M-system 20VS8-202N Instruction Manual

Summary of 20VS8-202N

  • Page 1

    20vs8-202n p. 1 / 4 em-4444 5-2-55, minamitsumori, nishinari-ku, osaka 557-0063 japan phone: +81(6)6659-8201 fax: +81(6)6659-8510 e-mail: info@m-system.Co.Jp isolation amplifier (small size, input isolation, dip type) model 20vs8-202n instruction manual before use .... Thank you for choosing m-syste...

  • Page 2

    20vs8-202n p. 2 / 4 em-4444 5-2-55, minamitsumori, nishinari-ku, osaka 557-0063 japan phone: +81(6)6659-8201 fax: +81(6)6659-8510 e-mail: info@m-system.Co.Jp ■ connection diagram 3 38 37 20 22 36 1 2 19 18 input v in + feedback com (–) non-inverting input inverting input reference volt. Source for i...

  • Page 3

    20vs8-202n p. 3 / 4 em-4444 5-2-55, minamitsumori, nishinari-ku, osaka 557-0063 japan phone: +81(6)6659-8201 fax: +81(6)6659-8510 e-mail: info@m-system.Co.Jp foraninvertingamplification,theout- put is following. V o = -(r 1 / r 2 ) × v in when r 1 =20kΩ,r 2 =10kΩ,thenthe outputisfollowing.Thegainis-...

  • Page 4

    20vs8-202n p. 4 / 4 em-4444 5-2-55, minamitsumori, nishinari-ku, osaka 557-0063 japan phone: +81(6)6659-8201 fax: +81(6)6659-8510 e-mail: info@m-system.Co.Jp maintenance regular calibration procedure is explained below: ■ calibration warmuptheunitforatleast10minutes.Apply0%,25%, 50%, 75% and 100% in...