Medtronic 24965 Technical Manual

Manual is about: Patient Connector Reveal LINQ Mobile Manager

Summary of 24965

  • Page 1

    24965 patient connector draft technical manual caution: federal law (usa) restricts this device to sale by or on the order of a physician..

  • Page 2

    The following list includes trademarks or registered trademarks of medtronic in the united states and possibly in other countries. All other trademarks are the property of their respective owners. Carelink, medtronic.

  • Page 3

    Contents 1 introduction to the 24965 patient connector . . . . . . . . . . . . . . 4 1.1 explanation of packaging and product symbols . . . . . . . . . . . 4 1.2 description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 1.3 intended use . . . . . . . . . . . . . . . . . . . . . ....

  • Page 4

    1 introduction to the 24965 patient connector 1.1 explanation of packaging and product symbols refer to the package label and product to see which symbols apply to this product. Caution consult instructions for use ingress protection use only with specified power supply type bf applied part humidity...

  • Page 5

    Accessories china rohs bluetooth connection follow instructions for use (blue) low battery keep dry acma (australian communications and media authority and the new zealand ministry of economic development radio spectrum management standards) symbol for australia and new zealand for us audiences only...

  • Page 6

    1.3 intended use the patient connector is a portable electronic device using low frequency inductive telemetry to communicate with medtronic implanted heart devices. The patient connector uses bluetooth technology to transmit implanted heart device data to a medtronic mobile app for further processi...

  • Page 7

    Use of unapproved power supply – use only the medtronic-supplied power supply with the patient connector. Use of an unapproved power supply may damage equipment or impact user or patient safety. 1.6 precautions attaching the tether kit – do not overtighten the screw when attaching the tether kit. Au...

  • Page 8

    Normal use. However, electronic devices are susceptible to many environmental stresses including, but not limited to, the following examples. • the unit is designed to be used indoors in a clinic or hospital. • the unit should not be dropped or mishandled in such a manner as to cause physical damage...

  • Page 9

    1.9.2 what to do if you suspect the patient connector has been compromised discontinue use of the patient connector if it does not behave as expected. Contact your medtronic representative for further assistance. 1.10 compatible components the following compatible components are available for the pa...

  • Page 10

    Patient connector – provides the communication link between the app and the patient’s implanted device. The patient connector contains a radio-frequency (rf) transmitter and receiver. It must be held over the implanted device during a communication session. Mobile device (not supplied) – when runnin...

  • Page 11

    2.1.1 indicator lights figure 2. 1 telemetry status 2 bluetooth connection status 3 battery status table 1. Table with title and rules above and below header icon indicator color description telemetry status green when lit, indicates that the patient connector is com- municating with the implanted d...

  • Page 12

    2.2.1 charging the patient connector battery 1. Open the power supply connector cover located on the bottom edge of the patient connector. 2. Connect power supply to the patient connector. 3. Plug the power cord into the ac power outlet (ac mains). Warning: use only the medtronic-supplied power supp...

  • Page 13

    4. Install the app. 5. Touch the app icon to open it. 6. Follow the instructions in the app to complete the installation. 2.2.3 pairing the patient connector with the app the patient connector uses a bluetooth connection to communicate with the mobile device. You will need to pair each patient conne...

  • Page 14

    3 conducting a patient session 3.1 position the patient connector for an implantable device, the patient connector should be held directly against the patient’s skin. The face of the patient connector must be parallel to and typically within 5 cm of the implantable device. Optimum position of the pa...

  • Page 15

    1. Disinfect the patient connector by using a 10% bleach and water solution and a sterile saturated gauze pad or sponge. 2. Wipe down all external surfaces and maintain a wet or damp exposure time of 15 minutes. 3. Allow to air dry approximately 5 minutes or until dry. 4.2 software updates software ...

  • Page 16

    Table 2. Patient connector specifications (continued) electrical safety requirements per iec 60601-1:2005, clause 8.7 touch leakage current ≤0.1 ma patient leakage current a.C. ≤0.1 ma patient auxiliary current d.C. ≤ 0.01 ma patient auxiliary current a.C. ≤0.1 ma iec 60529 degrees of protection pro...

  • Page 18

    Medtronic, inc. 710 medtronic parkway minneapolis, mn 55432 usa www.Medtronic.Com +1 763 514 4000 medtronic usa, inc. Toll-free in the usa (24-hour technical consultation for physicians and medical professionals) bradycardia: +1 800 505 4636 tachycardia: +1 800 723 4636 europe/middle east/africa med...