LaMotte 1749 Manual

Manual is about: EC/TDS/SAL Tracer PockeTester Kit. Conductivity/TDS/Salinity/Temperature/Meter

Summary of 1749

  • Page 1

    Tracer conductivity/tds/salinity/temperature/meter ec/tds/sal code 1749 pocketester tm new! Direct reading tds and salt!.

  • Page 2

    Warning! This set contains chemicals that may be harmful if misused. Read cautions on individual containers carefully. Not to be used by children except under adult supervision.

  • Page 3: Tracer

    Tracer ec/tds/sal pocketester tm • code 1749 table of contents introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 parts & access...

  • Page 4: Introduction

    Introduction congratulations on your purchase of the ec/tds/sal (electro conductivity/total dissolved solids/salinity) tracer pocketester. The tracer is a revolutionary, first of its kind measurement device that offers direct reading of conductivity, tds and salinity with one electrode. Careful use ...

  • Page 5: Contents

    Contents ec/tds/sal tracer pocketester kit code 1749 includes: conductivity standard, 12,880 µs code 6317-g sample cup w/cap † † not sold individually. See below. Parts & accessories ec/tds/sal replacement electrode code 1765 weighted stand w/sample cups (5) code 1746 sample cups w/caps (24) code 17...

  • Page 6: Meter Description

    Meter description front panel description 1. Battery compartment cap 2. Lcd display 3. Mode/hold button - change mode, hold data, store data 4. Cal/recall button - calibration, change temperature units, recall data 5. On/off button 6. Electrode collar 7. Electrode (note: the electrode cap is not sho...

  • Page 7: Basic Operation

    Basic operation powering the tracer the tracer uses four cr2032 lithium ion batteries. If the batteries are weak, the bat indicator will appear on the display. Press the on/off key to turn the tracer on or off. The auto power off feature will shut the tracer off automatically after 10 minutes after ...

  • Page 8

    Changing the displayed temperature units to change the displayed temperature units between °f or °c: 1. With the tracer off, press and hold the cal/recall button. 2. With the cal/recall button pressed, momentarily press the on/off button. When self cal appears in the display, release the cal/recall ...

  • Page 9: Testing

    Testing getting started 1. Remove the cap from the bottom of the tracer to expose the electrode. 2. Before the first use, rinse the electrode in deionized water and dry. 3. For best results, calibrate for conductivity with a standard in the expected range of the sample. For maximum accuracy calibrat...

  • Page 10

    Measurement 1. Fill a sample cup to the 20 ml line with the test sample. Sample depth must be greater than or equal to 1.5 inches. 2. Immerse the tracer electrode in the sample. Make sure the electrode is completely submersed. 3. Press the on/off button. (8888 and then self cal will appear in the di...

  • Page 11

    7. Press the on/off button to turn the tracer on. Make sure the meter is in the tds mode. 8. Immerse the electrode in the filtrate. Make sure the tip of the electrode is completely immersed. 9. Stir the filtrate with the electrode to create a homogeneous solution. 10. Gently stir the filtrate with t...

  • Page 12: Calibration

    Clearing the stored memory turn the tracer on. Press and hold the on/off button for 4 second the display will briefly desplay “clr” when the memory is cleared. Calibration for the most accurate results, allow sufficient time for the temperature of the probe to reach the temperature of the sample bef...

  • Page 13

    Note: each time the calibration mode is entered all calibration range indicators will be cleared, but only the calibration data for the currently selected range will be replaced. The calibrations for the other two ranges will be saved even though the indicators for those ranges are no longer display...

  • Page 14: Operational Matrix

    Operational matrix function/resulting action power mode key press sequence on/off on or off any momentary press of on/off button calibration on con press & hold cal/recall button for 2 seconds until cal is displayed store reading on any momentary press of mode/hold button hold release on hold moment...

  • Page 15: Function/resulting

    Function/resulting action power mode key press sequence enter con/tds ratio on tds (ppm or mg/l) press and release the cal/recall button twice in quick succession. Change con/tds ratio on tds (ppm or mg/l) press and release the cal/recall button twice in quick succession. Momentary press of mode/hol...

  • Page 16: Maintenance

    Maintenance storage 1. Rinse the electrode in distilled or deionized water. 2. Store the electrode dry with the cap on. 3. Always rinse the electrode in deionized water between measurements to avoid cross contamination. Double rinsing is recommended when high accuracy is required. Battery replacemen...

  • Page 17: Contaminant

    Electrode cleaning recommendations do not soak the electrode in the solutions for longer than the recommended length of time. To do so may cause a reference potential shift which will cause a degradation in performance or failure. When cleaning the probe, do not scratch or damage the platinized elec...

  • Page 18: Troubleshooting

    Troubleshooting problem check action reading is frozen hold mode press mode/hold button to exit hold mode “bat” message batteries low replace batteries meter will not calibrate in conductivity mode trapped air bubbles tap probe or stir sample to release air bubbles dirty probe clean conductivity pro...

  • Page 19: Warranty

    Warranty lamotte company warrants this instrument to be free of defects in parts and workmanship for 1 year from the date of shipment and the probe to be free of defects in parts and workmanship for 6 months from the date of shipment. If it should become necessary to return the instrument for servic...

  • Page 20: Lamotte Company