Magnetrol eclipse 705 Safety Manual

Manual is about: Guided Wave Radar Level Transmitter

Summary of eclipse 705

  • Page 1

    Guided wave radar level transmitter sil certified safety manual for enhanced model 705-51ax-xxx software v3.X this manual complements and is intended to be used with the eclipse ® enhanced model 705 installation and operating manual (bulletin 57-600 dated august 2005 or later). Safety function the h...

  • Page 2: Eclipse

    57-651 eclipse ® sil certified safety manual for enhanced model 705-51ax-xxx table of contents 1.0 introduction ...................................................................3 1.1 theory of operation.................................................3 1.2 product description ........................

  • Page 3

    3 57-651 eclipse ® sil certified safety manual for enhanced model 705-51ax-xxx 1.0 introduction 1.1 theory of operation guided wave radar is based upon the principle of tdr (time domain reflectometry). Tdr utilizes high frequency pulses of electromagnetic energy transmitted down a probe. When a puls...

  • Page 4

    4 57-651 eclipse ® sil certified safety manual for enhanced model 705-51ax-xxx 3. All products chosen for use in the sif must meet the requirements of iec 61508 for the given sil capability level or be justified based on proven in use data. Eclipse enhanced model 705 meets the hardware architec- tur...

  • Page 5

    5 57-651 eclipse ® sil certified safety manual for enhanced model 705-51ax-xxx 2.2 miscellaneous electrical considerations the following are miscellaneous electrical items that must be considered in any installation: 2.2.1 pollution degree 2 the enhanced eclipse model 705 is designed for use in cate...

  • Page 6

    6 57-651 eclipse ® sil certified safety manual for enhanced model 705-51ax-xxx 5.0 instructions 5.1 systematic limitations the following must be observed to avoid systematic failures: 5.1.1 application as the probe configuration establishes the fundamental per- formance characteristics of the system...

  • Page 7

    7 57-651 eclipse ® sil certified safety manual for enhanced model 705-51ax-xxx 5.4 storage the device should be stored in its original shipping box and not be subjected to temperatures outside the storage temper- ature (-50 to +80 °c) shown in the eclipse enhanced model 705 installation and operatin...

  • Page 8

    5.6.2 sis configuration requirements ensure the gwr transmitter parameters have been properly configured for the specific application and probe. Special consideration should be given to the following con- figuration parameters: fault: do not choose hold for this parameter as a fault will not be annu...

  • Page 9

    9 57-651 eclipse ® sil certified safety manual for enhanced model 705-51ax-xxx refer to section 2.6.4 of the enhanced eclipse model 705 installation and operating manual 57-600 for additional information regarding password protection. Ensure that an exclamation mark (!) appears as the last char- act...

  • Page 10

    10 57-651 eclipse ® sil certified safety manual for enhanced model 705-51ax-xxx 5.8.2 troubleshooting refer to section 3.3 of the enhanced eclipse model 705 i & o manual 57-600 for troubleshooting the various device status messages, which can be in the form of warnings and faults. To assist in findi...

  • Page 11

    11 57-651 eclipse ® sil certified safety manual for enhanced model 705-51ax-xxx 6.1.4 proof test procedure a suggested proof test is described below. This test will detect approximately 94% of possible du failures in model 705-51a*-*** version of the enhanced eclipse model 705. Ensure that all neces...

  • Page 12

    12 57-651 eclipse ® sil certified safety manual for enhanced model 705-51ax-xxx 7.0 report: lifetime of critical components although a constant failure rate is assumed by the proba- bilistic estimation method, this only applies if the useful lifetime of components is not exceeded. Beyond the useful ...

  • Page 13

    13 57-651 eclipse ® sil certified safety manual for enhanced model 705-51ax-xxx identifier assumptions for safety allocated ssa 1 the user sif will detect and properly handle annunciation of detected fault conditions signaled by the alarm level output according to the specific requirements of the si...

  • Page 14

    14 57-651 eclipse ® sil certified safety manual for enhanced model 705-51ax-xxx 8.2 fmeda report: exida management summary.

  • Page 15

    15 57-651 eclipse ® sil certified safety manual for enhanced model 705-51ax-xxx.

  • Page 16

    Bulletin: 57-651.4 effective: january 2018 supersedes: march 2012 705 enterprise street • aurora, illinois 60504-8149 • 630-969-4000 info@magnetrol.Com • www.Magnetrol.Com copyright © 2018 magnetrol international, incorporated service policy owners of magnetrol controls may request the return of a c...