NARDA SRM-3006 Command Reference Manual

Other manuals for SRM-3006: Manual, Manual
Manual is about: Selective Radiation Meter

Summary of SRM-3006

  • Page 1

    Srm-3006 selective radiation meter command reference guide v 1.5.0

  • Page 2

    Narda safety test solutions gmbh sandwiesenstraße 7 72793 pfullingen, germany © 2017 ® names and logo are registered trademarks of narda safety test solutions gmbh and l3 communications holdings, inc. Trade names are trademarks of the owners. Order no.: 3006/98.22 version: 31/03.2017, s/n a ... (fir...

  • Page 3

    Command reference guide narda srm-3006 i contents 1 general information ...........................................................................................................................1 backus-naur format .......................................................................................

  • Page 4

    Command reference guide ii srm-3006 narda dl_data? (spectrum) ............................................................................................................. 29 dl_data? (scope) ...............................................................................................................

  • Page 5

    Command reference guide narda srm-3006 iii mode? ..................................................................................................................................... 86 mr_list? ............................................................................................................

  • Page 6

    Command reference guide iv srm-3006 narda srv_list? ............................................................................................................................. 125 srv_sel .................................................................................................................

  • Page 8

    Command reference guide 2 srm-3006 narda data types abbreviation description example string ascii character string “funny fox” enum enumerator, fixed texts on, off, remote float/double floating point number 12.34 or 1.234e1 byte 8 bit whole number without sign 123 shortint 16 bit whole number with s...

  • Page 9: 2  Command Groups

    Command reference guide narda srm-3006 3 2 command groups general dev_id? Queries the device identification number dev_info? Queries device information that describes the individual device dev_option activates a device option dev_option? Queries the options prepared and activated in the device error...

  • Page 10

    Command reference guide 4 srm-3006 narda mode spectrum spectrum? Queries one or more measurement runs in spectrum mode spectrum _avg_config sets the current averaging method spectrum _avg_config? Queries the current averaging method spectrum _avg_list? Queries the current possible averaging settings...

  • Page 11

    Command reference guide narda srm-3006 5 clock date sets the current date setting for the device date? Queries the current date setting for the device time sets the current system time of the device time? Queries the current system time of the device data logger dl_clr deletes a selected data set dl...

  • Page 12: 3  Command Descriptions

    Command reference guide 6 srm-3006 narda 3 command descriptions date description sets the current date setting for the device input date ; output ; parameter list parameter type description date dd.Mm.Yy dword error code example (input) date 06.07.12; (output) 0;.

  • Page 13: Date?

    Command reference guide narda srm-3006 7 date? Description queries the current date setting for the device input date?; output ,; parameter list parameter type description date dd.Mm.Yy dword error code example (input) date?; (output) 03.05.10,0;.

  • Page 14: Dev_Id?

    Command reference guide 8 srm-3006 narda dev_id? Description queries the device identification number. Input dev_id?; output ,; parameter list parameter type description string device identification number dword error code example (input) dev_id?; (output) "f89aef31cd344840",0;.

  • Page 15: Dev_Info?

    Command reference guide narda srm-3006 9 dev_info? Description queries all device information that describes the individual device. Input dev_info?; output ,,,, ,, ,,; parameter list parameter type description string product name (e.G. “srm-3006”) string production identification number string seria...

  • Page 16: Dev_Option

    Command reference guide 10 srm-3006 narda dev_option description activates a device option. Options are additional measurement modes that the device is prepared for. Each device needs a separate activation code for each option, which can be purchased. The current device firmware determines whether a...

  • Page 17: Dev_Option?

    Command reference guide narda srm-3006 11 dev_option? Description queries the options prepared and activated in the device. Options are additional measurement modes that the device is prepared for. Each device needs a separate activation code for each option, which can be purchased. The current devi...

  • Page 18: Dl_Clr

    Command reference guide 12 srm-3006 narda dl_clr description deletes a selected data set (datalogger). All associated sub data sets are also deleted. Input dl_clr ; output ; parameter list parameter type description integer index of the selected dataset dword error code example (input) dl_clr 3; (ou...

  • Page 19: Dl_Clr_All

    Command reference guide narda srm-3006 13 dl_clr_all description deletes all data sets (datalogger) and all associated sub data sets. Input dl_clr_all; output ; parameter list parameter type description dword error code example (input) dl_clr_all; (output) 0;.

  • Page 20: Dl_Data?

    Command reference guide 14 srm-3006 narda dl_data? (general) description queries a sub data set (datalogger) a data set always comprises one or more sub data sets. The store mode determines whether each data set is saved separately or as a group of sub data sets. Data sets can be substantial in size...

  • Page 21: Dl_Data?

    Command reference guide narda srm-3006 15 dl_data? (levelrecording) description queries a sub data set (datalogger) in levelrecording mode input dl_data? ,; output ,,,,,, ,,,, ,,, ,,,,, ,,,, ,,,,,,,, , ,,,,,,, ,, ,,,, {,,,,} ; input parameter list parameter type description short index of the select...

  • Page 22

    Command reference guide 16 srm-3006 narda output parameter list, section: general parameter type description dword unique id of the sub-dataset enum level enum man manual storing, executed by the user cond_first auto-store at first true condition cond_all auto-store for all true conditions time auto...

  • Page 23

    Command reference guide narda srm-3006 17 output parameter list, section: setup common parameter type description float reference level (measurement range) [unit] enum unit enum unit category: a, b, c, d a: referred to rf input 1 w (@ 50 ohms) b: for connected antennas 1 w/m 2 (@ 377 ohms) c: for co...

  • Page 24

    Command reference guide 18 srm-3006 narda output parameter list, section: trace common parameter type description long number of sweeps performed short 0 – 100 [%] progress of averaging long 0 – 999 999 number of sweeps used for spatial averaging short number of traces output parameter list, section...

  • Page 25: Dl_Data?

    Command reference guide narda srm-3006 19 dl_data? (safetyevaluation, condensed) description queries a sub data set (datalogger) in safetyevaluation mode and in ‘condensed’ format. Services are grouped together as a result of service names with the same prefix. Input dl_data? ,; output ,,,,,, ,,,, ,...

  • Page 26

    Command reference guide 20 srm-3006 narda output parameter list, section: general parameter type description dword unique id of the sub-dataset enum safety_condensed enum man manual storing, executed by the user cond_first auto-store at first true condition cond_all auto-store for all true condition...

  • Page 27

    Command reference guide narda srm-3006 21 output parameter list, section: setup common parameter type description float reference level (measurement range) [unit] enum unit enum unit category: a, b, c, d see dl_data? (level) float value required to determine yref when changing between unit categorie...

  • Page 28

    Command reference guide 22 srm-3006 narda output parameter list, section: trace common parameter type description long number of sweeps performed integer sweep time [ms] short 0 – 100 [%] progress of averaging long 0 – 999 999 number of sweeps used for spatial averaging integer number of channels (s...

  • Page 29

    Command reference guide narda srm-3006 23 example (input) dl_data? 25,1; (output) 8068156,safety_condensed,man,10.05.10,11:43:30,no,no,gps,2d,0,0,0,0,no,"", "sw-0003",12.03.10,"v0.0.2 beta1","",01.01.01,"serial no",21.09.09, 0.0071,v/m,b,-13.0103,single,"icnirp 1998 general public","Österreich umts"...

  • Page 30: Dl_Data?

    Command reference guide 24 srm-3006 narda dl_data? (safetyevaluation, detailed) description queries a sub data set (datalogger) in safetyevaluation mode and in ‘detailed’ format. Input dl_data? ,; output ,,,,,, ,,,, ,,, ,,,,, ,,,, ,,,,,,,, , ,,,,,, ,,,,,, ,,,,, ,,,,,, ,{,,,,,,} ; input parameter lis...

  • Page 31

    Command reference guide narda srm-3006 25 output parameter list, section: general parameter type description dword unique id of the sub-dataset enum safety enum man manual storing, executed by the user cond_first auto-store at first true condition cond_all auto-store for all true conditions time aut...

  • Page 32

    Command reference guide 26 srm-3006 narda output parameter list, section: setup common parameter type description float reference level (measurement range) [unit] enum unit enum unit category: a, b, c, d see dl_data? (level) float value required to determine yref when changing between unit categorie...

  • Page 33

    Command reference guide narda srm-3006 27 output parameter list, section: trace common parameter type description long number of sweeps performed integer sweep time [ms] short 0 – 100 [%] progress of averaging long 0 – 999 999 number of sweeps used for spatial averaging short number of traces, 1 - 7...

  • Page 34

    Command reference guide 28 srm-3006 narda example (input) dl_data? 22,1; (output) 2996901,safety,man,10.05.10,10:01:34,no,no,gps,2d,0,0,0,0,no,"", "sw-0003",12.03.10,"v0.0.2 beta1","",01.01.01,"serial no",21.09.09, 0.0071,v/m,b,-13.0103,single,"icnirp 1998 general public","swisscom","","ant zero", n...

  • Page 35: Dl_Data?

    Command reference guide narda srm-3006 29 dl_data? (spectrum) description queries a sub data set (datalogger) in spectrum mode. Input dl_data? ,; output ,,,,,, ,,,, ,,, ,,,,, ,,,, ,,,,,,,, , ,,,,,,,, ,, ,,,,,, ,,,{,} ; input parameter list parameter type description short index of the selected datas...

  • Page 36

    Command reference guide 30 srm-3006 narda output parameter list, section: general parameter type description dword unique id of the sub-dataset enum spectrum enum man manual storing, executed by the user cond_first auto-store at first true condition cond_all auto-store for all true conditions time a...

  • Page 37

    Command reference guide narda srm-3006 31 output parameter list, section: setup common parameter type description float reference level (measurement range) [unit] enum unit enum unit category: a, b, c, d see dl_data? (level) float value required to determine yref when changing between unit categorie...

  • Page 38

    Command reference guide 32 srm-3006 narda output parameter list, section: trace specific parameter type description enum available traces: act actual value avg averaged value max maximum value max_avg maximum of the averaged values min minimum value min_avg minimum of the averaged values std values ...

  • Page 39: Dl_Data?

    Command reference guide narda srm-3006 33 dl_data? (scope) description queries a sub data set (datalogger) in scope mode. Input dl_data? ,; output ,,,,,, ,,,, ,,, ,,,,, ,,,, ,,,,,,,, , ,,,,,, ,,,,,, ,, ,,, {,,,{,} } ,{,} ; parameter list parameter type description short index of the selected dataset...

  • Page 40

    Command reference guide 34 srm-3006 narda output parameter list, section: general parameter type description dword unique id of the sub-dataset enum scope enum man manual storing, executed by the user cond_first auto-store at first true condition cond_all auto-store for all true conditions time auto...

  • Page 41

    Command reference guide narda srm-3006 35 output parameter list, section: setup common parameter type description float reference level (measurement range) [unit] enum unit enum unit category: a, b, c, d see dl_data? (level) float value required to determine yref when changing between unit categorie...

  • Page 42

    Command reference guide 36 srm-3006 narda output parameter list, section: trace specific parameter type description enum max, avg, min, [std] or act, [std] trace types are depending on rbw and time resolution settings. For time resolution >= 4 / rbw only trace types max, avg, min, [std] will always ...

  • Page 43: Dl_Data?

    Command reference guide narda srm-3006 37 dl_data? (umts) description queries a sub data set (datalogger) in umts mode. Input dl_data? ,; output ,,,,,, ,,,, ,,, ,,,,, ,,,, ,,,,,,,, , ,,;,,, ,,,, ,,,, {,,,,,, {,,,} } ; input parameter list parameter type description short index of the selected datase...

  • Page 44

    Command reference guide 38 srm-3006 narda output parameter list, section: general parameter type description dword unique id of the sub-dataset enum umts enum man manual storing, executed by the user cond_first auto-store at first true condition cond_all auto-store for all true conditions time auto-...

  • Page 45

    Command reference guide narda srm-3006 39 output parameter list, section: setup common parameter type description float reference level (measurement range) [unit] enum unit enum unit category: a, b, c, d see dl_data? (level) float value required to determine yref when changing between unit categorie...

  • Page 46

    Command reference guide 40 srm-3006 narda output parameter list, section: trace common parameter type description long number of sweeps performed short sweep time [ms] short 0 – 100 [%] progress of averaging short number of traces output parameter list, section: trace specific parameter type descrip...

  • Page 47: Dl_Data?

    Command reference guide narda srm-3006 41 dl_data? (lte) the lte commands applies for the lte fdd optional mode description queries a sub data set (datalogger) in lte mode. Input dl_data? ,; output ,,,,,, ,,,, ,,, ,,,,, ,,,, ,,,,,,,, , ,,,;,,, ,,, ,,,, {,,,} , {,,,,, {,,,, {value,} } ; input paramet...

  • Page 48

    Command reference guide 42 srm-3006 narda output parameter list, section: general parameter type description dword unique id of the sub-dataset enum lte enum man manual storing, executed by the user cond_first auto-store at first true condition cond_all auto-store for all true conditions time auto-s...

  • Page 49

    Command reference guide narda srm-3006 43 output parameter list, section: setup common parameter type description float reference level (measurement range) [unit] enum unit enum unit category: a, b, c, d see dl_data? (level) float value required to determine yref when changing between unit categorie...

  • Page 50

    Command reference guide 44 srm-3006 narda output parameter list, section: trace common parameter type description long number of sweeps performed short sweep time [ms] short 0 – 100 [%] progress of averaging short >= 0 short 0 4 503 short >= 1 enum yes, no short number of traces output parameter lis...

  • Page 51

    Command reference guide narda srm-3006 45 example (input) dl_data? 7,1; (output) 1913414,lte,man,18.06.14, 9:50:16,no,actual,gps,3d,6,504,48.4585316667,9.23057166667,no,"", "d-0051",14.09.10,"v1.3.2","",01.01.01,"",01.01.01, -30,dbm,a,0,single,"icnirp 1998 general public","germany alle funkdienste",...

  • Page 52: Dl_Data?

    Command reference guide 46 srm-3006 narda dl_data? (lte tdd) description queries a sub data set (datalogger) in lte tdd mode. Input dl_data? ,; output ,,,,,, ,,,, ,,, ,,,,, ,,,, ,,,,,,,, , ,,,;,,, ,,,,, ,,,, {,,,} , {,,,,, {,,,, {value,} } ; input parameter list parameter type description short inde...

  • Page 53

    Command reference guide narda srm-3006 47 output parameter list, section: general parameter type description dword unique id of the sub-dataset enum lte_tdd enum man manual storing, executed by the user cond_first auto-store at first true condition cond_all auto-store for all true conditions time au...

  • Page 54

    Command reference guide 48 srm-3006 narda output parameter list, section: setup common parameter type description float reference level (measurement range) [unit] enum unit enum unit category: a, b, c, d see dl_data? (level) float value required to determine yref when changing between unit categorie...

  • Page 55

    Command reference guide narda srm-3006 49 output parameter list, section: trace common parameter type description long number of sweeps performed short sweep time [ms] short 0 – 100 [%] progress of averaging short >= 0 short 0 4 503 short >= 1 enum yes, no short number of traces output parameter lis...

  • Page 56

    Command reference guide 50 srm-3006 narda example (input) dl_data? 7,1; (output) 1913414,lte_tdd,man,18.06.14, 9:50:16,no,actual,gps,3d,6,504,48.4585316667,9.23057166667,no,"", "d-0051",14.09.10,"v1.3.2","",01.01.01,"",01.01.01, -30,dbm,a,0,single,"icnirp 1998 general public","germany alle funkdiens...

  • Page 57: Dl_Info?

    Command reference guide narda srm-3006 51 dl_info? Description queries overview information for a data set. Input dl_info? ; output ,,,,,,, ,; parameter list parameter type description integer index of the desired dataset integer number of related sub-datasets enum spectrum, safety, umts, scope, lev...

  • Page 58: Dl_Info_Sub?

    Command reference guide 52 srm-3006 narda dl_info_sub? Description queries overview information for a sub data set. Input dl_info? ,subds_idx; output ,,,,; parameter list parameter type description integer index of the desired dataset integer index of the desired sub-dataset enum spectrum, safety, u...

  • Page 59: Dl_Memory?

    Command reference guide narda srm-3006 53 dl_memory? Description queries the free memory space available for storing data sets. The response is a relative number expressed in %. Input dl_memory?; output ,; parameter list parameter type description integer free memory space, 0 - 100 [%] dword error c...

  • Page 60: Dl_Number?

    Command reference guide 54 srm-3006 narda dl_number? Description queries the number of saved data sets. Input dl_number?; output ,; parameter list parameter type description integer number of stored datasets dword error code example (input) dl_number?; (output) 33,0;.

  • Page 61: Dl_Store

    Command reference guide narda srm-3006 55 dl_store description saves a data set in the current measurement mode. Input dl_store; output ; parameter list parameter type description dword error code example (input) dl_store; (output) 0;.

  • Page 62: Dl_Voice?

    Command reference guide 56 srm-3006 narda dl_voice? Description queries the voice comment of a data set in pcm wave format. Input dl_voice? ,; output ,,; parameter list parameter type description integer index of the desired dataset long number of characters per line (block-wise transfer) 0 no verti...

  • Page 63: Error?

    Command reference guide narda srm-3006 57 error? Description queries the current error code. Input error?; output ,; parameter list parameter type description enum current error code: 0 no error 401 remote command is not implemented in the remote module 402 invalid parameter 403 invalid count of par...

  • Page 64: Level?

    Command reference guide 58 srm-3006 narda level? Description queries the different measurement runs (traces) that are available in levelrecorder mode. Levelrecorder mode must be selected as the current operating mode. Input level? ; output ,,,, ,,,,; parameter list parameter type description enum rm...

  • Page 65: Level_Avg_Config

    Command reference guide narda srm-3006 59 level_avg_config description sets the current averaging method. The device has 1 averaging method: • average over a time period (time) input level_avg_config ; output ; parameter list parameter type description float time [s]: 0.48, 0.96, 1.2, 2.4, 3.6, 6, 1...

  • Page 66: Level_Avg_Config?

    Command reference guide 60 srm-3006 narda level_avg_config? Description queries the current averaging method. The device has 1 averaging methods: • average over a time period (time) input level_avg_config?; output ,; parameter list parameter type description float time [s]: 0.48, 0.96, 1.2, 2.4, 3.6...

  • Page 67: Level_Avg_List?

    Command reference guide narda srm-3006 61 level_avg_list? Description queries the current possible averaging time settings for levelrecorder mode. The averaging method for levelrecorder mode is averaging over a time period. The output is a complete list of valid settings for this mode. Can be easily...

  • Page 68: Level_Config

    Command reference guide 62 srm-3006 narda level_config description sets the current measurement parameters for levelrecorder mode. Levelrecorder mode must be selected as the current operating mode. Input level_config ,,,,; output ; parameter list parameter type description double center frequency [h...

  • Page 69: Level_Config?

    Command reference guide narda srm-3006 63 level_config? Description queries the current measurement parameters for levelrecorder mode. Levelrecorder mode must be selected as the current operating mode. Input level_config?; output ,,,,,; parameter list parameter type description double center frequen...

  • Page 70: Livescreen?

    Command reference guide 64 srm-3006 narda livescreen? Description queries the current device display as a screenshot in png format. Individual values are themselves represented by binary numbers in the range 0-255 (0x00 – 0xff). The values are transmitted in ascii-hex. Example: the value 0x89 is tra...

  • Page 71: Lte?

    Command reference guide narda srm-3006 65 lte? The lte commands applies for the lte fdd optional mode description queries the different measurement runs (traces) that are available in lte mode. Lte must be selected as the current operating mode. Input lte? ; output ,,,, {,,,} , {,,,,, {,,,, {value,}...

  • Page 72

    Command reference guide 66 srm-3006 narda output parameter list parameter type description long number of sweeps performed short sweep time [ms] short 0 – 100 [%] progress of averaging short >= 0 short 0 4 503 short >= 1 enum yes, no short number of traces enum available traces: act actual value avg...

  • Page 73

    Command reference guide narda srm-3006 67 example (input) lte? Act; (output) 5,229,100,3, 335,2,yes, 334,2,yes, 333,2,yes, 1, act,yes,-62.13,unchecked,9, pss,yes,-72.78,3,-78.27,-74.23,-999.00, sss,yes,-73.35,3,-79.88,-74.44,-999.00, rs_avg,yes,-74.85,3,-80.51,-76.23,-999.00, rs_max,yes,-73.12,3,-78...

  • Page 74: Lte_Avg_Config

    Command reference guide 68 srm-3006 narda lte_avg_config the lte commands applies for the lte fdd optional mode description sets the current averaging method. The device has 2 averaging methods: • average over a number of individual measurements (number) • average over a time period (time) there is ...

  • Page 75: Lte_Avg_Config?

    Command reference guide narda srm-3006 69 lte_avg_config? The lte commands applies for the lte fdd optional mode description queries the current averaging method. The device has 2 averaging methods: • average over a number of individual measurements (number) • average over a time period (time) there...

  • Page 76: Lte_Avg_List?

    Command reference guide 70 srm-3006 narda lte_avg_list? The lte commands applies for the lte fdd optional mode description queries the current possible averaging settings for lte mode. The device has 2 averaging methods: • average over a number of individual measurements (number) • average over a ti...

  • Page 77: Lte_Config

    Command reference guide narda srm-3006 71 lte_config the lte commands applies for the lte fdd optional mode description sets the current measurement parameters for lte mode. Lte must be selected as the current operating mode. Input lte_config ,,,,,,; output ; parameter list parameter type descriptio...

  • Page 78: Lte_Config?

    Command reference guide 72 srm-3006 narda lte_config? The lte commands applies for the lte fdd optional mode description queries the current measurement parameters for lte mode. Lte must be selected as the current operating mode. A list of all cbws available can be queried with the rbw_list? Command...

  • Page 79: Lte_Reset_Table

    Command reference guide narda srm-3006 73 lte_reset_table the lte commands applies for the lte fdd optional mode description resets the table for lte results (cell ids). Input lte_reset_table; output ; parameter list parameter type description dword error code example (input) lte_reset_table; (outpu...

  • Page 80: Lte_Tdd?

    Command reference guide 74 srm-3006 narda lte_tdd? Description queries the different measurement runs (traces) that are available in lte tdd mode. Lte tdd must be selected as the current operating mode. Input lte_tdd? ; output ,,,, {,,,} , {,,,,, {,,,, {value,} } } ; input parameter list parameter t...

  • Page 81

    Command reference guide narda srm-3006 75 output parameter list parameter type description long number of sweeps performed short sweep time [ms] short 0 – 100 [%] progress of averaging short >= 0 short 0 4 503 short >= 1 enum yes, no short number of traces enum available traces: act actual value avg...

  • Page 82

    Command reference guide 76 srm-3006 narda example (input) lte_tdd? Act; (output) 5,229,100,3, 335,2,yes, 334,2,yes, 333,2,yes, 1, act,yes,-62.13,unchecked,9, pss,yes,-72.78,3,-78.27,-74.23,-999.00, sss,yes,-73.35,3,-79.88,-74.44,-999.00, rs_avg,yes,-74.85,3,-80.51,-76.23,-999.00, rs_max,yes,-73.12,3...

  • Page 83: Lte_Tdd_Avg_Config

    Command reference guide narda srm-3006 77 lte_tdd_avg_config description sets the current averaging method. The device has 2 averaging methods: • average over a number of individual measurements (number) • average over a time period (time) there is one setting parameter for each method, which is sel...

  • Page 84: Lte_Tdd_Avg_Config?

    Command reference guide 78 srm-3006 narda lte_tdd_avg_config? Description queries the current averaging method. The device has 2 averaging methods: • average over a number of individual measurements (number) • average over a time period (time) there is one setting parameter for each method, which is...

  • Page 85: Lte_Tdd_Avg_List?

    Command reference guide narda srm-3006 79 lte_tdd_avg_list? Description queries the current possible averaging settings for lte tdd mode. The device has 2 averaging methods: • average over a number of individual measurements (number) • average over a time period (time) the output is a complete list ...

  • Page 86: Lte_Tdd_Config

    Command reference guide 80 srm-3006 narda lte_tdd_config description set the current measurement parameters for lte tdd mode. Lte tdd must be selected as the current operating mode. Input lte_tdd_config ,,,,,, factor>,; output ; parameter list parameter type description double center frequency [hz] ...

  • Page 87: Lte_Tdd_Config?

    Command reference guide narda srm-3006 81 lte_tdd_config? Description query the current measurement parameters for lte tdd mode. Lte tdd must be selected as the current operating mode. A list of all cbws available can be queried with the rbw_list? Command. Input lte_tdd_config?; output ,,,,,, factor...

  • Page 88: Lte_Tdd_Reset_Table

    Command reference guide 82 srm-3006 narda lte_tdd_reset_table description resets the table for lte tdd results (cell ids). Input lte_tdd_reset_table; output ; parameter list parameter type description dword error code example (input) lte_tdd_reset_table; (output) 0;.

  • Page 89: Meas_Start

    Command reference guide narda srm-3006 83 meas_start description starts the cyclical measurement in the current operating mode. A started measurement is recognized by a query, e.G. Spectrum? And observation of the sweepcounter, which increments when the measurement is started. Input meas_start; outp...

  • Page 90: Meas_Stop

    Command reference guide 84 srm-3006 narda meas_stop description stops the cyclical measurement in the current operating mode. A stopped measurement is recognized by a query, e.G. Spectrum? And observation of the sweepcounter, which stops incrementing when the measurement is stopped. Input meas_stop;...

  • Page 91: Mode

    Command reference guide narda srm-3006 85 mode description sets the current operating mode. Input mode ; output ; parameter list parameter type description enum spectrum safety umts scope level lte lte_tdd dword error code example (input) mode spectrum; (output) 0;.

  • Page 92: Mode?

    Command reference guide 86 srm-3006 narda mode? Description queries the current operating mode. Input mode?; output ,, parameter list parameter type description enum spectrum safety umts scope level lte lte_tdd dword error code example (input) mode?; (output) spectrum,0;.

  • Page 93: Mr_List?

    Command reference guide narda srm-3006 87 mr_list? Description queries the current possible reference level settings (measurement range). The requested list reflects all the possible settings and guarantees consistency with the other device settings. The parameter can be used directly in the applica...

  • Page 94: Prog_Exit

    Command reference guide 88 srm-3006 narda prog_exit description ends the measurement program and switches the device to standby mode. Input prog_exit; output ; parameter list parameter type description dword error code example (input) prog_exit; (output) 0;.

  • Page 95: Rbw_List?

    Command reference guide narda srm-3006 89 rbw_list? Description queries the current possible rbw settings. The requested list reflects all the possible settings and guarantees consistency with the other device settings. The parameter can be used directly in the application to produce the same appear...

  • Page 96: Remote

    Command reference guide 90 srm-3006 narda remote description activates / deactivates remote communication with the srm. “remote on” is the condition required for utilizing remote access through all device interfaces. Note command also useable in “remote off” mode input: remote ; output ; parameter l...

  • Page 97: Remote?

    Command reference guide narda srm-3006 91 remote? Description checks communication with the srm, and queries readiness for remote operation. “remote on” is the condition required for utilizing remote access through all device interfaces. Note command also useable in “remote off” mode input remote?; ...

  • Page 98: Result_Type_Reset

    Command reference guide 92 srm-3006 narda result_type_reset description resets all measurement runs (traces) to default values as is done at the start of the measurement. Input result_type_reset; output ; parameter list parameter type description dword error code example (input) result_type_reset; (...

  • Page 99: Result_Type_Reset_Min_Max

    Command reference guide narda srm-3006 93 result_type_reset_min_max description resets the min and max measurement runs (traces) to default values as is done at the start of the measurement. Input result_type_reset_min_max; output ; parameter list parameter type description dword error code example ...

  • Page 100: Safety?

    Command reference guide 94 srm-3006 narda safety? Description queries one or more measurement runs (traces) in safetyevaluation mode. A measurement run is the number of all the measurement values determined during a measurement cycle (sweep). Safetyevaluation must be selected as the current operatin...

  • Page 101

    Command reference guide narda srm-3006 95 parameter list parameter type description enum sets the desired traces: act actual value avg averaged value max maximum value max_avg maximum of the averaged values min minimum value min_avg minimum of the averaged values std values of the selected standard ...

  • Page 102

    Command reference guide 96 srm-3006 narda string name of the channel (service) double resolution bandwidth [hz] being used for this channel (service) double lower frequency of the channel (service) [hz] double upper frequency of the channel (service) [hz] dword error code example (input) safety? Act...

  • Page 103: Safety_Avg_Config

    Command reference guide narda srm-3006 97 safety_avg_config description sets the current averaging method. The device has 2 averaging methods: • average over a number of individual measurements (number) • average over a time period (time) there is one setting parameter for each method, which is sele...

  • Page 104: Safety_Avg_Config?

    Command reference guide 98 srm-3006 narda safety_avg_config? Description queries the current averaging method. The device has 2 averaging methods: • average over a number of individual measurements (number) • average over a time period (time) there is one setting parameter for each method, which is ...

  • Page 105: Safety_Avg_List?

    Command reference guide narda srm-3006 99 safety_avg_list? Description queries the current possible averaging settings for safetyevaluation mode. The device has 2 averaging methods: • average over a number of individual measurements (number) • average over a time period (time) the output is a comple...

  • Page 106: Safety_Config

    Command reference guide 100 srm-3006 narda safety_config description sets the current measurement parameters for safetyevaluation mode. Safetyevaluation must be selected as the current operating mode. Input safety_config? ,,,; output ; parameter list parameter type description enum on the power leve...

  • Page 107: Safety_Config?

    Command reference guide narda srm-3006 101 safety_config? Description queries the current measurement parameters for safetyevaluation mode. Safetyevaluation must be selected as the current operating mode. Input safety_config?; output ,,,,; parameter list parameter type description enum on the power ...

  • Page 108: Scr_Clr

    Command reference guide 102 srm-3006 narda scr_clr description deletes a selected screenshot. Input scr_clr ; output ; parameter list parameter type description integer index of the screen shot dword error code example (input) scr_clr 3; (output) 0;.

  • Page 109: Scr_Clr_All

    Command reference guide narda srm-3006 103 scr_clr_all description deletes all screenshots. Input scr_clr_all; output ; parameter list parameter type description dword error code example (input) scr_clr_all; (output) 0;.

  • Page 110: Scr_Data?

    Command reference guide 104 srm-3006 narda scr_data? Description queries a selected screenshot. Individual values are themselves represented by binary numbers in the range 0-255 (0x00 – 0xff). The values are transmitted in ascii-hex. Example: the value 0x89 is transmitted as ‘8’ ‘9’, and the associa...

  • Page 111: Scr_Info?

    Command reference guide narda srm-3006 105 scr_info? Description queries the additional information for a screenshot. Input scr_info? ; output ,,,,; parameter list parameter type description integer index of the screen shot enum spectrum, safety, umts, scope, level date dd.Mm.Yy time hh:mm:ss string...

  • Page 112: Scr_Number?

    Command reference guide 106 srm-3006 narda scr_number? Description queries the number of existing screenshots. Input scr_number?; output ,; parameter list parameter type description integer number of screen shots available dword error code example (input) scr_number?; (output) 6,0;.

  • Page 113: Send_Key

    Command reference guide narda srm-3006 107 send_key description activates an operating panel key by remote communication. Note command also useable in “remote off” mode input send_key ; output ; parameter list parameter type description enum hsk1 horizontal soft key 1, (hsk1 4 hsk6, from left to rig...

  • Page 114: Send_Rot_Knob

    Command reference guide 108 srm-3006 narda send_rot_knob description activates the operating panel rotary control by remote communication. Note command also useable in “remote off” mode input send_rot_knob ; output ; parameter list parameter type description integer counter-clockwise rotation decrem...

  • Page 115: Spectrum?

    Command reference guide narda srm-3006 109 spectrum? Description queries one or more measurement runs (traces) in spectrum mode. A measurement run is the number of all the measurement values determined during a measurement cycle (sweep). Spectrum must be selected as the current operating mode. Input...

  • Page 116

    Command reference guide 110 srm-3006 narda example (input) spectrum? Act; (output) 397,27,100,0,993282300,52083.3333333,1, act,no,21, -12.26127,-12.55294,-11.70693,-11.97045,-15.70837,-18.4338,-16.36422,-14.76947, -15.36936,-14.26438,-14.78028,-16.47095,-15.76123,-12.88897,-11.72068,-12.01601, -12.8...

  • Page 117: Spectrum_Avg_Config

    Command reference guide narda srm-3006 111 spectrum_avg_config description sets the current averaging method. The device has 2 averaging methods: • average over a number of individual measurements (number) • average over a time period (time) there is one setting parameter for each method, which is s...

  • Page 118: Spectrum _Avg_Config?

    Command reference guide 112 srm-3006 narda spectrum _avg_config? Description queries the current averaging method. The device has 2 averaging methods: • average over a number of individual measurements (number) • average over a time period (time) there is one setting parameter for each method, which...

  • Page 119: Spectrum_Avg_List?

    Command reference guide narda srm-3006 113 spectrum_avg_list? Description queries the current possible averaging settings for spectrum mode. The device has 2 averaging methods: • average over a number of individual measurements (number) • average over a time period (time) the output is a complete li...

  • Page 120: Spectrum_Bi_Value?

    Command reference guide 114 srm-3006 narda spectrum_bi_value? Description band integration over frequency. This command is only available in spectrum mode. The spectral power level is integrated over a selected frequency band. The following conditions apply: • the band limits must lie within the ran...

  • Page 121: Spectrum_Config

    Command reference guide narda srm-3006 115 spectrum_config description sets the current measurement parameters for spectrum mode. Spectrum must be selected as the current operating mode. Input spectrum_config ,,,,,; output ; parameter list parameter type description double center frequency [hz] doub...

  • Page 122: Spectrum_Config?

    Command reference guide 116 srm-3006 narda spectrum_config? Description queries the current measurement parameters for spectrum mode. Spectrum must be selected as the current operating mode. Input spectrum_config?; output ,,,,,,; parameter list parameter type description double center frequency [hz]...

  • Page 123: Spectrum_Mrk_Highest?

    Command reference guide narda srm-3006 117 spectrum_mrk_highest? Description queries the marker values for the highest measurement value. This command is only available in spectrum mode. Input spectrum_mrk_highest? ; output ,,; parameter list parameter type description enum act, avg, max, max_avg, m...

  • Page 124: Spectrum_Mrk_Idx_Value?

    Command reference guide 118 srm-3006 narda spectrum_mrk_idx_value? Description queries the marker values for a selected marker. This command is only available in spectrum mode. The marker is selected by means of an index. The index describes a bin (single measured value) in the spectrum. Up to 27,51...

  • Page 125: Spectrum_Mrk_Value?

    Command reference guide narda srm-3006 119 spectrum_mrk_value? Description queries the marker values for a specified frequency point. This command is only available in spectrum mode. The discrete resolution of the spectrum means that the marker closest to the specified frequency point will be select...

  • Page 126: Spectrum_Pkt_Table?

    Command reference guide 120 srm-3006 narda spectrum_pkt_table? Description queries a configurable peak table. This command is only available in spectrum mode. Input spectrum_pkt_table? ,,,; output ,{,,}; parameter list parameter type description enum sets the maximum number of peaks for the peak tab...

  • Page 127: Srv_Clr

    Command reference guide narda srm-3006 121 srv_clr description deletes a service table. Input srv_clr ; output ; parameter list parameter type description string long name of the channel (service) table, up to 35 chr dword error code example (input) srv_clr "china_overview"; (output) 0;.

  • Page 128: Srv_Clr_All

    Command reference guide 122 srm-3006 narda srv_clr_all description deletes all service tables. Input srv_clr_all; output ; parameter list parameter type description dword error code example (input) srv_clr_all; (output) 0;.

  • Page 129: Srv_Ds

    Command reference guide narda srm-3006 123 srv_ds description creates a service table. Input srv_ds ,, ,{,,,,}; output ; parameter list parameter type description string short name of the channel (service) table, up to 15 chr string long name of the channel (service) table, up to 35 chr short number...

  • Page 130: Srv_Ds?

    Command reference guide 124 srm-3006 narda srv_ds? Description queries a service table. Input srv_ds? ; output ,, ,{,,,,}; parameter list parameter type description string requested long name of the channel (service) table, up to 35 chr string short name of the channel (service) table, up to 15 chr ...

  • Page 131: Srv_List?

    Command reference guide narda srm-3006 125 srv_list? Description queries all service tables without details. Input srv_list?; output ,{,}; parameter list parameter type description short total number of channel (service) tables available string long name of the channel (service) table, up to 35 chr ...

  • Page 132: Srv_Sel

    Command reference guide 126 srm-3006 narda srv_sel description sets the current active service table. Input srv_sel ; output ; parameter list parameter type description string long name of the channel (service) table, up to 35 chr dword error code example (input) srv_sel "my_longname"; (output) 0;.

  • Page 133: Srv_Sel?

    Command reference guide narda srm-3006 127 srv_sel? Description queries the current active service table. Input srv_sel?; output ,,; parameter list parameter type description string long name of the channel (service) table, up to 35 chr string short name of the channel (service) table, up to 15 chr ...

  • Page 134: Su_Clr

    Command reference guide 128 srm-3006 narda su_clr description deletes a selected device setup. Input su_clr ; output ; parameter list parameter type description string name of the setup, up to 35 chr dword error code example (input) su_clr "aaa"; (output) 0;.

  • Page 135: Su_Clr_All

    Command reference guide narda srm-3006 129 su_clr_all description deletes all device setups. Input su_clr_all; output ; parameter list parameter type description dword error code example (input) su_clr_all; (output) 0;.

  • Page 136: Su_Default

    Command reference guide 130 srm-3006 narda su_default description resets the device to the default (works) settings. This affects all parameters that can be changed by a device setup. This command is useful for setting the device to a defined state. Input su_default; output ; parameter list paramete...

  • Page 137: Su_Ds

    Command reference guide narda srm-3006 131 su_ds description transfers a setup file to the device. Setup files cannot be created or modified manually. They are always created by completely saving the current device settings on the device itself (su_save). For this reason, the contents of a setup fil...

  • Page 138: Su_Ds?

    Command reference guide 132 srm-3006 narda su_ds? Description reads a setup file out of the device. Setup files cannot be created or modified manually. They are always created by completely saving the current device settings on the device itself (su_save). For this reason, the contents of a setup fi...

  • Page 139: Su_List?

    Command reference guide narda srm-3006 133 su_list? Description lists the names of the setups stored in the device. Input su_list?; output ,{,}; parameter list parameter type description short number of setups available string name of the setup, up to 35 chr dword error code example (input) su_list?...

  • Page 140: Su_Recall

    Command reference guide 134 srm-3006 narda su_recall description activates a selected device setup. Input su_recall ; output ; parameter list parameter type description string name of the setup, up to 35 chr dword error code example (input) su_recal “3*28_bins”; (output) 0;.

  • Page 141: Su_Store

    Command reference guide narda srm-3006 135 su_store description saves the current device settings as a new setup in the device. A name that has not been used yet must be specified as the setup name. Input su_store ; output ; parameter list parameter type description string name of the setup, up to 3...

  • Page 142: Sweep_State?

    Command reference guide 136 srm-3006 narda sweep_state? Description queries the status and progress of the current measurement. The can be used to detect the start of a new measurement so that the measured values can be fetched only when required. Input sweep_state?; output ,,,,; parameter list para...

  • Page 143: Time

    Command reference guide narda srm-3006 137 time description sets the current system time of the device. Input time,; output ; parameter list parameter type description time hh:mm:ss dword error code example (input) time 15:16:17; (output) 0;.

  • Page 144: Time?

    Command reference guide 138 srm-3006 narda time? Description queries the current system time of the device. Input time?; output ,; parameter list parameter type description time hh:mm:ss dword error code example (input) time?; (output) 14:29:58,0;,0;.

  • Page 145: Umts?

    Command reference guide narda srm-3006 139 umts? Description queries the different measurement runs (traces) that are available in umts mode. Umts must be selected as the current operating mode. Input umts? ; output ,,,, ,,,,,, {,,}; parameter list parameter type description enum act actual value av...

  • Page 146

    Command reference guide 140 srm-3006 narda example (input) umts? Act; (output) 463,317,100,1, act,no,-84.3169,-76.30199,unchecked,3, 501,-85.56746,yes,339,-90.33401,yes,34,-999,yes, 0; example (input) umts? All; (output) 5464,298,100,6, act,no,-82.21515,-75.25188,unchecked,3, 501,-84.02314,yes,339,-...

  • Page 147: Umts_Avg_Config

    Command reference guide narda srm-3006 141 umts_avg_config description sets the current averaging method. The device has 2 averaging methods: • average over a number of individual measurements (number) • average over a time period (time) there is one setting parameter for each method, which is selec...

  • Page 148: Umts _Avg_Config?

    Command reference guide 142 srm-3006 narda umts _avg_config? Description queries the current averaging method. The device has 2 averaging methods: • average over a number of individual measurements (number) • average over a time period (time) there is one setting parameter for each method, which is ...

  • Page 149: Umts_Avg_List?

    Command reference guide narda srm-3006 143 umts_avg_list? Description queries the current possible averaging settings for umts mode. The device has 2 averaging methods: • average over a number of individual measurements (number) • average over a time period (time) the output is a complete list of va...

  • Page 150: Umts_Config

    Command reference guide 144 srm-3006 narda umts_config description sets the current measurement parameters for umts mode. Umts must be selected as the current operating mode. Input umts_config? ,,,; output ; parameter list parameter type description double center frequency [hz] enum on, off float 0....

  • Page 151: Umts_Config?

    Command reference guide narda srm-3006 145 umts_config? Description queries the current measurement parameters for umts mode. Umts must be selected as the current operating mode. Input umts_config?; output ,,,,; parameter list parameter type description double center frequency [hz] enum on, off floa...

  • Page 152: Umts_Reset_Table

    Command reference guide 146 srm-3006 narda umts_reset_table description resets the table for umts results (scrambling codes). This command is commonly used at the beginning of a new measurent. Input umts_reset_table; output ; parameter list parameter type description dword error code example (input)...

  • Page 153: Unit

    Command reference guide narda srm-3006 147 unit description sets the current unit setting. Input unit ; output ; parameter list parameter type description enum measurement results can be displayed in one of the following units: dbm, dbv, dbmv, dbuv, dbv/m, dbmv/m, dbuv/m, dba/m v/m, a/m, w/m^2, w/cm...

  • Page 154: Unit?

    Command reference guide 148 srm-3006 narda unit? Description queries the current unit setting. Input unit?; output ,error>; parameter list parameter type description enum measurement results can be displayed in one of the following units: dbm, dbv, dbmv, dbuv, dbv/m, dbmv/m, dbuv/m, dba/m v/m, a/m, ...

  • Page 155: Unit_List?

    Command reference guide narda srm-3006 149 unit_list? Description queries the current possible unit settings (unit). The requested list reflects all the possible settings and guarantees consistency with the other device settings. The parameter can be used directly in the application to produce the s...

  • Page 156: Vbw_List?

    Command reference guide 150 srm-3006 narda vbw_list? Description queries the current possible vbw settings. The requested list reflects all the possible settings and guarantees consistency with the other device settings. The parameter can be used directly in the application to produce the same appea...

  • Page 157: Version?

    Command reference guide narda srm-3006 151 version? Description queries the versions of the current firmware modules. Input version? ; output ,; parameter list parameter type description enum app application firmware bl bootloader os operating system versionstring “v0.0.0” to “v99.99.99” example (in...

  • Page 158

    Command reference guide 152 srm-3006 narda 4 remote application examples time synchronization with a pc the device system time can differ from the local time. In such cases, the device time should be synchronized with the pc clock to ensure that the measurements are time stamped correctly. The follo...

  • Page 159

    Command reference guide narda srm-3006 153 communications trace (input) remote on; (output) 0; (input) date?; (output) 01.01.10,0; (input) time?; (output) 12:00:00,0; (input) date 14.06.10; (output) 0; (input) time 15:31:00; (output) 0; (input) remote off; (output) 0;.

  • Page 160

    Command reference guide 154 srm-3006 narda synchronizing queries of measured values with the sweep state a large number of measured values can occur for certain device settings when a spectrum is queried. Synchronization with the device status (sweep state) is necessary to avoid reading the same mea...

  • Page 161

    Command reference guide narda srm-3006 155 communications trace (input) remote on; (output) 0; (input) mode spectrum; (output) 0; (input) sweep_state?; (output) 26,383,26,100,0; (input) sweep_state?; (output) 26,383,36,100,0; (input) sweep_state?; (output) 26,383,54,100,0; (input) sweep_state?; (out...

  • Page 162

    Command reference guide 156 srm-3006 narda carrier signal monitoring with alarms simultaneous monitoring of several carrier signals (services) is a typical application for multi-channel power (safety evaluation) mode. The services are defined in a so-called service table for this. An alarm is to be ...

  • Page 163

    Command reference guide narda srm-3006 157 communications trace (input) remote on; (output) 0; (input) su_recall “test“; (output) 0; (input) sweep_state?; (output) 1,433,100,25,0; (input) sweep_state?; (output) 1,433,100,25,0; (input) sweep_state?; (output) 2,421,100,50,0 (input) safety? Act; (outpu...

  • Page 166

    Narda safety test solutions gmbh sandwiesenstraße 7 72793 pfullingen, germany phone: +49 7121 9732 0 fax: +49 7121 9732 790 e-mail: support.Narda-de@l3t.Com www.Narda-sts.Com narda safety test solutions 435 moreland road hauppauge, ny 11788, usa phone: +1 631-231-1700 fax: +1 631-231-1711 e-mail: na...