Medtronic LIFEPAK 500 Operating Instructions Manual

Other manuals for LIFEPAK 500: Operating Instructions Manual
Manual is about: automated external defibrillator

Summary of LIFEPAK 500

  • Page 1

    Operating instructions lifepak ® 500 automated external defibrillator.

  • Page 3

    Lifepak ® 500 automated external defibrillator operating instructions.

  • Page 4

    Ii lifepak 500 automated external defibrillator operating instructions important federal (usa) law restricts this device to sale by or on the order of a physician. This automated external defibrillator (aed) is to be used by authorized personnel only. Device tracking the u.S. Food and drug administr...

  • Page 5

    Lifepak 500 automated external defibrillator operating instructions iii ©1996–2005 medtronic emergency response systems, inc. Table of contents preface about defibrillation ..................................................................................................................................

  • Page 6

    Iv lifepak 500 automated external defibrillator operating instructions transferring setup to another lifepak 500 aed...................................................................................2-13 connecting electrodes to the aed ..................................................................

  • Page 7

    Lifepak 500 automated external defibrillator operating instructions v ©1996–2005 medtronic emergency response systems, inc. List of figures figure 2-1 lifepak 500 aed controls, indicators, and connectors .................................................................2-2 figure 2-2 accessories for ...

  • Page 8

    Vi lifepak 500 automated external defibrillator operating instructions figure 5-5 sla battery capacity while installed in an aed for 3 months at 20°c (68°f) without recharging ...............................................................................................................................

  • Page 9

    Lifepak 500 automated external defibrillator operating instructions vii ©1996–2005 medtronic emergency response systems, inc. List of tables table 2-1 controls, indicators, and connectors.........................................................................................................2-2 tabl...

  • Page 10

    Viii lifepak 500 automated external defibrillator operating instructions table e-1 guidance and manufacturer’s declaration - electromagnetic emissions ....................................E-3 table e-2 guidance and manufacturer’s declaration - electromagnetic immunity....................................

  • Page 11

    Lifepak 500 automated external defibrillator operating instructions ix ©1996–2005 medtronic emergency response systems, inc. Pr ef ac e preface about defibrillation page x operator considerations x indications for use xi lifepak 500 automated external defibrillator xi features of the lifepak 500 aut...

  • Page 12

    X lifepak 500 automated external defibrillator operating instructions preface about defibrillation defibrillation is a recognized means of terminating certain potentially fatal arrhythmias. A direct current defibrillator applies a brief, high-energy pulse of electricity to the heart muscle. The life...

  • Page 13

    Lifepak 500 automated external defibrillator operating instructions xi ©1996–2005 medtronic emergency response systems, inc. Pref ac e preface indications for use the lifepak 500 aed is to be used only on patients in cardiopulmonary arrest. The patient must be unconscious, not breathing normally, an...

  • Page 14

    Xii lifepak 500 automated external defibrillator operating instructions preface automated operation the operator controls aed operation with two or three top-panel buttons ( on/off , analyze [optional], and shock ). For lifepak 500 aeds that do not have an analyze button, the aed operates in auto an...

  • Page 15

    Lifepak 500 automated external defibrillator operating instructions xiii ©1996–2005 medtronic emergency response systems, inc. Pref ac e preface readiness display most lifepak 500 aeds with the biphasic waveform include a readiness display on the device’s handle that can be seen at all times. Ok dis...

  • Page 17

    Lifepak 500 automated external defibrillator operating instructions 1-1 ©1996–2005 medtronic emergency response systems, inc. 1s af et y i nf or m at ion safety information this section provides important information to help you operate the lifepak 500 automated external defibrillator (aed). Familia...

  • Page 18

    1-2 lifepak 500 automated external defibrillator operating instructions safety information terms the following terms are used either in this manual or on the lifepak 500 aed: danger: immediate hazards that will result in serious personal injury or death. Warning: hazards or unsafe practices that cou...

  • Page 19

    Lifepak 500 automated external defibrillator operating instructions 1-3 ©1996–2005 medtronic emergency response systems, inc. Safety information 1s af et y i nf or m at ion symbols the symbols below may be found in this manual or on various configurations of the lifepak 500 aed and accessories: warn...

  • Page 20

    1-4 lifepak 500 automated external defibrillator operating instructions safety information indicator, steady display indicates battery is low, replace battery; flashing (key panel only) indicates replace battery immediately indicator, steady display indicates device requires service; flashing (key p...

  • Page 21

    Lifepak 500 automated external defibrillator operating instructions 1-5 ©1996–2005 medtronic emergency response systems, inc. Safety information 1s af et y i nf or m at ion mark of conformity according to the european medical device directive 93/42/eec canadian standards association certification fo...

  • Page 23

    Lifepak 500 automated external defibrillator operating instructions 2-1 ©1996–2005 medtronic emergency response systems, inc. 2g et tin g r ea dy getting ready this section provides a basic orientation to the lifepak 500 automated external defibrillator (aed) and describes how to prepare the aed for...

  • Page 24

    2-2 lifepak 500 automated external defibrillator operating instructions getting ready unpacking and initial inspection remove the lifepak 500 aed from the shipping container. Examine the aed and accessories for any sign of damage during shipping. Make sure that all the required supplies and accessor...

  • Page 25

    Lifepak 500 automated external defibrillator operating instructions 2-3 ©1996–2005 medtronic emergency response systems, inc. 2g et tin g r ea dy getting ready * accent marks are not included in operating messages for international languages. 3 orange shock button delivers energy. The led flashes to...

  • Page 26

    2-4 lifepak 500 automated external defibrillator operating instructions getting ready figure 2-2 accessories for the lifepak 500 aed table 2-2 accessories for the lifepak 500 aed 15 lifepak 500 nonrechargeable lithium battery pak provides power for the lifepak 500 aed. 16 lifepak 500 rechargeable sl...

  • Page 27

    Lifepak 500 automated external defibrillator operating instructions 2-5 ©1996–2005 medtronic emergency response systems, inc. 2g et tin g r ea dy getting ready about batteries use either of the following battery types to power the lifepak 500 aed: • lifepak 500 rechargeable sealed lead-acid (sla) ba...

  • Page 28

    2-6 lifepak 500 automated external defibrillator operating instructions getting ready battery removal to remove the battery: 1 turn off the aed. 2 lift the latch release on the battery and slide it out. Note: when a battery is removed from the aed, battery and service indicators appear on the readin...

  • Page 29

    Lifepak 500 automated external defibrillator operating instructions 2-7 ©1996–2005 medtronic emergency response systems, inc. 2g et tin g r ea dy getting ready 3 to set the hour: • press the s button to increase the value. • press the x button to advance to the next field. 4 to set the minutes: • pr...

  • Page 30

    2-8 lifepak 500 automated external defibrillator operating instructions getting ready modem selection the modem selection option determines the initialization string for the modems listed in table 2-4. Select the number that matches your modem. If you select 0, you must define the modem initializati...

  • Page 31

    Lifepak 500 automated external defibrillator operating instructions 2-9 ©1996–2005 medtronic emergency response systems, inc. 2g et tin g r ea dy getting ready for the lifepak 500 aed with the biphasic defibrillation waveform, energy level 1 is fixed at 200 joules; however, choices are available for...

  • Page 32

    2-10 lifepak 500 automated external defibrillator operating instructions getting ready note: when a no shock advised message occurs immediately after a shock, the cpr period is the same as cpr time 1. Note: cpr time 0 is not available if auto analyze 2 is selected on aeds that have an analyze button...

  • Page 33

    Lifepak 500 automated external defibrillator operating instructions 2-11 ©1996–2005 medtronic emergency response systems, inc. 2g et tin g r ea dy getting ready motion detection the motion detection option determines whether or not the motion detection system is active during analysis. The factory d...

  • Page 34

    2-12 lifepak 500 automated external defibrillator operating instructions getting ready transfer setup once the setup in one lifepak 500 aed has been customized, the transfer setup option supports the transfer of this setup to other lifepak 500 aeds. Setup transfers are possible only between lifepak ...

  • Page 35

    Lifepak 500 automated external defibrillator operating instructions 2-13 ©1996–2005 medtronic emergency response systems, inc. 2g et tin g r ea dy getting ready transferring setup to another lifepak 500 aed you can transfer the clock setting and all setup information except device id from one lifepa...

  • Page 36

    2-14 lifepak 500 automated external defibrillator operating instructions getting ready connecting electrodes to the aed you can connect the quik-combo electrodes with the redi-pak preconnect system to the aed before patient care to save time. To connect the redi-pak-type quik-combo electrodes: 1 ins...

  • Page 37

    Lifepak 500 automated external defibrillator operating instructions 3-1 ©1996–2005 medtronic emergency response systems, inc. 3 u sin g th e li fe pa k 50 0 ae d using the lifepak 500 aed this section describes how to use the lifepak 500 automated external defibrillator (aed) for ecg analysis and de...

  • Page 38

    3-2 lifepak 500 automated external defibrillator operating instructions using the lifepak 500 aed warnings and cautions preparing the aed for operation follow these steps to help ensure that the aed is always ready for use: • properly maintain the aed and batteries as described on page 5-7 of this m...

  • Page 39

    Lifepak 500 automated external defibrillator operating instructions 3-3 ©1996–2005 medtronic emergency response systems, inc. 3 u sin g th e li fe pa k 50 0 ae d using the lifepak 500 aed quik-combo and fast-patch electrodes are pre-gelled, self-adhesive electrodes that allow hands- free defibrillat...

  • Page 40

    3-4 lifepak 500 automated external defibrillator operating instructions using the lifepak 500 aed special situations for electrode placement when placing electrodes on the patient, be aware of the following special situations. Obese patients or patients with large breasts apply the electrodes to a f...

  • Page 41

    Lifepak 500 automated external defibrillator operating instructions 3-5 ©1996–2005 medtronic emergency response systems, inc. 3 u sin g th e li fe pa k 50 0 ae d using the lifepak 500 aed aed prompts the following paragraphs describe typical scenarios that might occur during aed operation. Topics in...

  • Page 42

    3-6 lifepak 500 automated external defibrillator operating instructions using the lifepak 500 aed shock advised if the aed detects a shockable ecg rhythm, it displays this message: you will hear the shock advised voice prompt. The aed begins charging for shock #1. A rising tone indicates that the ae...

  • Page 43

    Lifepak 500 automated external defibrillator operating instructions 3-7 ©1996–2005 medtronic emergency response systems, inc. 3 u sin g th e li fe pa k 50 0 ae d using the lifepak 500 aed after no shock advised , the aed enters cpr time if cpr time is set to 15 seconds or more. If cpr time is set to...

  • Page 44

    3-8 lifepak 500 automated external defibrillator operating instructions using the lifepak 500 aed the aed then alternately displays two messages. The cpr countdown timer indicates cpr time remaining. You will hear the corresponding voice prompt. If cpr time is set to 15, 30, 45, 60, 90, 120, or 180 ...

  • Page 45

    Lifepak 500 automated external defibrillator operating instructions 3-9 ©1996–2005 medtronic emergency response systems, inc. 3 u sin g th e li fe pa k 50 0 ae d using the lifepak 500 aed you will hear the corresponding voice prompt. Note: if auto analyze 2 is selected or for lifepak 500 aeds that d...

  • Page 46

    3-10 lifepak 500 automated external defibrillator operating instructions using the lifepak 500 aed you will hear the push analyze voice prompt accompanied by a warning tone. You should: • stop all patient and vehicle movement. • confirm that the patient is in cardiac arrest. • press analyze . Stay c...

  • Page 47

    Lifepak 500 automated external defibrillator operating instructions 4-1 ©1996–2005 medtronic emergency response systems, inc. 4 d at a ma na ge m ent data management this section describes how to store and transfer lifepak 500 automated external defibrillator (aed) data to a computer or a modem. Top...

  • Page 48

    4-2 lifepak 500 automated external defibrillator operating instructions data management overview of data storage and retrieval every time you use the lifepak 500 aed on a patient, data is stored digitally inside the aed. This data allows post-incident review for quality control, training, and resear...

  • Page 49

    Lifepak 500 automated external defibrillator operating instructions 4-3 ©1996–2005 medtronic emergency response systems, inc. Data management 4 d at a ma na ge m ent figure 4-2 comparison of data stored for the current patient and previous patient the aed stores all data for the current patient (b)....

  • Page 50

    4-4 lifepak 500 automated external defibrillator operating instructions data management if you do not connect electrodes to a patient or a simulator, you can turn on the aed and not affect the current patient. For example, you can turn on the aed to test it with the external test load or to transfer...

  • Page 51

    Lifepak 500 automated external defibrillator operating instructions 4-5 ©1996–2005 medtronic emergency response systems, inc. Data management 4 d at a ma na ge m ent sending data to a computer by modem these paragraphs describe the resources, equipment connections, and procedures required to send li...

  • Page 52

    4-6 lifepak 500 automated external defibrillator operating instructions data management procedure for sending data perform these steps to send data: 1 make sure that the equipment at the destination site is properly connected. 2 make sure that the destination computer power is on and that the quik-v...

  • Page 53

    Lifepak 500 automated external defibrillator operating instructions 4-7 ©1996–2005 medtronic emergency response systems, inc. Data management 4 d at a ma na ge m ent 9 if you answered yes , you will see the message: incident id xxxxxxxxx • press s to scroll through and select from the alphanumeric c...

  • Page 54

    4-8 lifepak 500 automated external defibrillator operating instructions data management sending data to a computer by direct connection these paragraphs describe the resources, equipment connections, and procedures required to send aed data to a computer by direct connection. Required resources tabl...

  • Page 55

    Lifepak 500 automated external defibrillator operating instructions 4-9 ©1996–2005 medtronic emergency response systems, inc. Data management 4 d at a ma na ge m ent note: if you leave the lifepak 500 aed unattended during data transfer, the aed automatically turns off after 15 minutes of no activit...

  • Page 56

    4-10 lifepak 500 automated external defibrillator operating instructions data management 4 connect the equipment as shown in figure 4-6. • connect the printer cable to the aed and the printer. Figure 4-6 connecting the aed to a printer 5 after a few seconds, you will see the message: to print push x...

  • Page 57

    Lifepak 500 automated external defibrillator operating instructions 4-11 ©1996–2005 medtronic emergency response systems, inc. Data management 4 d at a ma na ge m ent event log summary this report summarizes important events for a particular patient record. Code summary report this report includes t...

  • Page 58

    4-12 lifepak 500 automated external defibrillator operating instructions data management figure 4-7 example of event log report and event log summary incident id no: incident date: 15may04 operator id no: device type: lifepak 500 device serial no: 00001203 device id: rfd#6 25mm/sec, 1.0 cm/mv event ...

  • Page 59

    Lifepak 500 automated external defibrillator operating instructions 4-13 ©1996–2005 medtronic emergency response systems, inc. Data management 4 d at a ma na ge m ent figure 4-8 example of code summary report incident date: 15may04.

  • Page 60

    4-14 lifepak 500 automated external defibrillator operating instructions data management figure 4-9 example of code summary report (cont.) incident date: 15may04.

  • Page 61

    Lifepak 500 automated external defibrillator operating instructions 4-15 ©1996–2005 medtronic emergency response systems, inc. Data management 4 d at a ma na ge m ent figure 4-10 example of code summary report (cont.) incident date: 15may04.

  • Page 62

    4-16 lifepak 500 automated external defibrillator operating instructions data management figure 4-11 test log report example device type: lifepak 500 device serial no: 00001203 device id: rfd#6 test log report self test: pass self test: pass self test: pass self test: pass self test: pass self test:...

  • Page 63

    Lifepak 500 automated external defibrillator operating instructions 5-1 ©1996–2005 medtronic emergency response systems, inc. 5 m aint en an ce maintenance this section describes how to perform operator-level maintenance and testing on the lifepak 500 automated external defibrillator (aed). For trou...

  • Page 64

    5-2 lifepak 500 automated external defibrillator operating instructions maintenance maintenance and testing scheduling the lifepak 500 aed performs an automatic self-test every 24 hours. If the automatic self-test detects a low battery condition or a condition that requires service, the monophasic a...

  • Page 65

    Lifepak 500 automated external defibrillator operating instructions 5-3 ©1996–2005 medtronic emergency response systems, inc. 5 m aint en an ce maintenance aeds with readiness display: observe readiness display ok none needed. Battery indicator displayed replace battery immediately. Service indicato...

  • Page 66

    5-4 lifepak 500 automated external defibrillator operating instructions maintenance cleaning clean the lifepak 500 aed and accessories as described in table 5-2. Use only the cleaning agents listed in the table. Table 5-2 recommended cleaning methods testing this section describes the aed automatic ...

  • Page 67

    Lifepak 500 automated external defibrillator operating instructions 5-5 ©1996–2005 medtronic emergency response systems, inc. 5 m aint en an ce maintenance turn the aed off and on. If the call service message disappears, you can still use the aed if it is needed for patient therapy. However, you sho...

  • Page 68

    5-6 lifepak 500 automated external defibrillator operating instructions maintenance on a regular basis, the daily auto test will test for low or replace battery conditions. The daily auto test is not performed if the aed is already turned on at 0300 or if the battery is not installed. If the aed is ...

  • Page 69

    Lifepak 500 automated external defibrillator operating instructions 5-7 ©1996–2005 medtronic emergency response systems, inc. 5 m aint en an ce maintenance 4 press analyze . If auto analyze 2 is selected or you have an aed that does not have an analyze button, the aed will start analyzing automatica...

  • Page 70

    5-8 lifepak 500 automated external defibrillator operating instructions maintenance note: when a battery pak is removed from the aed, battery and service indicators appear on the readiness display. After replacing the battery pak, turn on the device to reset the readiness display. Nonrechargeable ba...

  • Page 71

    Lifepak 500 automated external defibrillator operating instructions 5-9 ©1996–2005 medtronic emergency response systems, inc. 5 m aint en an ce maintenance figures 5-2 through 5-4 illustrate the effect of these factors with three different usage levels. Figure 5-4 active life, patient use every two ...

  • Page 72

    5-10 lifepak 500 automated external defibrillator operating instructions maintenance to properly maintain nonrechargeable lithium battery paks: • do not attempt to recharge (lithium battery paks cannot be connected to the battery charger used to recharge the rechargeable sla battery paks). • do not ...

  • Page 73

    Lifepak 500 automated external defibrillator operating instructions 5-11 ©1996–2005 medtronic emergency response systems, inc. 5 m aint en an ce maintenance in the usa, environmental protection agency and department of transportation regulations allow disposal of lithium batteries with ordinary hous...

  • Page 74

    5-12 lifepak 500 automated external defibrillator operating instructions maintenance • recharge until the battery charger charge led is green. This indicates that the battery charger has completed the fast-charge cycle. Undercharging can cause battery damage. • recharge only at temperatures between ...

  • Page 75

    Lifepak 500 automated external defibrillator operating instructions 5-13 ©1996–2005 medtronic emergency response systems, inc. 5 m aint en an ce maintenance a fully charged battery is not harmed if it remains connected to the battery charger. However, if a battery is disconnected and then reconnecte...

  • Page 76

    5-14 lifepak 500 automated external defibrillator operating instructions maintenance • preparation the device should be clean and contaminant-free prior to being recycled. • recycling of disposable electrodes after disposable electrodes are used, follow your local clinical procedures for recycling. ...

  • Page 77

    Lifepak 500 automated external defibrillator operating instructions 5-15 ©1996–2005 medtronic emergency response systems, inc. 5 m aint en an ce maintenance specifications table 5-4 lists the specifications for the monophasic, biphasic, and public safety (dps) lifepak 500 aeds. The specifications ap...

  • Page 78

    5-16 lifepak 500 automated external defibrillator operating instructions maintenance safety classification internally powered equipment iec 60601-1/en 60601-1, 5.1. Waveform monophasic: monophasic pulse (edmark) per aami df2-1989, 3.2.1.5.1 biphasic and dps: truncated exponential, with voltage and d...

  • Page 79

    Lifepak 500 automated external defibrillator operating instructions 5-17 ©1996–2005 medtronic emergency response systems, inc. 5 m aint en an ce maintenance event documentation type internal digital memory. Memory capacity 20 minutes audio recording (optional). Ecg and event log of operator/device a...

  • Page 80

    5-18 lifepak 500 automated external defibrillator operating instructions maintenance vibration monophasic: mil-std-810e, method 514.4, category 10 biphasic: mil-std-810e, method 514.4, helicopter- category 6 (3.75 g rms ) and ground mobile - category 8 (3.15 g rms ). Rtca/do-160c, table 8-2 fixed wi...

  • Page 81

    Lifepak 500 automated external defibrillator operating instructions 5-19 ©1996–2005 medtronic emergency response systems, inc. 5 m aint en an ce maintenance capacity typical: 312 full discharges or 14 hours of “on time” with a new battery. Minimum: 230 full discharges with a new battery at 20°c. 0°–...

  • Page 82

    5-20 lifepak 500 automated external defibrillator operating instructions maintenance table 5-5 lifepak 500 aed battery charger specifications waveform biphasic truncated exponential waveform general safety classification class ii (double insulation), iec 60601/en 60601, 5.1 input 100–240v 0.7–0.4a 5...

  • Page 83

    Lifepak 500 automated external defibrillator operating instructions 5-21 ©1996–2005 medtronic emergency response systems, inc. 5 m aint en an ce maintenance clinical summary: defibrillation of ventricular fibrillation and ventricular tachycardia background medtronic conducted a multi-centered, prosp...

  • Page 84

    5-22 lifepak 500 automated external defibrillator operating instructions maintenance ventricular tachycardia seventy-two episodes of ventricular tachycardia (vt), induced in 62 patients, were treated with randomized shocks. High rates of conversion were observed with biphasic and monophasic shocks. ...

  • Page 85

    Lifepak 500 automated external defibrillator operating instructions 6-1 ©1996–2005 medtronic emergency response systems, inc. 6 tr ou ble sh oo tin g troubleshooting this section describes how to troubleshoot lifepak 500 automated external defibrillator (aed) operating problems. This section also de...

  • Page 86

    6-2 lifepak 500 automated external defibrillator operating instructions troubleshooting troubleshooting during patient care table 6-1 troubleshooting during patient care observation possible cause corrective action 1 connect electrodes message appears. Inadequate connection to aed. • check for compl...

  • Page 87

    Lifepak 500 automated external defibrillator operating instructions 6-3 ©1996–2005 medtronic emergency response systems, inc. 6 tr ou ble sh oo tin g troubleshooting 6 service indicator on key panel flashing and call service message appears. A fault requiring immediate service. • turn aed off and on...

  • Page 88

    6-4 lifepak 500 automated external defibrillator operating instructions troubleshooting try again, to send push or cannot send messages. (continued) cable is not properly connected. • check connections. Modem is not connected to an analog telephone line. • confirm that the telephone line is analog (...

  • Page 89

    Lifepak 500 automated external defibrillator operating instructions 6-5 ©1996–2005 medtronic emergency response systems, inc. 6 tr ou ble sh oo tin g troubleshooting table 6-4 troubleshooting during setup transfer printer does not print. (continued) wrong type of printer. • check the printer to make...

  • Page 90

    6-6 lifepak 500 automated external defibrillator operating instructions troubleshooting table 6-5 lifepak 500 aed screen messages screen message description analyzing now the aed is analyzing the patient ecg rhythm. Asystole the aed has analyzed the patient’s ecg and detected persistent asystole. As...

  • Page 91

    Lifepak 500 automated external defibrillator operating instructions 6-7 ©1996–2005 medtronic emergency response systems, inc. 6 tr ou ble sh oo tin g troubleshooting motion detected the aed detects motion during ecg analysis, thereby inhibiting analysis. Motion detection setup mode message for motio...

  • Page 92

    6-8 lifepak 500 automated external defibrillator operating instructions troubleshooting table 6-6 lifepak 500 aed voice prompts voice prompt description analyzing now, stand clear the aed is analyzing the patient ecg rhythm. Asystole the aed has analyzed the patient ecg and detected persistent asyst...

  • Page 93

    Lifepak 500 automated external defibrillator operating instructions 6-9 ©1996–2005 medtronic emergency response systems, inc. 6 tr ou ble sh oo tin g troubleshooting table 6-7 lifepak 500 aed event types possible event types * * these events and all voice prompts may appear in the event log report. ...

  • Page 95

    Lifepak 500 automated external defibrillator operating instructions ©1996–2005 medtronic emergency response systems, inc. Ap pe nd ix a appendix a shock advisory system this section describes the basic function of the shock advisory system (sas)..

  • Page 97

    Lifepak 500 automated external defibrillator operating instructions a-1 ©1996–2005 medtronic emergency response systems, inc. Shock advisory system ap pe nd ix a overview of the shock advisory system the shock advisory system (sas) is an ecg analysis system built into the lifepak 500 aed that advise...

  • Page 98

    A-2 lifepak 500 automated external defibrillator operating instructions shock advisory system the lifepak 500 defibrillator was also tested using ecgs acquired from hospitalized pediatric patients ranging in age from nonshockable: asystole 43 >95% specificity 100.0% [94.8%] nonshockable: all other r...

  • Page 99

    Lifepak 500 automated external defibrillator operating instructions a-3 ©1996–2005 medtronic emergency response systems, inc. Shock advisory system ap pe nd ix a control of shock therapy operator control of shock therapy the shock advisory system causes the aed to charge automatically when it detect...

  • Page 101

    Lifepak 500 automated external defibrillator operating instructions ©1996–2005 medtronic emergency response systems, inc. Appe ndix b appendix b lifepak 500 operator’s checklist this operator’s checklist may be reproduced..

  • Page 103

    Lifepak ® 500 automated external defibrillator operator’s checklist unit serial no.: location: instruction recommended corrective action date initials 1 examine the aed case, connector, and battery well for: foreign substances clean the device. Damage or cracks contact authorized service personnel. ...

  • Page 105

    Lifepak 500 automated external defibrillator operating instructions ©1996–2005 medtronic emergency response systems, inc. Appen dix c appendix c fast-patch defibrillation cable instructions for use.

  • Page 107

    Lifepak 500 automated external defibrillator operating instructions c-1 ©1996–2005 medtronic emergency response systems, inc. Fast-patch ® defibrillation cable for lifepak ® 500 automated external defibrillator instructions for use introduction to use fast-patch disposable defibrillation/ecg electro...

  • Page 108

    Fast-patch ® defibrillation cable for lifepak ® 500 automated external defibrillator instructions for use (continued) c-2 lifepak 500 automated external defibrillator operating instructions disconnecting from disposable electrodes disconnect the defibrillation cable from the electrode by pulling the...

  • Page 109

    Lifepak 500 automated external defibrillator operating instructions ©1996–2005 medtronic emergency response systems, inc. Ap pe nd ix d appendix d quik-combo defibrillation cable instructions for use.

  • Page 111

    Quik-combo ™ defibrillation cable for lifepak ® 500 automated external defibrillator instructions for use lifepak 500 automated external defibrillator operating instructions d-1 ©1996–2005 medtronic emergency response systems, inc. Introduction if using standard quik-combo pacing/ defibrillation/ecg...

  • Page 112

    Quik-combo ™ defibrillation cable for lifepak ® 500 automated external defibrillator instructions for use (continued) d-2 lifepak 500 automated external defibrillator operating instructions connecting to quik-combo pacing/ defibrillation/ecg electrodes properly connect the defibrillation cable to th...

  • Page 113

    Lifepak 500 automated external defibrillator operating instructions ©1996–2005 medtronic emergency response systems, inc. Appe ndix e appendix e declarations of conformity / electromagnetic compatibility guidance.

  • Page 115

    Lifepak 500 automated external defibrillator operating instructions e-1 ©1996–2005 medtronic emergency response systems, inc. Appe ndix e declarations of conformity / electromagnetic compatibility guidance * see emc tables ec declaration of conformity manufacturer’s name: medtronic emergency respons...

  • Page 116

    E-2 lifepak 500 automated external defibrillator operating instructions declarations of conformity / electromagnetic compatibility guidance ec declaration of conformity we hereby declare under our sole responsibility that the product model(s) bcwa-042000-100a and bcwa-042000-100n (medtronic lifepak ...

  • Page 117

    Lifepak 500 automated external defibrillator operating instructions e-3 ©1996–2005 medtronic emergency response systems, inc. Appe ndix e declarations of conformity / electromagnetic compatibility guidance essential performance the lifepak 500 aed maintains safe and effective performance of the defi...

  • Page 118

    E-4 lifepak 500 automated external defibrillator operating instructions declarations of conformity / electromagnetic compatibility guidance table e-2 guidance and manufacturer’s declaration - electromagnetic immunity guidance and manufacturer’s declaration - electromagnetic immunity the lifepak 500 ...

  • Page 119

    Lifepak 500 automated external defibrillator operating instructions e-5 ©1996–2005 medtronic emergency response systems, inc. Appe ndix e declarations of conformity / electromagnetic compatibility guidance table e-3 guidance and manufacturer’s declaration - electromagnetic immunity guidance and manu...

  • Page 120

    E-6 lifepak 500 automated external defibrillator operating instructions declarations of conformity / electromagnetic compatibility guidance table e-4 recommended separation distances recommended separation distances between portable and mobile rf communications equipment and the lifepak 500 aed the ...

  • Page 121

    Lifepak 500 automated external defibrillator operating instructions index-1 ©1996–2005 medtronic emergency response systems, inc. In de x index a accessories battery charger , 2-4 carrying case , 2-4 cleaning , 5-4 data cable , 2-4 inspection , 5-2 ordering , 5-14 part numbers , 5-14 quik-combo elec...

  • Page 122

    Index-2 lifepak 500 automated external defibrillator operating instructions index audio recording definition , 4-2 setup option , 2-11 audio recording setup option , 2-11 auto analyze setup option , 2-9 auto tests , 5-5 daily auto test , 5-5 extended auto test , 5-6 b battery installation , 2-5 low ...

  • Page 123

    Lifepak 500 automated external defibrillator operating instructions index-3 ©1996–2005 medtronic emergency response systems, inc. Index in de x intended use statements , x l lcd general description , 2-3 messages and definitions , 6-6 location tracking of defibrillators , ii log report. See event lo...

  • Page 124

    Index-4 lifepak 500 automated external defibrillator operating instructions index t telephone number for modem , 2-7 test load test , 5-6 test load, general description , 2-4 test log report , 4-4 description , 4-11 example , 4-16 test mode message , 5-6 testing the aed , 5-4 auto tests , 5-5 daily ...

  • Page 125

    Device tracking the u.S. Food and drug administration classifies defibrillators as a medical device that requires tracking (knowing where the device is). As such, federal regulations require that manufacturers maintain tracking information for each device distributed. We rely on our customers to pro...

  • Page 128

    Medtronic emergency response systems 11811 willows road northeast redmond, wa 98052-2003 usa telephone: 425.867.4000 toll free (usa only): 800.442.1142 fax: 425.867.4121 internet: www.Medtronic-ers.Com www.Medtronic.Com medtronic europe s.A. Medtronic emergency response systems rte. Du molliau 31 ca...