  • Page 1

    Changing the silhouette® infusion set using a revel™ insulin pump to rewind the piston in the pump: to fill the reservoir: 1. Remove reservoir from package. 2.Pull plunger rod down to fill reservoir with air 3. After wiping top of insulin vial with alcohol, hold blue transfer guard and press it down...

  • Page 2

    Changing the silhouette® infusion set using a revel™ insulin pump to connect the tubing: 1. Remove silhouette from package. 2. Hold reservoir in one hand. With other hand gently push connector onto top of reservoir and turn clockwise until it slides and locks into place. 3. Tap side of reservoir to ...

  • Page 3

    Changing the silhouette® infusion set using a revel™ insulin pump to fill the tubing: 5. Press and hold act to fill tubing. The pump will beep as tubing fills. Release act when you see drops at tip of connector needle. Then press esc . 6.If you see drops at end of connector needle and there are no a...

  • Page 4

    Changing the silhouette® infusion set using a revel™ insulin pump to insert silhouette® infusion set: method 2 - by sil-serter® 1. Use thumb to press down on lock & release handle of sil-serter®. 2. Place silhouette® insertion piece inside sil-serter® and let go of the lock & release handle. The ser...

  • Page 5

    Changing the silhouette® infusion set using a revel™ insulin pump to remove introducer needle: 1. Peel paper from front part of adhesive and press adhesive securely to skin. 2. Squeeze arms on gray needle handle and pull backwards to remove introducer needle. 3.Peel rest of paper without lifting inf...