Oakton ION 5 Acorn Series Operating Instructions Manual

Manual is about: pH/Ion/°C Meter

Summary of ION 5 Acorn Series

  • Page 1

    Distributed by: 00702-79 operating instructions oakton wd-35613-21, -30, -31, -33 ion 5 and ion 6 acorn series meters printed in the u.S.A. 8/99 ion 6 acorn series ph/ion/ ° c meter on off cal hold enter mode 10. Warranty oakton warrants this meter to be free from significant deviations in material ...

  • Page 2

    3 2 table of contents 1. Introduction ............................................................................................4 2. Display and keypad functions .................................................................5 3. Preparation..........................................................

  • Page 3

    5 4 1. Introduction thank you for purchasing an oakton ® ion acorn meter. These economical, microprocessor-based meters measure the ion concentration of various ions (mono and divalent). The acorn ion meters feature: • large lcd for clear and easy reading • ion 5 meter allows ion and mv measurement ...

  • Page 4

    3. Preparation inserting the batteries the battery compartment is found at the back of the instrument. To open the battery compartment, push the lid up. See figure note the polarity of battery before insert- ing the batteries into position. After replacement, place the cover back into its position a...

  • Page 5

    6. If you do wish to calibrate to this point, press the enter key. The meter will display the mv output of your electrode in the standard solution. Allow the instrument to stabilize. Stabilization may take a few minutes depending on electrode and standard concentration. Electrode response in low con...

  • Page 6

    11 10 ph buffer set selection (ion 6 meter only) the ion 6 meter is capable of up to three-point ph calibration to ensure accuracy across the entire range of the meter. The ion 6 meter can be set to recognize either of two buffer sets: usa or nist buffers. Usa —“usa”: 4.01; 7.00 and 10.01 nist –“nst...

  • Page 7

    Note : this meter has automatic buffer recogni- tion which identifies the correct ph buffer values during calibration. If buffers other than 4.01, 7.00 or 10.00 are used, or the electrode has worn out, the lcd will flash "er1". See figure press cal key to clear. Temperature calibration (ion 6 meter ...

  • Page 8

    15 14 5. Measurement measuring ion concentration 1. Switch on meter. The ion 5 and ion 6 meters will automatically enter the ion measurement mode. No annunciator will appear. See figure 2. Prepare samples as necessary (i.E., add ionic strength adjuster). Sample preparation varies depending on ion ty...

  • Page 9

    17 16 6. Electrode care and maintenance ion selective electrode care for ion selective electrode care, consult the instruction manual included with your electrode. Ph electrode care (ion 6 meter only) storage for best results, keep the ph bulb wet. Store the ph bulb in the electrode soaker bottle fi...

  • Page 10

    19 18 8. Specifications mode ion temperature mv ph (ion 6 only) range 0.1 to 1999 ppm 0.00 to -500 to 0.0 to 100°c +500 mv 14.00 ph 0.1 ppm from 0.1 mv, resolution 0.1 ppm-199.9 ppm 0.1°c ±199.9 mv; 0.01 ph 1 ppm from 1 mv outside 200-1999 ppm this range ±0.2 mv, accuracy ±0.5% of reading ±0.5°c fro...

  • Page 11

    21 20 wd-35812-42 sodium ion selective electrode, includes 15 ml of reference electrolyte and a filling pipette. Shpg wt 0.5 lb (230 g). Wd-35812-93 sodium ion selective solution kit, includes 475 ml ion strength adjuster, 475 ml calibration standard, 60 ml reference electrolyte, and syringe dispens...

  • Page 12

    23 22.