Medtronic CONTOUR User Manual

Manual is about: NEXT LINK Meter

Summary of CONTOUR

  • Page 1

    C ontour ® n ext link meter.

  • Page 2: Contour

    C ontour ® n ext link product overview contour ® n ext link for use only with: compatible with: paradigm® veo™ meter: contour® next test strips paradigm® real-time (x22).

  • Page 3: Meter Readings

    C ontour ® n ext link accuracy • industry standards have an acceptable error range of ±20% to lab results. 1 • bayer’s contour® next link meter gives even more accurate readings, ±10% to lab results. 2 1. International organization for standardization. Iso 15197:2003 section 7.4 requires results wit...

  • Page 4

    Contour® next link meter features.

  • Page 5

    Charging the meter the bayer meter has a permanent rechargeable battery. Before use, the meter needs to be fully charged. To charge the meter: 1) remove the cap with on it to reveal the usb connector. 2) plug the usb connector into a computer. The computer must be on and not in sleep, hibernate or p...

  • Page 6

    Connecting the pump and meter (1/5) follow these steps to connect your meter to your pump. 1) hold the menu button until the meter turns on. 2) scroll to your language and press ok. Press ok to confirm. 3) press ok when asked connect to a minimed pump? 4) press ok. 5) stop. Do not press any buttons....

  • Page 7

    Connecting the veo pump and meter (2/5) 6) from the main menu of your pump, select utilities and press act 7) select connect devices and press act 8) from the device options screen, select meters and press act 9) select on and press act 10) the meter id menu will appear. Select add id and press act.

  • Page 8

    Connecting the pump and meter (3/5) 11) use the up and down arrow buttons to enter the meter id shown on the meter screen. Press act after entering each character. After entering the meter id number into the pump, press next on the meter 12) after you set the last character of the id, the screen wil...

  • Page 9

    Connecting the pump and meter (4/5) choosing meter send options 13) choose the send option. Always = meter will always send results. Ask me = meter will ask you before sending results. Never = meter will never send results..

  • Page 10

    Connecting the pump and meter (5/5) *for more detail on this feature, see your meter user guide . Additional meter settings autolog 15) accept or turn on the autolog feature*. Autolog allows you to mark a test result as before meal, after meal, or fasting target 16) accept or change the target range...

  • Page 11

    Auto calibrate feature in the veo should remain off calibration reminder.

  • Page 12

    C ontour ® n ext l ink useful tips test strips c ontour ® n ext l ink meter only uses c ontour n ext test trip port light to turn on the port light, double press the menu button when the meter is off underfill sampling if the blood drop is too small, the meter will beep twice and this screen will be...

  • Page 13

    Comparison to the previous contour link meter features contour® link contour® next link test strip contour® test strip contour® next test strip test time 5 seconds 5 seconds calibration no coding no coding sample size 0.6 ul 0.6 ul second-chance® sampling no yes memory 480 tests 1000 tests alternate...