  • Page 3: Tracer

    Tracer orp pocketester tm code 1742 table of contents introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 parts & accessories ....

  • Page 4: Introduction

    Introduction congratulations on your purchase of the tracer orp pocketester. The advantages of the tracer to the user include: easy to use and maintain, high accuracy, automatic calibration, and fast response. The tracer a 15-reading memory storage. Careful use and maintenance will provide years of ...

  • Page 5: Meter Description

    Meter description front panel description 1. Battery compartment cap 2. Lcd display 3. Mode/hold button 4. Cal/recall button 5. On/off button 6. Electrode collar 7. Electrode (note: the electrode cap is not shown) display description 1 . Reading 2 . Units 3 . Bargraph 4 . Data hold 5 . Scale 6 . Low...

  • Page 6: Basic Operation

    Basic operation powering the tracer if the batteries are weak, the bat indicator will appear on the lcd. Press the on/off key to turn the tracer on or off. The auto power off feature will shut the tracer off automatically after 10 minutes of inactivity. Electrode recognition when the tracer is turne...

  • Page 7: Button Immediately After

    Orp measurement 1. Press the on/off button to turn the meter on. Place the electrode ½” to 1” into the test sample . 2. Stir once and let the reading stabilize for 1 to 3 minutes. The main display will indicate the orp in numeric units from 0 to 1000mv. The bar graph will also display the orp value....

  • Page 8: Maintenance

    Maintenance storage to preserve electrode life keep the sponge in the protective cap soaked with tap water. Dot not use deionized or distilled water.Cap tracer when not in use. Store vertically. Battery replacement 1. Twist off the battery compartment cap. 2. Hold the battery housing in place with o...

  • Page 9: Troubleshooting

    Troubleshooting problem cause action power on but no display batteries insert batteries batteries verify correct polarity batteries replace batteries unstable readings electrode immerse electrode more deeply in sample electrode condition electrode before first use. Soak in tap water for 10 minutes. ...

  • Page 10: Expand Your Tracer

    Expand your tracer interchangeable electrodes are available to convert the orp tracer to a total chlorine tracer or a ph tracer. Remember to ask for instructions and appropriate reagent or buffer tablets when ordering total chlorine or ph electrodes. Total chlorine tracer electrode, 0.00-10.00 ppm c...

  • Page 11

    Lamotte company will service and repair out-of-warranty products at a nominal charge..

  • Page 12: Lamotte Company

    Lamotte company helping people solve analytical challenges ® po box 329 • chestertown • maryland • 21620 • usa 800-344-3100 • 410-778-3100 (outside u.S.A.) • fax 410-778-6394 visit us on the web at www.Lamotte.Com 5/08.