LeCroy ZD200 Instruction Manual

Manual is about: Differential Probe

Summary of ZD200

  • Page 1

    Instruction m anual zd200 differential probe.

  • Page 2: Zd200 Differential

    Zd200 differential probe instruction manual march, 2011.

  • Page 3: Lecroy Corporation

    Lecroy corporation 700 chestnut ridge road chestnut ridge, ny, 10977-6499 tel: (845) 578-6020, fax: (845) 578 5985 warranty lecroy warrants this oscilloscope accessory for normal use and operation within specification for a period of one year from the date of shipment. Spare parts, replacement parts...

  • Page 4: Zd200 Differential Probe

    Zd200 differential probe zd200-gsm-reva iv table of contents introduction .......................................................................................... 6 key benefits ....................................................................................... 6 standard accessories ............

  • Page 5: Zd200 Differential Probe

    Zd200 differential probe zd200-gsm-reva v verification procedure...................................................................... 20 a. Lf attenuation accuracy .................................................................. 20 reference material .................................................

  • Page 6: Zd200 Differential Probe

    Zd200 differential probe zd200-gsm-reva 6 introduction the zd200 differential probe is ideally suited for automotive and serial data signals. The wide dynamic range (+/- 20v differential) and 1 mohm input resistance make the zd200 ideally suited for a wide range of applications. The zd200 probe can ...

  • Page 7: Instruction Manual

    Instruction manual zd200-gsm-reva 7 standard accessories the zd200 probe is shipped with the following standard accessories: standard accessory quantity straight tip 6 hook clip 2 y leadset 1 micro ic clips 2 micro-grabbers 2 extension leads (5 cm) 2 extension leads (10 cm) 2 dual pin set (12.8 mm) ...

  • Page 8: Zd200 Differential Probe

    Zd200 differential probe zd200-gsm-reva 8 features and accessories the zd200 probe is provided with numerous features and accessories to make probing and connecting to different test points easier than ever. The small, low mass probe head is designed for ease of use and high performance. The probe t...

  • Page 9: Instruction Manual

    Instruction manual zd200-gsm-reva 9 num part description 1 hook clips 2 micro-grabbers the micro- and mini-grabbers are ideal for connecting to small ic legs or pins very tightly spaced. 3 micro ic grabbers the micro- and mini-grabbers are ideal for connecting to small ic legs or pins very tightly s...

  • Page 10: Zd200 Differential Probe

    Zd200 differential probe zd200-gsm-reva 10 probe operation handling the probe the zd200 probe is a precision test instrument. Exercise care when handling and storing the probe. Always handle the probe by the probe body or compensation box. Avoid putting excessive strain or exposing the probe cable t...

  • Page 11: Instruction Manual

    Instruction manual zd200-gsm-reva 11 control through the oscilloscope’s interface can be found on the channel dialog that corresponds with the connected probe. Refer to your oscilloscope’s manual for specific operation instructions. Turning the volts/div knob controls the oscilloscope’s scale factor...

  • Page 12: Zd200 Differential Probe

    Zd200 differential probe zd200-gsm-reva 12 care and maintenance cleaning the exterior of the probe and cable should be cleaned, using a soft cloth moistened with water. The use of abrasive agents, strong detergents, or other solvents may damage the probe. Always ensure that the input leads are free ...

  • Page 13: Instruction Manual

    Instruction manual zd200-gsm-reva 13 returning a probe for calibration or service return a probe for calibration or service by contacting your local lecroy sales representative. They tell you where to return the product. All returned products should be identified by both model and serial number. Pro...

  • Page 14: Zd200 Differential Probe

    Zd200 differential probe zd200-gsm-reva 14 description of failure 5. Package the probe case in a cardboard shipping box with adequate padding to avoid damage in transit. 6. Mark the outside of the box with the shipping address given to you by the lecroy representative; be sure to add the following: ...

  • Page 15: Instruction Manual

    Instruction manual zd200-gsm-reva 15 replacement parts the probe connection accessories and other common parts can be ordered through the north america customer care centers. Replacement part part number accessory kit pacc-zd007 y-lead adapter pacc-zd008.

  • Page 16: Zd200 Differential Probe

    Zd200 differential probe zd200-gsm-reva 16 performance verification performance verification overview this procedure can be used to verify the warranted characteristics of the zd200 high impedance active probe. The recommended calibration interval for the model zd200 is one year. The complete perfor...

  • Page 17: Instruction Manual

    Instruction manual zd200-gsm-reva 17 because the input and output connector types may vary on different brands and models of test instruments, additional adapters or cables may be required. Description minimum requirement test equipment examples digital oscilloscope probus interface; windows-based w...

  • Page 18: Zd200 Differential Probe

    Zd200 differential probe zd200-gsm-reva 18 description minimum requirement test equipment examples cable terminator, precision, bnc 50 Ω ± 0.05% lecroy term-cf01 banana plug adapter (2 ea.) female bnc to dual banana plug pomona 1269 bnc to mini- grabber bnc mail to mini- grabber cable, 36” pomona 51...

  • Page 19: Instruction Manual

    Instruction manual zd200-gsm-reva 19 6. Disconnect the probus extender cable from the oscilloscope. Verify that the scale factor changes from 20 mv/div to 2 mv/div. 7. Reconnect the probus extender cable to the oscilloscope. The warranted characteristics of the zd200 are valid at any temperature wit...

  • Page 20: Zd200 Differential Probe

    Zd200 differential probe zd200-gsm-reva 20 functional check the functional check will verify the basic operation of the probe functions. It is recommended that the functional check be performed prior to the performance verification procedure. 1. Return to the factory default settings by: a. Selectin...

  • Page 21: Instruction Manual

    Instruction manual zd200-gsm-reva 21 9. Connect the probus-cf01 bnc male output (the probe end) to the precision 50Ω termination bnc input. The probe should remain powered. 10. Install the straight tips on the zd200 inputs. 11. Connect the signal generator output to the zd200 inputs using the bnc to...

  • Page 22: Zd200 Differential Probe

    Zd200 differential probe zd200-gsm-reva 22 reference material specifications note: specifications are subject to change without notice. Please refer to the lecroy website at www.Lecroy.Com for detailed specification information. Contact lecroy for support use the following regional contacts to find ...

  • Page 23: Instruction Manual

    Instruction manual zd200-gsm-reva 23 contact your local lecroy office for sales and technical assistance email (sales and support): contact.Sa@lecroy.Com web site: www.Lecroy.Com/europe china phone (sales and support): ++86 28- 86527180 / 7181 / 7182 fax (sales and support): +86 28- 8652 7183 email ...

  • Page 24: Zd200 Differential Probe

    Zd200 differential probe zd200-gsm-reva 24 safety symbols the following symbols appear on the accessory or in this manual and indicate important safety considerations. Refer to information near this symbol to protect against personal injury or damage to the instrument. The caution sign indicates a p...

  • Page 25: Instruction Manual

    Instruction manual zd200-gsm-reva 25 operating environment the accessory is intended for indoor use and should be operated in a clean, dry environment. Before using this product, ensure that its operating environment is maintained within these parameters: temperature: 5 to 40 °c. Humidity: maximum r...

  • Page 26: Zd200 Differential Probe

    Zd200 differential probe zd200-gsm-reva 26 safety requirements caution avoid personal injury or damage to your accessory or any equipment connected to it by reviewing and complying with the following safety precautions: use only as intended. The accessory is intended to be used only with the compati...

  • Page 27: Instruction Manual

    Instruction manual zd200-gsm-reva 27 compliance information this lecroy accessory is ce compliant and bears the ce mark. Ce declaration of conformity this zd200 probe meets intent of the european council directive 2006/95/eec for product safety and 2004/108/eec for electromagnetic compatibility. Com...

  • Page 28: Zd200 Differential Probe

    Zd200 differential probe zd200-gsm-reva 28 o en 61000-4-2:2001 electrostatic discharge immunity (±4 kv contact discharge, ±4 kv air discharge, ±4 kv vertical/horizontal coupling planes ) o en 61000-4-3:2006 rf radiated electromagnetic field immunity (3 v/m, 80 mhz to 1 ghz; 3 v/m, 1.4 ghz to 2 ghz; ...

  • Page 29: Instruction Manual

    Instruction manual zd200-gsm-reva 29 appendix a - performance verification test record this record can be used to record the results of measurements made during the performance verification of the zd200 probe. Photocopy this page and record the results on the copy. File the completed record as requi...

  • Page 30: Zd200 Differential Probe

    Zd200 differential probe zd200-gsm-reva 30 equipment used instrument model serial number calibration due date oscilloscope digital multimeter function generator note: the function generator used in this performance verification procedure is used for making relative measurements. The output of the ge...

  • Page 31: Instruction Manual

    Instruction manual zd200-gsm-reva 31 thank you for purchasing a zd200 differential probe..