Appendix 1
Guidance and manufacture’s declaration – electromagnetic immunity –
for all EQUIPMENT and SYSTEMS
Guidance and manufacture’s declaration – electromagnetic immunity
The CMS50QA Pulse Oximeter is intended for use in the electromagnetic environment specified below. The customer or
the user of CMS50QA Pulse Oximeter should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±
6 kV contact
±
8 kV air
±
6 kV contact
±
8 kV air
Floors should be wood, concrete
or ceramic tile. If floor are
covered with synthetic material,
the relative humidity should be at
least 30%.
Power frequency
(50/60Hz) magnetic
field
IEC61000-4-8
3A/m
3A/m
Power frequency magnetic fields
Should be at levels characteristic
of a typical location in a typical
commercial
or
hospital
environment.
NOTE U
T
is the a.c. mains voltage prior to application of the test level.