Medtronic MiniMed Pro-set Change instructions
Changing the
MiniMed
®
Pro-set
™
Infusion Set:
MiniMed
®
Revel
™
Insulin Pump
For more information please see the MiniMed Paradigm REAL-Time Revel insulin pump User Guide.
1) Wash your hands. Remove the old infusion set from your body and the
reservoir from the pump.
2) Press
ACT
for the MAIN MENU, and select
Reservoir + Set then select
Reservoir Setup to open the REWIND screen. Press
ACT
to rewind. Stop
at the
REWIND COMPLETE screen.
1) Remove the reservoir
from the package.
6) Check the reservoir for
air bubbles. Tap the side
of the reservoir to force
any bubbles to rise to the
top. Gently push up on the
plunger to move the air
bubbles from the reservoir
into the insulin vial.
3) Wipe the top of the insulin vial
with alcohol. Hold the
blue
transfer guard and press it
down onto the insulin vial.
7) After the air bubbles are
removed, slowly pull the
plunger down to fill the
reservoir with enough
insulin to last 2-3 days.
4) Push down on the plunger to
push air from the reservoir
into the vial. Continue to hold
the plunger down.
8) To avoid getting insulin on the
top of the reservoir, flip the vial
over, so the vial is upright. Hold
the transfer guard, and turn the
reservoir counter-clockwise.
Pull the reservoir straight up to
disconnect it from the transfer
guard. Discard the transfer
guard into a sharps container.
5) Flip the vial over so it is now
on top. Slowly pull the plunger
down to fill the reservoir.
1
Reservoir Setup
Fill Cannula
History
RESERVOIR + SET MENU
If insulin, or any liquid, gets inside the tubing connector, it can temporarily
block the vents that allow the pump to properly prime the infusion set.
This may result in the delivery of too little or too much insulin, which can
cause hyperglycemia or hypoglycemia. To prevent insulin from getting
inside the tubing connector of the MiniMed
®
Pro-set
™
infusion set, after
you fill the reservoir make sure you hold the insulin vial upright when you
remove the reservoir from the transfer guard. If you do not hold the insulin
vial upright, insulin can get on the top of the reservoir and could transfer liquid into the tubing connector.
Top of reservoir
Tubing Connector
PART 1: REWIND THE PISTON IN THE PUMP
PART 2: FILL THE RESERVOIR
2) Pull the plunger down to fill
the reservoir with air.
plunger
rod
reservoir
transfer
guard
WARNING: