MediSure Blood glucose monitoring system User Manual - Indications For Use

Summary of Blood glucose monitoring system

  • Page 1

    (multi-user model) medi + sure canada inc. 31a-328 speedvale ave. East, suite 160, guelph , ontario, n1e 0j4, canada. Www.Medisure.Ca toll free: 1-855-mdi-sure (634-7873) fax: 855-329-5302 e-mail: info@medisure.Ca user guide ® ® medi sure blood glucose monitoring system medi sure ®.

  • Page 2: Medi

    Medi + sure ® multi-patient user guide - table of contents introduction ....................................................................................................... 1 testing principle .................................................................................................. 2 ind...

  • Page 3: Introduction

    1 introduction we would like to thank you for making a smart decision. The medi + sure® blood glucose monitoring system is designed for canadians. The compact and portable medi + sure® blood glucose meter is easy to use, and displays test results in only 6 seconds. In addition, the meter automatical...

  • Page 4: Testing Principle

    2 testing principle the medi + sure® blood glucose monitoring system is designed to provide an easy, accurate method for determining capillary blood glucose values. This analysis is based on amperometric technology using glucose oxidase that is specific for the blood glucose measurement..

  • Page 5: Indications For Use

    3 indications for use the medi + sure® blood glucose monitoring system is intended for use in the quantitative measurement of glucose in fresh capillary whole blood from the finger. It is to be used as an aid in monitoring the effectiveness of a diabetes control program. It is not intended for the d...

  • Page 6: The Medi

    4 important safety instructions users must adhere to standard precautions when handling or using this device. All parts of the glucose monitoring system should be considered potentially infectious and are capable of transmitting blood-borne pathogens between patients and healthcare professionals. Th...

  • Page 7: Kit Contents

    5 contents and specifications kit contents medi + sure® multi-patient blood glucose meter 3-volt lithium coin battery (cr2032) user's guide getting started guide carrying case meter specifications storage conditions temperature: -20°~+50°c humidity rh ≦ 80% memory capacity 960 test results with resp...

  • Page 8: System Operating Conditions

    6 system operating conditions test strips use only medi+sure blood glucose test strips operating conditions temperature: 10°~40° c humidity: rh ≦ 80% sample type capillary whole blood in finger hematocrit of tested blood 25%~60% altitude limit to be used no higher than 10183 feet above sea level the...

  • Page 9: About The Medi

    7 about the medi + sure® system medi + sure® lcd display show blood glucose test results and messages medi + sure® meter back.

  • Page 10: Medi

    8 medi + sure® blood glucose test strip how to determine whether blood sample is sufficient note: if the sample is sufficient, the sampling window on the test strip should be full with no white space showing..

  • Page 11: How Your System Works

    9 how your system works installing or replacing the battery your medi + sure® blood glucose monitoring system is packed with one 3-volt lithium coin battery that needs to be installed before testing. The battery will last for approximately 1,000 tests. Always keep spare batteries on hand. When the b...

  • Page 12: Alarm Setting

    10 10. Press the button. The hour flashes. 11. Press the button or button until the correct hour is displayed. 12. Press the button. The minutes flash. 13. Press the button or button until the correct minutes are displayed. 14. Press the button. The unit “mmol/l” is displayed. 15. Press the button. ...

  • Page 13: The Medi

    11 reviewing results in memory the medi + sure® blood glucose meter automatically stores up to 960 test results with date and time. You can review test results from the newest to the oldest. Results stored in memory will not be lost when you replace the battery. 1. Press the button to turn on the me...

  • Page 14

    12 14. To delete all the results in the memory, press the button for at least 3 seconds. “----” message will be shown on the screen. This indicates that all results in the memory have been deleted. If you don’t want to clear delete all test results, scroll to the individual result and press the butt...

  • Page 15: Alarm Notes

    13 3. Press the button and the temperature, and thermometer symbol will be shown on the screen for 3 seconds and the meter will turn off automatically. Note: to quickly adjust a certain number, keep pressing the or button until the one you want appears. At any steps if you want to shut the meter off...

  • Page 16: Blood Sampling And Testing

    14 blood sampling and testing finger tip blood sampling 1. Wash hands in warm, soapy water. Rinse well and dry thoroughly. 2. Press a single use lancing device firmly onto the chosen site and draw a sufficient blood sample. 3. Dispose of lancet device in a suitable sharps container. Notes • only an ...

  • Page 17

    15 automatically. You will hear a beep letting you know the countdown has begun. 7. Look at the sampling window on the test strip. If you have enough blood on the strip, you will not see any white. The sampling window will be completely filled with blood. If it doesn’t have enough blood, it will loo...

  • Page 18: Testing Control Samples

    16 testing control samples prepare the test strip and glucose control solutions. 1. Press the button. The meter will power on. 2. Insert medi + sure® strip into the strip port. 3. Press the button to enter control mode. The button “l1” refers to level 1. You can select l1, l2 or blood by pressing th...

  • Page 19: Step 1:

    17 how to clean and disinfect your device step 1: clean before disinfecting 1. Dispense caviwipes towelette. 2. Clean surface of your device with towelette to remove debris and bioburden. 3. Discard used towelette. Step 2 : disinfect after cleaning 1. Dispense caviwipes towelette. 2. Disinfect pre-c...

  • Page 20: Whole Case

    18 4. If the meter is being operated by a second person who is providing testing assistance to the user, the meter should be disinfected prior to use by the second user. 5. Wash hands thoroughly with soap and water after handling the meter, or test strips. Why is the disinfection of this device requ...

  • Page 21: Sure® Customer Assistance

    19 information of caviwipe disinfectant caviwipes disinfecting towelettes are purchased from metrex research corp, and the epa reg. Is 46781-8. Caviwipes are non-woven disposable towelette pre-saturated with cavicide, and are intended for use in health care settings such as hospital, dental offices,...

  • Page 22: Important Notes

    20 important notes caring for your meter taking care of your meter is important. Please follow the guidelines shown below: 1. Do not disassemble your meter. If you have any questions, please contact medi + sure®. 2. Please use your meter carefully. Avoid throwing, shaking, or dropping it, which may ...

  • Page 23: Sample Application Message:

    21 no memory in the meter message: if there is no memory in the meter, you press button two times and the meter shows “----”. Sample application message: this requires you to apply the blood sample or control solution to the test strip after inserting a test strip. Apply blood sample apply control s...

  • Page 24: Hi Message:

    22 hi message: if the test result is higher than 41.7 mmol /l the “hi” message is displayed. Lo message: if the test result is lower than 1.1 mmol /l the “lo” message is displayed..

  • Page 25: Cause:

    23 error messages and troubleshooting when you operate the meter incorrectly, error messages are displayed. Each message is defined below. Cause: test strip was removed during testing. Damaged or used test strip was used. The blood sample was applied before inserting the test strip. The blood sample...

  • Page 26: Cause:

    24 cause: the system is operating outside the appropriate temperature range 10°~40°c. Action: move your meter and test strips to a place where the temperature is appropriate for running a test. Wait at least 30 minutes and test again. Cause: the battery is low. Action: replace the battery soon..

  • Page 27: Expected Values

    25 expected values 1 expected blood glucose levels for people without diabetes: time range (mmol/l) fasting two hours after meal 1 american diabetes association: diabetes care, volume 35, supplement 1, january 2012, s4.

  • Page 28: Meter Warranty

    26 meter warranty medi+sure canada inc. Warrants, to the original purchasers of the medi + sure® blood glucose meter, that your meter shall be free of defects in materials and workmanships under a life time warranty from the date of purchase. If the medi + sure® meter does not work properly because ...

  • Page 29: Medi

    27 medi + sure® log book date time test result ac / pc notes/patient name ac pc ac pc ac pc ac pc ac pc ac pc ac pc ac pc ac pc ac pc.

  • Page 30

    28 notes:.

  • Page 31: Name

    29 personal information name address home phone mobile phone hospital physician email diagnosis date additional information:.