Megger OTS60SX/2 User Manual

Manual is about: Semi Automatic Oil Test Set

Summary of OTS60SX/2

  • Page 1

    M ots60sx/2 semi automatic oil test set user guide.

  • Page 2: Contents

    Contents safety warnings 3 general description 4 preparing the oil test set 5 operation 5 oil sampling 5 preparing the oil test vessel 5 control panel 6 breakdown testing 6 withstand testing 6 warning messages 7 testing to national standards 7 sampling and testing oil for dielectric strength 8 oil t...

  • Page 3

    3. Safety warnings ■ safety warnings and precautions must be read and understood before the instrument is used. They must be observed during use. ■ the oil test set must be properly earthed. ■ the test chamber must be kept scrupulously clean. ■ if the test chamber cover is cracked or damaged in any ...

  • Page 4

    Mains input panel general description 4. The ots60sx/2 is a lightweight, semi-automatic, oil dielectric strength test set. The instrument is suitable for protected field use and can be powered from a range of mains supplies. The maximum 60 kv output allows tests to be performed on oil from a wide va...

  • Page 5

    Ensure that the instrument is properly earthed. The test set must be connected to a socket (receptacle) with a protective earth (ground) conductor. Before connecting the instrument to the mains power supply, ensure the voltage selector located on the mains input panel is set to the required voltage....

  • Page 6

    Control panel the control panel contains the supply on/off switch, the high voltage display, voltage ramp rate selection and the three control keys. The operation of the instrument is very simple. When the instrument is turned on, the on/off switch is illuminated, and the software edition code will ...

  • Page 7

    7 warning messages door open if the instrument overheats, the words ‘ ‘ and ‘ ‘ will be alternately displayed and the operation disabled. This will reset automatically, when cooled, after a few minutes. Testing to national standards the ots60sx/2 can be used to test to any international standard. A ...

  • Page 8

    The following notes should be read in conjunction with the relevant oil testing specifications. Causes of bad oil transformers and switchgear oil may be rendered unsuitable for further use due to four main reasons:- 1. Low dielectric strength. 2. High acidity. 3. High sludge content. 4. Excessive fr...

  • Page 9

    9 particular care to prevent contamination is necessary when a test is to be made on a bulk delivery tanker, and on all occasions when a high value of dielectric strength is anticipated. The procedure suggested above should be rigorously applied..

  • Page 10

    10 oil testing specifications test specification as1767 astm d 877 astm d 1816 astm d 1816 bs 148 bs 5730a bs 5874 cei 344 en 60156 iec 156 ip 295 ijis c 2101 ksc 2101 nfc 27-131 nfc 27-221 roct 6581 sabs 555 stas 286 une 21 vde 0370 electrode shape a,b c a a a,b a,b a,b a a,b a,b a,b b b a,b a,b a,...

  • Page 11

    Specification 11 output voltage 0 - 60 kv rms test frequency 61,8 hz transformer rating 500 va output disconnection within 1 ms of detection of breakdown power supply 115 v (99 - 132 v), 230 v (198 - 264 v) (switch selectable) 50/60 hz, 80 va fuses 500 ma (t) hbc to iec127/5 for 220/240 v supply 1 a...

  • Page 12

    12 accessories supplied with instrument part number spacing gauge (0,5, 1,0, 1,5, 2,0, 2,5, 2,54 & 4mm) 6132-009 accessory pouch 6320-232 fuses: 2 x 500 ma (t) for 230 v supply (fitted to the instrument) 25950-014 2 x 1 a (t) for 115 v supply 25950-004 mains lead (fit your own plug) 25424-860 mains ...

  • Page 13

    Repair and warranty 13 the instrument circuit contains static sensitive devices, and care must be taken in handling the printed circuit board. If the protection of an instrument has been impaired it should not be used, and be sent for repair by suitably trained and qualified personnel. The protectio...

  • Page 14

    Megger limited megger megger archcliffe road dover 4271 bronze way, dallas, z.A. Du buisson de la couldre kent ct17 9en england tx 75237-1019 usa 23 rue eugène henaff t +44 (0)1 304 502101 t +1 800 723 2861 78190 trappes france f +44 (0)1 304 207342 t +1 214 333 3201 t +33 (1) 30.16.08.90 f +1 214 3...