Novametrix Medical Systems 8100 User Manual - Thank You . . .

Manual is about: CO2SMO Plus, respiratory profile monitor

Summary of 8100

  • Page 1

    Novametrix medical systems inc. Po box 690 5 technology drive wallingford, connecticut, u.S.A. 06492 respiratory profile monitor user’s manual model 8100 october 10, 1997 catalog no. 6758-23-02.

  • Page 3: Revision History

    Rev. 02 co 2 smo plus user’s manual iii revision history 08-jan-97 release, rev 00 17-jan-97 production release, rev 01 10-oct-97 revision 02 software, rev 02 co 2 smo plus! And capnostat are registered trademarks and y-sensor, superbright and oxysnap are trade- marks of novametrix medical systems i...

  • Page 4: Thank You . . .

    Iv co 2 smo plus user’s manual rev. 02 thank you . . . . . . For selecting the co 2 smo plus! Respiratory profile monitor. You are the operator of one of the most advanced, state-of-the-art respiratory instruments available today. Our engineering staff, in collaboration with the medical community, h...

  • Page 5: Contents

    Rev. 01 co 2 smo plus user’s manual v contents general description ............................................................................................................ 1 indication for use, unpacking, operational overview of the monitor, and principles of operation. Safety .....................

  • Page 6

    Vi co 2 smo plus user’s manual rev. 01.

  • Page 7: Section 1

    Rev. 02 co 2 smo plus user’s manual 1 section 1 general description through the combination of capnography, airway flow and pressure, and pulse oximetry, the co 2 smo plus! Provides the clinician with a respiratory profile of the patient. This includes information that was previously difficult to ob...

  • Page 8: Operational Overview

    Section 1 operational overview 2 co 2 smo plus user’s manual rev. 02 operational overview parameters measured the co 2 smo plus! Measures and calculates various respiratory parameters for patient/ventilator management. Refer to “monitoring” on page 47 for a complete parameters list. Front panel cont...

  • Page 9

    Operational overview general description rev. 02 co 2 smo plus user’s manual 3 if the or key is not pressed within five seconds, trends will be retained. The monitor is now ready for operation. Contrast press or hold to adjust the display’s contrast for different up/down viewing angles. Alert silenc...

  • Page 10

    Section 1 operational overview 4 co 2 smo plus user’s manual rev. 02 data screen press to switch between data screens (numeric values only), and data entry screen (to enter weight and paco 2 - see “data screens” on page 56 for details). Setup/configuration press to display the setup screen, or hold ...

  • Page 11

    Operational overview general description rev. 02 co 2 smo plus user’s manual 5 co 2 /pl e th fl o w /p r e s s t and a rd d a t a s cree n s tr en d - c hanges ti me base - m o ves point e r - pr ess t o change value p r o t ocol d a t a scre ens note: s creens m a y vary wit h fut u re so ft ware u...

  • Page 12

    Section 1 operational overview 6 co 2 smo plus user’s manual rev. 02 indicators battery alert icon illuminates when the unit is on battery power and the battery charge is critically low. When battery is near exhaustion an audible alert will sound and “battery very low plug in ac power” will appear i...

  • Page 13

    Operational overview general description rev. 02 co 2 smo plus user’s manual 7 alert bar flashes red when alert conditions are detected. When alerts latched in the system configuration is set to on, the alert bar is latched. This means that when an alert is active, the alert bar will flash even if t...

  • Page 14

    Section 1 rear panel connections and labeling 8 co 2 smo plus user’s manual rev. 02 rear panel connections and labeling equipotentiality: connection to monitor’s chassis patient isolation: identifies connection as type bf attention: consult manual for detailed information mains fuse rating for repla...

  • Page 15: Principle Of Operation

    Principle of operation general description rev. 02 co 2 smo plus user’s manual 9 principle of operation co 2 co 2 smo plus! Uses the capnostat co 2 sensor to measure co 2 by using the infrared absorbtion technique, which has endured and evolved in the clinical setting for over two decades and remain...

  • Page 16

    Section 1 principle of operation 10 co 2 smo plus user’s manual rev. 02 pulse rate is calculated by measuring the time interval between peaks of the infrared light waveform. The inverse of this measurement is displayed as pulse rate. The oxygen saturation and pulse rate values are updated once each ...

  • Page 17: Section 2

    Rev. 02 co 2 smo plus user’s manual 11 section 2 safety the co 2 smo plus! Monitor is electrically isolated. Patient leakage current flowing from the instrument to ground is limited to less than 100 ua at 120/220 vac, 50-60hz. For maximum patient and operator safety, you must follow the following wa...

  • Page 18

    Section 2 12 co 2 smo plus user’s manual rev. 02 • periodically check the flow sensor and tubing for excessive moisture or secretion build up. Although the co 2 smo plus! Automatically purges the lines, excessive moisture or secretions may still remain. • while using the flow sensor, a system leak, ...

  • Page 19

    Safety rev. 02 co 2 smo plus user’s manual 13 • when a new capnostat co 2 sensor is attached to the monitor, or is moved from one monitor to another, it must be initialized before use. The capnostat co 2 sensor does not have to be initialized again as long as it is used with the same monitor. • if y...

  • Page 20

    Section 2 14 co 2 smo plus user’s manual rev. 02 [this page intentionally blank.].

  • Page 21: Section 3

    Rev. 02 co 2 smo plus user’s manual 15 section 3 preparation for use ac/battery operation press the power key to place the unit into operate mode (on) or to place the unit into standby (off). The status of the unit is dependent upon both the key and the power source: • no ac power, power icon off, n...

  • Page 22

    Section 3 setup and configuration settings 16 co 2 smo plus user’s manual rev. 02 battery is not sufficiently charged. Allow the battery to charge for 12 hours before initial use; refer to the following diagram for charging instructions. To charge the battery, connect the line cord to an ac source a...

  • Page 23

    Setup and configuration settings preparation for use rev. 02 co 2 smo plus user’s manual 17 setup options are described below: gas comp: use this setting to enter the gas composition in order to compensate the flow measurement for gas density. Any setting other than the default (0% agent, balance ai...

  • Page 24

    Section 3 setup and configuration settings 18 co 2 smo plus user’s manual rev. 02 assumed inspired concentrations based on selected expired concentrations: • inspired co 2 =0 • inspired o 2 =expired setting + 4 volume % • inspired n 2 =expired setting • inspired n 2 o=expired setting + 1 volume % • ...

  • Page 25

    Setup and configuration settings preparation for use rev. 02 co 2 smo plus user’s manual 19 with ventilators that have a prvc (pressure regulated volume control) mode, the spontaneous threshold setting will have to be adjusted. This mode combines mechanical ventilation with pressure support for spon...

  • Page 26

    Section 3 setup and configuration settings 20 co 2 smo plus user’s manual rev. 02 co 2 units: select the desired units for both the capnogram and etco 2 values. Co 2 avg (co 2 averaging): select the interval from which the displayed value of end tidal co 2 (etco 2 ) is calculated. Vco 2 avg (vco 2 a...

  • Page 27

    Setup and configuration settings preparation for use rev. 02 co 2 smo plus user’s manual 21 alert delay: when set to “on” the monitor will delay the audible alert for 10 seconds upon the detection of an alert violation. Alerts latched: the monitor will indicate an alert condition once detected. If t...

  • Page 28

    Section 3 setup and configuration settings 22 co 2 smo plus user’s manual rev. 02 [this page intentionally blank.].

  • Page 29: Section 4

    Rev. 02 co 2 smo plus! User’s manual 23 section 4 sensors and patient connections the co 2 smo plus! Calculates various flow, co 2 and pulse oximetry related parameters. Flow, co 2 and pulse oximetry sensors can be used simultaneously or in any combination. Initializing the capnostat co 2 sensor whe...

  • Page 30

    Section 4 24 co 2 smo plus! User’s manual rev. 02 3. The monitor sounds a short “beep” and will display “sensor passed.” after a few seconds, the message “put co2 sensor on adapter in room air, ok: cancel: ” will display. The capnostat co 2 sensor has been initialized but is not ready for patient us...

  • Page 31

    Sensors and patient connections rev. 02 co 2 smo plus! User’s manual 25 3. The monitor will respond “zeroing co2 on airway adapt: xx” or “zeroing co2 on cannula:xx” if the sampling pump is on (xx = seconds remaining). On-cable verifier cells the on-cable verifier cells consist of a zero (-0-) and a ...

  • Page 32: Combined Co

    Section 4 combined co 2 and flow 26 co 2 smo plus! User’s manual rev. 02 combined co 2 and flow to monitor flow (respiratory mechanics) and co 2 , a combined co 2 airway adapter and flow sensor (co 2 /flow sensor) is available. 1. Select a series 3 co 2 /flow sensor based on the patient: 2. Connect ...

  • Page 33

    Combined co 2 and flow sensors and patient connections rev. 02 co 2 smo plus! User’s manual 27 3. Connect the capnostat co 2 sensor to the combined co 2 /flow sensor. 4. Connect the combined co 2 /flow sensor to the patient breathing circuit. • do not place the airway adapter between the et tube and...

  • Page 34

    Section 4 monitoring flow (respiratory mechanics) separately 28 co 2 smo plus! User’s manual rev. 02 • for routine performance of airway care, separate the system between the et tube and the airway adapter (neonatal circuit), or between the et tube and elbow (pediatric/adult circuit). Lavage and suc...

  • Page 35: Monitoring Co

    Monitoring co 2 separately sensors and patient connections rev. 02 co 2 smo plus! User’s manual 29 monitoring co 2 separately to monitor co 2 without flow (respiratory mechanics): 1. Select an airway adapter based on the patient and monitoring situation. 2. Attach the capnostat co 2 sensor to the mo...

  • Page 36

    Section 4 monitoring co 2 separately 30 co 2 smo plus! User’s manual rev. 02 sampling adapter the sampling airway adapter with tubing (catalog no. 5843) should be used in conjunction with a nasal sampling cannula to monitor non-intubated patients. 1. Verify that the windows are clean and dry. Clean ...

  • Page 37

    Monitoring co 2 separately sensors and patient connections rev. 02 co 2 smo plus! User’s manual 31 7. Adapter zero is necessary when the pump is first turned on to set the sampling system alert thresholds and adjust to the optical characteristics of the sampling adapter. After the first sampling ada...

  • Page 38: Flow Calibration/zeroing

    Section 4 flow calibration/zeroing 32 co 2 smo plus! User’s manual rev. 02 flow calibration/zeroing flow sensor calibration is not necessary due to the consistency from one flow sensor to the next. Also, the flow signal is automatically zeroed periodically by internal valves. The flow zero will free...

  • Page 39: Spo

    Spo 2 sensors sensors and patient connections rev. 02 co 2 smo plus! User’s manual 33 spo 2 sensors this section explains how to select an spo 2 sensor, how to connect the sensor to the monitor, and how to apply the sensor to the patient. To attach a superbright sensor or sensor extension cable to c...

  • Page 40

    Section 4 spo 2 sensors 34 co 2 smo plus! User’s manual rev. 02 oxysnap ™ connectors to connect an oxysnap extension cable to an oxysnap superbright sensor: 1. Align the arrows on the oxysnap connectors and press the connectors together. 2. To disconnect, grasp the connectors at the finger grips and...

  • Page 41

    Spo 2 sensors sensors and patient connections rev. 02 co 2 smo plus! User’s manual 35 3. Release the finger grips. 4. To remove sensor, gently squeeze grips and slide the sensor from the finger. Finger sensor quick check 1. Is “probe off” displayed when the sensor is connected to the monitor but not...

  • Page 42

    Section 4 spo 2 sensors 36 co 2 smo plus! User’s manual rev. 02 y-sensor application using y-strip tapes or foam wrap select a y-strip or foam wrap based on the patient type and intended sensor location. Wrap style tapes: y-strip tapes are available in two color coded sizes: 25 mm tapes have green l...

  • Page 43

    Spo 2 sensors sensors and patient connections rev. 02 co 2 smo plus! User’s manual 37 to use the y-strip tapes: 1. Remove the portion of the release liner containing the holes. 2. (omit this step if using the 25 mm y-strip tape.) if using the 20 mm y-strip tape, carefully remove the y-sensor center ...

  • Page 44

    Section 4 spo 2 sensors 38 co 2 smo plus! User’s manual rev. 02 leave slack in the wires between the tape and the sensor. To use the non-adhesive foam wrap: 1. With the blue side of the foam wrap facing up, press the buttons on the back of each y-sensor head through the holes in the foam wrap. The w...

  • Page 45

    Spo 2 sensors sensors and patient connections rev. 02 co 2 smo plus! User’s manual 39 2. Face the blue side of the wrap toward the skin and wrap around the site (cut excess foam if necessary on small version). Secure with the velcro ® tab. 3. Ensure the sensor heads are directly opposite each other ...

  • Page 46

    Section 4 spo 2 sensors 40 co 2 smo plus! User’s manual rev. 02 y-sensor application using ear clip 1. Remove center strip from the y-sensor. 2. Slide each y-sensor head into the ear clip‘s receptacles. The heads should face each other. 3. Gently squeeze the end of the ear clip (shown in diagram), a...

  • Page 47

    Spo 2 sensors sensors and patient connections rev. 02 co 2 smo plus! User’s manual 41 if a satisfactory reading cannot be obtained, rub the site and/or use adhesive dots for better response. The adhesive dots (catalog no. 8700) included with the ear clips will also help in preventing the ear clip fr...

  • Page 48

    Section 4 spo 2 sensors 42 co 2 smo plus! User’s manual rev. 02 single patient use spo 2 sensors select a y-strip or foam wrap based on the patient type and intended sensor location. Single patient use pediatric/adult sensor (catalog no. 6455): • the single patient use spo 2 sensor can be used when ...

  • Page 49

    Spo 2 sensors sensors and patient connections rev. 02 co 2 smo plus! User’s manual 43 single patient use spo 2 sensor application 1. Connect the db-9 extension cable to the co 2 smo plus! Rear panel connector. 2. Press the connector on the end of the extension cable into the connector on the end of ...

  • Page 50

    Section 4 spo 2 sensors 44 co 2 smo plus! User’s manual rev. 02 4. Select the appropriate size sensor based on the patient type. Pediatric/adult sensor neonatal/pediatric sensor adult finger adult toe pediatric toe neonatal hand neonatal foot.

  • Page 51

    Spo 2 sensors sensors and patient connections rev. 02 co 2 smo plus! User’s manual 45 5. To apply the sensor, place the blue side of the sensor wrap against the skin, wrap it around the site and secure with velcro † tab. The velcro tab on the neonatal/pediatric version is removable to allow the foam...

  • Page 52

    Section 4 spo 2 sensors 46 co 2 smo plus! User’s manual rev. 02 [this page intentionally blank.].

  • Page 53: Section 5

    Rev. 02 co 2 smo plus user’s manual 47 section 5 monitoring display of data the co 2 smo plus! Measures and displays respiratory mechanics, carbon dioxide and pulse oximetry data. Until valid data is recieved for any parameter, that parameter will show a dash “--”. When valid data is received, the v...

  • Page 54

    Section 5 48 co 2 smo plus user’s manual rev. 02 resp spon resp mech respiration rate of spontaneous and mechanical breaths. Data key pif peak inspiratory flow rate pef peak expiratory flow rate i:e ratio of inspiration time (ti) to expiration time (te) ti inspiration time (time taken for inspiratio...

  • Page 55

    Monitoring rev. 02 co 2 smo plus user’s manual 49 enter weight data key to enter/abg weight screen vt ml/kg mech tidal volume per kilogram of body weight mark abg: event key vd/vt phy deadspace (physiological) to tidal volume ratio vd phy physiological deadspace vd alv alveolar deadspace data set - ...

  • Page 56

    Section 5 50 co 2 smo plus user’s manual rev. 02 data key monitoring screen raw i * static raw i is measured with inspiratory pause and is available in adult mode only. Stat is displayed in place of the units dynamic inspired airway resistance raw e * static raw e is measured with inspiratory pause ...

  • Page 57

    Monitoring rev. 02 co 2 smo plus user’s manual 51 * an asterisk next to a parameter abbreviation denotes that the parameter is trended in co 2 smo plus! And that the trend is downloaded to co 2 smo plus! For windows. Data key mv spon * spontaneous minute volume mv mech * mechanical minute volume mv ...

  • Page 58

    Section 5 52 co 2 smo plus user’s manual rev. 02 parameters not displayed in addition to the parameters listed above, the following parameters are also calculated and communicated through the rs232 output for display on a pc but are not displayed on the co 2 smo plus! Screen: parameter abbreviation ...

  • Page 59: Screen Displays

    Screen displays monitoring rev. 02 co 2 smo plus user’s manual 53 screen displays the following screens are displayed on the co 2 smo plus! : waveform screens co 2 section spo 2 section flow parameters capnogram etco 2 value spo 2 value respiration rate pulse rate capnogram (press key for combinatio...

  • Page 60

    Section 5 screen displays 54 co 2 smo plus user’s manual rev. 02 refer to scale setting in the configuration screen (see “configuration settings” on page 19) to display or hide scales, and for manual or automatic scale adjustment. Flow waveform co 2 section spo 2 section etco 2 value spo 2 value res...

  • Page 61

    Screen displays monitoring rev. 02 co 2 smo plus user’s manual 55 trend screens on-screen trends are displayed as bar graphs, with 20 bars each. Using the key, the time base can be selected for 20 minutes, 1, 4, 8, 12, or 24 hours. The value of each bar is the average value in the corresponding time...

  • Page 62

    Section 5 screen displays 56 co 2 smo plus user’s manual rev. 02 data screens data set - standard • vt exp mech (avg) • vt alveolar spon • vt exp spon (avg) • vt alveolar mech • raw inspired • mv alveolar total • raw expired • mv alveolar spon • map (mean airway pressure) • mv alveolar mech • pip (p...

  • Page 63

    Screen displays monitoring rev. 02 co 2 smo plus user’s manual 57 to enter paco 2 (for calculation of vd/vt phy, vd phy, and vd alv): while this screen is displayed, press the key when a blood sample is drawn. A time and date marker will be placed below the paco 2 value. When the abg results are kno...

  • Page 64

    Section 5 screen displays 58 co 2 smo plus user’s manual rev. 02 data set - protocol calculated based on entered patient weight calculated based on entered paco 2 value.

  • Page 65

    Screen displays monitoring rev. 02 co 2 smo plus user’s manual 59 setup screen refer to “setup and configuration settings” on page 16 for more information. Alert settings screen refer to “alerts and messages” on page 73 for more information. Pointer press to move pointer to desired parameter adjust ...

  • Page 66

    Section 5 notes on patient monitoring 60 co 2 smo plus user’s manual rev. 02 notes on patient monitoring flow/co 2 sensor tubing purge the co 2 smo plus! Includes an automatic and manual purge feature which provides a flow rate of room air to keep the pressure sense lines free from water condensatio...

  • Page 67

    Notes on patient monitoring monitoring rev. 02 co 2 smo plus user’s manual 61 • water that accumulates in the neonatal combo co 2 /flow and neonatal flow sensors or the sensor tubing may cause the reported tidal volumes to be higher than set volumes. If reported values are higher (or lower) than exp...

  • Page 68

    Section 5 notes on patient monitoring 62 co 2 smo plus user’s manual rev. 02 below is an example of a waveform that exhibits a small leak in the flow sensor tubing. Replace flow sensor; if problem persists, refer monitor to qualified service personnel. Flow sensor zeroing the systems ability to main...

  • Page 69

    Notes on patient monitoring monitoring rev. 02 co 2 smo plus user’s manual 63 manual flow zero - neonatal mode a manual flow sensor zero may be used if a positive or negative offset of the flow waveform from zero baseline is observed. This condition mainly occurs if the orientation of the co 2 smo p...

  • Page 70

    Section 5 notes on patient monitoring 64 co 2 smo plus user’s manual rev. 02 [this page intentionally blank.].

  • Page 71: Section 6

    Transmitted parameters hp vuelink interface rev. 02 co 2 smo plus user’s manual 65 section 6 hp vuelink interface the co 2 smo plus! Interface to the hewlett packard vuelink module provides a pathway for co 2 smo plus! Data to be viewed on the hewlett packard patient monitoring system. Twenty eight ...

  • Page 72: Preparing For Use

    Section 6 preparing for use 66 co 2 smo plus user’s manual rev. 02 transmitted waveforms preparing for use this section details the setup and connection of the co 2 smo plus! To the hewlett packard vuelink module. To prepare the co 2 smo plus! For operation you will need the following: • co 2 smo pl...

  • Page 73

    Preparing for use hp vuelink interface rev. 02 co 2 smo plus user’s manual 67 component setup 1. Connect the capnostat, flow sensor, pulse oximeter probe and ac power cord as described in sections 3 and 4 of this user’s manual 2. Connect the 25 pin male connector on the vuelink interface cable to th...

  • Page 74: Module Setup

    Section 6 module setup 68 co 2 smo plus user’s manual rev. 02 module setup a total of 28 parameters are trended in the co 2 smo plus! Unit; 17 of these parameters can be transmitted to the hewlett packard patient monitor. From this list of 17 pre-selected parameters (shown below), you may choose up ...

  • Page 75

    Auxiliary plus task window hp vuelink interface rev. 02 co 2 smo plus user’s manual 69 scale wav e - this selection allows you to set the scales for the waveforms. When selected, the parameter table in the auxiliary plus window will clear and the available scale settings for the waveform currently b...

  • Page 76: Trends/calcs

    Section 6 trends/calcs 70 co 2 smo plus user’s manual rev. 02 7. You may select a new waveform by pressing the lighted right or left arrow key until you get to the desired selection. You may select a new parameter by pressing any of the lighted arrow keys until you get to the desired selection. 8. W...

  • Page 77: Smo Plus!

    To stop monitoring with the co2smo plus! Hp vuelink interface rev. 02 co 2 smo plus user’s manual 71 pressing the zoom in softkey will increase the resolution of the trend graph in increments of one-half hour, 1 hour, 2 hours and 3 hours. Zoom out - this selection decreases the trend resolution. The...

  • Page 78

    Section 6 to stop monitoring with the co2smo plus! 72 co 2 smo plus user’s manual rev. 02 [this page intentionally blank.].

  • Page 79: Section 7

    Rev. 01 co 2 smo plus user’s manual 73 section 7 alerts and messages alerts alerts are generated for etco 2 , respiration rate, spo 2 , and pulse rate. These alerts occur when the limits for a particular parameter are exceeded. There is also an alert when there is a loss of respiration for a consecu...

  • Page 80

    Section 7 alerts 74 co 2 smo plus user’s manual rev. 01 erase stored trends? Yes: no: this prompt appears each time the monitor is turned on. To erase stored trends press . To keep stored trends press , or allow the message to time out, or press any other key. Event marker this message appears in th...

  • Page 81

    Alerts alerts and messages rev. 01 co 2 smo plus user’s manual 75 respiratory mechanics (flow) messages oxygen saturation messages message meaning of message adult flow sensor identified optical encoders identified that a pediatric/adult flow sensor is connected. Adult flow/co2 sensor identified opt...

  • Page 82

    Section 7 alerts 76 co 2 smo plus user’s manual rev. 01 capnography messages probe off the spo 2 probe is off the patient. Pulse out of range the pulse rate is beyond the measurement range of the monitor (30-250 beats/min). Spo 2 hdw error an error with the monitor’s internal circuitry has been dete...

  • Page 83

    Alerts alerts and messages rev. 01 co 2 smo plus user’s manual 77 etco 2 out of range the detected etco 2 value is beyond the measurement range of the monitor (0-100mmhg, 0-12.5% or kpa). Faulty sensor: x capnostat co 2 sensor needs servicing (x = service code) gas comp gas compensation (setup scree...

  • Page 84

    Section 7 alerts 78 co 2 smo plus user’s manual rev. 01 zeroing co 2 on airway adapt: xx an airway adapter zero is in progress. Xx indicates the number of seconds remaining. Zeroing co 2 on -0- cell: xx capnostat co 2 sensor is currently attached to the zero cell, and a “cell zero” is in progress. Z...

  • Page 85: Battery Status And Alerts

    Battery status and alerts alerts and messages rev. 01 co 2 smo plus user’s manual 79 battery status and alerts the led on the front panel flashes when the battery is extremely low; this will be accompanied by an audible alert. When this occurs connect the power cord to the monitor’s rear panel power...

  • Page 86: Setting Alert Limits

    Section 7 setting alert limits 80 co 2 smo plus user’s manual rev. 01 setting alert limits the alert limits for the monitor can be manually or automatically adjusted. When the monitor is turned off the alert limit settings will be retained, even if ac power and the battery are disconnected. The moni...

  • Page 87: Alert Audio

    Alert audio alerts and messages rev. 01 co 2 smo plus user’s manual 81 alert audio an audible alert is generated any time an alert condition is detected provided that neither the 2 minute silence, nor the audible alert muting are enabled. When the alert delay in the configuration menu is set to on t...

  • Page 88

    Section 7 alert audio 82 co 2 smo plus user’s manual rev. 01 [this page intentionally blank.].

  • Page 89: Section 8

    Rev. 02 co 2 smo plus user’s manual 83 section 8 maintenance cleaning and sterilization follow the cleaning and sterilization instructions listed below to clean and/or sterilize the monitor and its accessories. Monitor • turn the monitor off and unplug the line cord from the ac power source before c...

  • Page 90

    Section 8 cleaning and sterilization 84 co 2 smo plus user’s manual rev. 02 • after cleaning or sterilizing the y-sensor, verify that the sensor is physically intact, with no broken or frayed wires or damaged parts. Make certain that the connectors are clean and dry, with no signs of contamination o...

  • Page 91

    Cleaning and sterilization maintenance rev. 02 co 2 smo plus user’s manual 85 • the neonatal airway adapter (cat. No. 7053) may be sterilized using eto (ethylene oxide) gas. Be sure to use appropriate aeration times. • before reusing the adapter, ensure that the windows are dry and residue-free, and...

  • Page 92: Maintenance Schedules

    Section 8 maintenance schedules 86 co 2 smo plus user’s manual rev. 02 8. Follow this with at least three more syringes of distilled water, followed by at least two more syringes of air to make sure that most of the distilled water has been drained. 9. Remove the syringe from the unit. Do not connec...

  • Page 94

    Section 8 mains voltage configuration 88 co 2 smo plus user’s manual rev. 02 4. Pry the fuse housing out from the power entry module. 5. Replace the blown fuse(s) with the proper type and rating. 6. Reinstall the fuse housing. When positioning the housing into the power entry module make sure that i...

  • Page 95

    Mains voltage configuration maintenance rev. 02 co 2 smo plus user’s manual 89 changing the mains voltage setting 1. Check that the monitor is off. 2. Set the rear panel power entry module switch to off (“o”). Remove the line cord from the power entry module. 3. Using a flat blade screwdriver, pry t...

  • Page 96

    Section 8 mains voltage configuration 90 co 2 smo plus user’s manual rev. 02 [this page intentionally blank.].

  • Page 97: Section 9

    Rev. 02 co 2 smo plus user’s manual 91 section 9 specifications general specifications for the novametrix co 2 smo plus! Monitor, model 8100, are listed for informational purposes only, and are subject to change without notice. Capnograph • principle of operation: non-dispersive infrared (ndir) abso...

  • Page 98: Pulse Oximeter

    Section 9 pulse oximeter 92 co 2 smo plus user’s manual rev. 02 pulse oximeter • oxygen saturation range: 0-100% accuracy: ± 2% for 80-100%, unspecified for 0-79% (approximately 68% of the observations are within the accuracy claim) display resolution: 1% averaging time: menu selectable 2 or 8 secon...

  • Page 99: Monitor Specifications

    Monitor specifications specifications rev. 02 co 2 smo plus user’s manual 93 monitor specifications • classification (iec601-1): class i/internal power source, type bf, continuous operating mode • operating environment: 50-104 ° f (10-40 ° c), 0-90% relative humidity (non-condensing) • size: height ...

  • Page 100

    Section 9 additional features 94 co 2 smo plus user’s manual rev. 02 [this page intentionally blank.].

  • Page 101: Section 10

    Rev. 02 co 2 smo plus user’s manual 95 section 10 warranty equipment manufactured or distributed by novametrix medical systems inc., is fully guaranteed, covering materials and workmanship, for a period of one year from the date of shipment, except for certain disposable products and products with s...

  • Page 102: Service Policy

    Section 10 service policy 96 co 2 smo plus user’s manual rev. 02 service policy novametrix medical systems inc. Will provide warranty service support to its customers within 48 hours of receiving a telephone request for technical support. This 48 hour period begins once a service request is placed t...

  • Page 103: Section 11

    Rev. 02 co 2 smo plus user’s manual 97 section 11 accessories co 2 smo plus! ( model 8100 ) respiratory profile monitor catalog no. Description 6758-00 co 2 smo plus! Respiratory profile monitor (model 8100) with capnostat co 2 sensor (cat. No. 7167), oxysnap superbright finger sensor (cat. No. 8744...

  • Page 104

    Section 11 98 co 2 smo plus user’s manual rev. 02 flow sensors for monitoring airway flow without an adapter for the capnostat co 2 sensor. 6717-00 series 3 pediatric/adult flow sensor (10 per box) 6718-00 series 3 neonatal flow sensor (10 per box) spo 2 sensors and accessories reusable oxysnap™ spo...

  • Page 105

    Accessories rev. 02 co 2 smo plus user’s manual 99 6455-25 pediatric/adult single patient use spo 2 sensor terminates in db9 connector (25 per box) 6480-00 neonatal/pediatric single patient use spo 2 sensor terminates in db9 connector (10 per box) 6480-25 neonatal/pediatric single patient use spo 2 ...

  • Page 106

    Section 11 100 co 2 smo plus user’s manual rev. 02 [this page intentionally blank.].