NDS ZEROWIRE G2 User Manual

Manual is about: Advanced Wireless HD-Video Transmission System

Summary of ZEROWIRE G2

  • Page 1

    Zerowire® g2 advanced wireless hd-video transmission system user manual e n g l i s h.

  • Page 3

    © 2015 nds surgical imaging. All rights reserved. Information in this document has been carefully checked for accuracy; however, no guarantee is given to the correctness of the contents. This document is subject to change without notice. Nds provides this information as reference only. Reference to ...

  • Page 5

    I table of contents section 1 warnings and cautions ------------------------------------------------------------------------------ ii recycling ------------------------------------------------------------------------------------------------- ii declarations of conformity ----------------------------...

  • Page 6

    Ii 1 recycling: follow local governing ordinances and recycling plans regarding the recycling or disposal of this equipment. Safety compliance: this device meets the requirements of en60601-1 so as to conform to the medical device directive 93/42/eec and 2007/47/ec (general safety information). Safe...

  • Page 7

    Iii declarations of conformity legal statement fcc and council directives of european standards: this device complies with part 15 of fcc rules and 93/42/eec and 2007/47/ec of the council directives of european standards. Operation is subject to the following two conditions: (1) this device may not ...

  • Page 9

    1 2 about this manual this manual is designed to assist the user with installation, setup and operation of the zerowire® g2 wireless video system. A numbered tab on the side of the page denotes the beginning of a section. The functional descriptions in this manual are representative of: zerowire g2 ...

  • Page 10

    2 overview (continued) zerowire g2 specifications wireless signal type 60 ghz wireless hd band (wihd) frequency band 57 - 64 ghz video compression none video inputs (tx) dvi-d, 3g-sdi video outputs (rx) dvi-d 3d video support dvi-d interleaved (line-by-line) hipaa support 256 bit aes encryption syst...

  • Page 11

    3 3 connector panels usb ports are used for installing updates to the transmitter or receiver firmware, and not for use as a general purpose i/o port. 3g - sdi input power connector dvi input transmitter link and status leds receiver link and status leds the link button is used to link a transmitter...

  • Page 12

    4 installation mounting brackets: refer to the zerowire g2 accessory kits by display table on page 18. Installation: (for mounted displays) note: this installation requires two people: one to support the display, and one to perform the installation. Replace the left-side mounting screws with two of ...

  • Page 13

    5 with the mounting blade facing the front of the display, slide the step bracket between the vesa mount and the back of the display until notches labeled for the display you are working with fit over the two left-hand screws. Replace the right-hand screws with the provided screws and tighten all sc...

  • Page 14

    6 zerowire g2 transmitters and receivers used with a supported 32" or smaller nds display can use an optional ‘y’ adapter cable for powering the zerowire unit. The two types of ‘y’ adapter cables are shown below. Where applicable, the appropriate y-cable is included in the accessory kit (see page 18...

  • Page 15

    7 wiring diagrams cable bend radius we recommend that the bend radius of metallic cables be no less than 2.5 inches (63 mm) or 7 times the diameter of the cable whichever is greater. Sharper bends may damage the cable and/or degrade the video signal. Zerowire g2 receiver wiring black wires are 24 vd...

  • Page 16

    8 note: the transmitter and receiver are not limited to line-of-sight operation, i.E., the pair will operate reliably despite the presence of intervening obstacles. However, neither unit should be completely surrounded by metallic objects. Passing video: with the linking completed, the status leds s...

  • Page 17

    9 for mission critical applications, we strongly recommend that a replacement zerowire g2 transmitter and receiver pair, and dvi cable, be immediately available. Additionally, we recommend that a display that is hard wired to the video source be immediately available whenever a surgical procedure is...

  • Page 18

    10 positioning and orientation due to the shape of the antenna’s signal field, both transmitter and receiver units should be installed so they are aligned vertically and horizontally within ±10 ° . Vertical alignment: vertical alignment should be within ±10 ° of vertical. Horizontal alignment: horiz...

  • Page 19

    11 elevation: the transmitter and receiver should be positioned at least 5 feet (1.5 m) above the floor, be at the same height, and, preferably, with the transmitter and receiver front edges parallel and facing each other. Separation: zerowire g2 operates correctly with a transmitter and receiver se...

  • Page 20

    12 non line of sight operation the zerowire g2 system can maintain a wireless link with partial obstructions. However, best performance is realized with a clear line-of-sight between units. In an or environment, the ceiling mounted boom system and its spring arm junctions, along with surgical light ...

  • Page 21

    13 channel usage the recommendations listed below should be used to determine the appropriate setting. 1. Zerowire g2 transmitter and/or receiver units cannot be used with earlier zerowire transmitter and/ or receiver units. 2. A maximum of two zerowire g2 systems may be operated in the same room. T...

  • Page 22

    14 when a receiver is connected to a monitor, self-explanatory diagnostic messages are displayed in the lower left corner of the screen to inform the user of the receiver’s current status. The messages will be displayed for 15 seconds, with the exception of “no transmitter found” and “signal lost.” ...

  • Page 23

    15 4 troubleshooting status led indicators led state description blue and blinking slowly the unit is scanning for a channel. Blue and flashing rapidly unit is attempting to link. Blue and steady the unit is sending (transmitter) or receiving (receiver) video data. Blue alternating on/off for 3 seco...

  • Page 24

    16 specifications *the effective range may vary depending upon the environment in which the product * is operating. Follow the product’s setup section to achieve optimum performance and * range. 5 wireless signal type 60 ghz wireless hd band (wihd) frequency band 57 - 64 ghz video inputs (tx) dvi-d,...

  • Page 25

    17 supported video modes format frame rate (hz) horizontal active pixels vertical active pixels horizontal total pixels vertical total pixels 1080p 60 1920 1080 2200 1125 1080p 50 1920 1080 2640 1125 1080i 30 1920 1080 2200 562 1080i 25 1920 1080 2640 562 720p 60 1280 720 1650 750 720p 50 1280 720 1...

  • Page 26

    18 zerowire g2 accessory kits by display zerowire g2 accessory kit display nds part # pair: (transmiƩer and receiver) single: (transmiƩer or receiver) endovue 24" (standard) 90k0010 90z0152 90z0153 china model 90k0011 disconƟnued model 90k0004 radiance full mmi 24" (standard) 90r0019 90z0152 90z0153...

  • Page 27

    19 cleaning instructions prior to cleaning, the unit should be turned off and disconnected from its power source. Cleaning: thoroughly wipe all exterior surfaces with a lint-free cloth that has been dampened with an acceptable cleaning agent. Acceptable cleaning materials are listed below. Remove re...

  • Page 28

    20 6 zerowire g2 has been tested in an or environment and did not interfere with electrosurgical, ultrasound, x-ray, medical displays and life monitoring equipment. Neither did the aforementioned equipment interfere with zerowire g2. Additionally, since zerowire g2 operates in 57 – 64 ghz, typically...

  • Page 29

    21 emc tables guidance and manufacturer’s declaration – electromagnetic emissions the product is intended for use in the electromagnetic environment specified below. The customer or the user of the product should assure that it is used in such an environment. Emissions compliance electromagnetic env...

  • Page 30

    22 emc tables guidance and manufacturer’s declaration 211; electromagnetic immunity the product is intended for use in the electromagnetic environment specified below. The customer or the user of the product should assure that it is used in such an environment. Immunity test iec 60601 test level com...

  • Page 31

    23 emc tables guidance and manufacturer’s declaration – electromagnetic immunity the product is intended for use in the electromagnetic environment specified below. The customer or the user of the product should assure that it is used in such an environment. Immunity test iec 60601 test level compli...

  • Page 32

    24 emc tables recommended separation distances between portable and mobile rf communications equipment and the product the product is intended for use in an electromagnetic environment in which radiated rf disturbances are controlled. The customer or the user of the product can help prevent electrom...

  • Page 34

    Asia pacific japan +81 3 5753 2466 info@ndssi.Jp europe the netherlands + 31 180 63 43 56 info-emea@ndssi.Com corporate headquarters san jose, ca usa 408 776 0085 info@ndssi.Com 60 g0571 rev b 10 november 2015.