    Instruction manual smartread ™ digital blood pressure monitor guidebook model 04-244-001 please read this instruction manual completely before operating this unit. Mabis . Thcare inc heal ™.

  • Page 2: Limited Five Year Warranty

    Limited five year warranty your mabis blood pressure monitor is guaranteed to be free of manufacturing defects for a period of five years from the date of purchase under normal use. If the unit fails to operate during warranty period, return it prepaid along with $5.00 for return shipping and insura...

  • Page 3: Index

    Introduction . . . . . . . . . . . . . . . . . . . . . . 2 product identification & caution . . . . . . . 3 general blood pressure information . . . . 4-6 important information before use . . . . . . 7 battery installation/replacement . . . . . . 8 ac adapter installation (optional) . . . . . 9 appl...

  • Page 4: Introduction

    To achieve the maximum benefit from your blood pressure monitor, we recommend that you first consult with your physician or trained health care professional. Thank you for purchasing a mabis automatic digital blood pressure monitor. With proper care and use, your monitor will provide you with many y...

  • Page 5: Product Identification

    Product identification 3 cuff power/start button digital display panel cuff jack ac jack cuff tubing tubing connector ring caution you can stop the inflation or deflation process of the cuff at any time by pressing the “power” button causing the cuff to immediately deflate. Batteries “aa” (alkaline ...

  • Page 6: What Is Blood Pressure?

    What is blood pressure? Blood pressure is the pressure that is exerted by blood flowing against the walls of the blood vessels throughout your body. Your heart, which is the center of the circulatory system, provides the force for the blood to flow or circulate. When your heart contracts or beats, t...

  • Page 7

    What influences blood pressure? Many factors such as genetics, age, sex, altitude, physical activity, anxiety, muscular development, certain medications or even the time of day can influence blood pressure. Influences such as sleep or relaxation decreases blood pressure, while anxiety or exercise in...

  • Page 8: Variations In Blood Pressure

    6 who blood pressure classifications standards for assessment of high or low blood pressure, regardless of age, have been established by the world health organization (who) as shown in the following chart. Reference material: investigation into adult diseases report by the ministry of health and soc...

  • Page 9

    7 1. Blood pressure measurements should only be interpreted by a physician or a trained health care professional who is familiar with your medical history. Through regular use of this device and recording of your measurements, you can keep your physician informed of the changes in your blood pressur...

  • Page 10: Note:

    8 battery installation/replacement low battery indicator 1. Remove the battery cover as shown in fig. 1. 2. Insert the batteries into the compartment, fig. 2, taking care to observe and match the polarities as indicated on battery compartment. 3. Replace the battery cover, fig. 3. Note: the batterie...

  • Page 11: Note:

    9 ac adapter installation (optional) fig. 1 your mabis blood pressure monitor will operate on 4 “aa” batteries (included) or a standard 6v/600ma ac adapter fitted with a din plug (not included). To use an ac adapter: 1. Insert the din plug into the socket located on the left side of the monitor, fig...

  • Page 12

    Applying your blood pressure cuff 10 fig. 1 1. Remove any constrictive clothing or jewelry that may interfere with the cuff placement. 2. You should be seated with your feet flat on the floor. 3. Position the cuff on a solid surface with the tubing facing up and away from you, fig. 1. The metal bar ...

  • Page 13

    11 applying your blood pressure cuff 5. Insert your left arm into the cuff (cylinder). The bottom edge of the cuff should be positioned one inch above the elbow joint, fig. 3. The tubing should extend towards your hand. 6. With your left arm extended in front of you and palm facing upward, postion t...

  • Page 14

    12 taking your blood pressure reading 1. Position the monitor on a flat, stable surface with the digital display panel in view. Insert the cuff tubing into the left side of the monitor, fig. A. 2. Rest your elbow on a table with your palm facing upward. Elevate your hand so that the cuff is at the s...

  • Page 15

    Taking your blood pressure reading 5. The display will show the increasing pressure in the cuff, fig. 4. 6. When a pulse is detected, the unit will sound for each beat and the display symbol will flash, fig. 5. 7. A long beep will indicate that the measurement process is complete and the cuff will a...

  • Page 16: Error Display Symbols

    Error display symbols display symbol condition/cause corrective action • cuff pressure drops to less than 20 mmhg before achieving reading. • movement is detected during reading or low inflation level. • cuff inflation takes too long. Wait a minimum of 15 minutes and remeasure. Loosen cuff to allow ...

  • Page 17: Care and Maintenance

    1. Only use a soft, dry cloth to clean your blood pressure monitor. 2. Avoid using any types of liquids on the monitor or cuff. 3. Do not store the unit where it will be exposed to direct sunlight, dust or humidity. Avoid extreme temperatures. 4. Never disassemble the monitor or cuff. 5. Dropping or...

  • Page 18: Condition

    16 if any abnormality occurs during use, please check and correct the following: condition correction display is blank check and correct the polarity when power is on. Of the installed batteries. Reinstall or replace batteries. Cuff pressure does not increase check and reconnect cuff after pressing ...

  • Page 19: Product Specifications

    17 name and model number smartread™ digital blood pressure monitor, 04-244-001 display system digital display system/lcd measuring method oscillometric method power source “aa” type alkaline batteries 1.5v x 4 measuring range 30 ⬃ 280 mmhg (pressure) 40 ⬃ 200 pulse/minute (pulse) accuracy within ⫾ 3...