Oakton PC 700 Instruction Manual - Table 1

Manual is about: pH/mV/Conductivity/ºC/ºF Bench Meter

Summary of PC 700

  • Page 1

    Instruction manual pc 700 ph/mv/conductivity/ºc/ºf bench meter technology made easy ... Part of thermo fisher scientific 68x541704 rev. 1 feb. 2010 www. Globaltestsupply .Com find quality products online at: sales@globaltestsupply.Com.

  • Page 2: Table of Contents

    Table of contents 1. Introduction .............................................................. 1 2. Getting started........................................................ 2 2.1 keypad functions ................................................................... 2 2.2 lcd annunciators ...............

  • Page 3: 10.

    9.6 3.3 buff (ph buffers & calibration points)—ph only ............. 13 9.7 3.3 atc (auto temp compensation)—con & tds only .......... 13 9.8 3.4 tds (tds factor)—con & tds only ................................... 14 9.9 3.5 t.Co (temperature coefficient)—con & tds only............ 14 9.10 3.6 t.Nr ...

  • Page 4: Introduction

    1 1. Introduction thank you for purchasing our pc 700 series benchtop meter. This microprocessor-based meter is economical and simple to use. The design incorporates a large lcd for clear viewing, yet offers a small footprint to conserve space. The pc 700 measures ph, mv (orp), conductivity, or tds ...

  • Page 5: Getting Started

    2. Getting started 2.1 keypad functions powers the meter on and off. Upon power on, the meter automatically begins in the mode that was last used. Calibration and memory values are retained even if meter is unplugged. Toggle between available measurement modes— ph/temp, mv/temp, conductivity/temp, o...

  • Page 6: 2.2  Lcd Annunciators

    2.2 lcd annunciators 2.3 meter connections dc power supply con/ temp 8-pin din connection for 2-cell con/tds/temp electrode ph bnc connection for ph, or orp (redox) electrode 3 www. Globaltestsupply .Com find quality products online at: sales@globaltestsupply.Com.

  • Page 7: Conductivity Electrode

    3. Conductivity electrode the pc 700 includes an electrode with a nominal cell constant of k = 1.0, built-in temperature sensor, and 1 meter cable. The ultem body housing has good chemical resistant properties. The electrode design offers fast temperature response and reduces air entrapment, ensurin...

  • Page 8: Ready

    5 e buffer group see section 9.6 to change th 1) press as needed to select ph. Dip the ph and atc electrodes into ph buffer and press 2) . The secondary display will lock on the appropriate buffer value. Provide ing for best results. When the ready indicator appears, pr stirr ess to accept. The prim...

  • Page 9: Note:

    6 tandard in which a meter/electrode system are d are match the orp value. 4.2 millivolt (mv) offset adjustment oxidization reduction potential (orp or redox) is not a precise measurement, but is useful as a relative indicator. As such, mv offset adjustment is not meant to enhance accuracy, but rath...

  • Page 10: Table 1

    7 isted in table 1 , the pc 700 must be set for manual calibration i the factory default is automati an ere i for e 1. Table if you only use calibration standards that are listed in table 1 , automatic calibration is recommended. If you intend to calibrate with one or more standards that are not l n...

  • Page 11: Ration Procedure

    8 his ual calibration so that it is applied to one range ration, perform a calibration in each est results. On point per range can be calibration and between samples y which will retain all calibration factory default value. E unstable and are very temperature dependent. As a result, reproducible ca...

  • Page 12: Ready

    4) when the ready indicator appears, press to accept. The t primary reading will flash briefly before returning to measuremen mode upon successful calibration. 5.5 manual conductivity & tds calibration procedure 1) press as needed to select conductivity ( μ s or ms) or tds (ppm or ppt) calibration. ...

  • Page 13: 6.1  Taking Measurements

    6. Conductivity and tds measurement 6.1 taking measurements be into the sample beyond the upper steel band side of the probe guard for y. G measurement. Erroneous results will occur while the probe guard is 1) rinse the electrode with de-ionized or distilled water before use to remove any impurities...

  • Page 14: Hold Function

    7. Hold function for prolonged observation of a reading, press during measurement mode to freeze the display. The “hold” indicator will display when the reading is held. To release the held value and resume measurement, press again or insert the held value into memory by pressing . 8. Storing and re...

  • Page 15: Setup Functions

    9. Setup functions use the setup feature to customize your instrument operation. First, select the appropriate measurement mode you wish to adjust by pressing until the desired units are displayed (i.E. Ph, mv, conductivity, or tds). During measurement, press and hold for 5 seconds to enter setup mo...

  • Page 16: 3.1

    13 ) 9.3 3.0 conf (configuration press to access set-up programs 3.1 thru 3.9. 3.1 9.4 rdy (ready / stability indicator) press . Press or to choose ready “on”, ready “off”, or auto hold. Press to confirm. 9.5 3.2 enheit) ºc ºf (celsius or fahr press . Press or to select ºc or ºf. Press to confirm. ....

  • Page 17: 9.8  3.4

    Press or to select “yes” or “no”. Press to confirm. 9.8 3.4 n & tds only ress tds (tds factor)—co p . Press or to select the desired tds factor (.40 to 1.00). Press to confirm. 9.9 3.5 t.Co (temp efficient)—con & tds only the tem c e amount of change in conductivity per his will erature co erature o...

  • Page 18: Acal

    Press . Press or to select the desired value (15.0 to 30 press .0). To confirm. Acal to nductivity calibration)—con & tds only 9.11 3.7 (au co see sectio for more deta atic conductivity calibration. Ress n 5.1 ils on autom . P press or to select “yes” (automatic) or “no” (manual). Press to confirm. ...

  • Page 19: 9.14  4.0 Rst (Reset)

    16 9.14 4.0 rst (reset) . Press or press to select “yes” (reset) or “no” (cancel). S if “yes”, pres or to select “cal” (calibration reset ete reset to factory default settings). Only) or “fct” (compl press to confirm. Clr (clear mem ry) press 9.15 5.0 o . Press or to select “yes” (erase memory) or “...

  • Page 20: 10.

    17 10. Calculating tds conversion factor you can calibrate tds using the value of the calibration standard solution at a standard temperature such as 25 ºc. To determine the conductivity- to-tds conversion factor use the following formula: factor = actual tds ÷ actual conductivity @ 25 ºc actual tds...

  • Page 21: 11.

    11. Calculating temperature coefficients to determine the temperature coefficient of your sample solution use this formula: where: tc = temperature coefficient 25 = 25 ºc c t1 = conductivity at temp 1 c t2 = conductivity at temp 2 t1 = temp 1 t2 = temp 2 note: a controlled temperature water bath is ...

  • Page 22: 12.

    19 12. Replacements and accessories part number ordering code item description eutech instruments oakton instruments pc 700 with ph electrode (ecfc7252101b / 59001- 70), conductivity/temp electrode (consen9501d / 35608-74), integral stand,100/240 vac ecpc70043s 01x543601 35413-00 pc 700 with integra...

  • Page 23: 13.

    13. Troubleshooting guide problem cause solution no display main power not switched on. Ac adapter socket not inserted properly. Switch on the power supply. Re-insert ac adapter. “ur” (under range) “or” (over range) in primary or secondary display measured value is out of range. Electrode not connec...

  • Page 24

    Calibration error. Buffer value does not match displayed value or electrode is disconnected or failing. Use fresh buffer solutions. Check electrode connection. Clean & recondition electrode. Replace electrode. Broken or worn electrode replace electrode unstable ph reading external ‘noises’ or induct...

  • Page 25: 14.

    22 14. Specifications ph range 0.00 to 14.00 ph resolution 0.01 ph accuracy 0.01 ph + 1 count calibration points up to 5 points with auto-buffer recognition buffer options usa : ph 1.68, 4.01, 7.00, 10.01,12.45 nist: ph 1.68, 4.01, 6.86, 9.18 ,12.45 slope display yes (with offset) mv range 2000 mv r...

  • Page 26

    23 cell constant 0.1, 1.0, 10.0 (selectable) tds factor 0.40 to 1.00 (adjustable) calibration points automatic (4 points); maximum 1 per range manual (5 points); maximum 1 per range auto-ranging yes hold function yes memory 100 data sets input bnc, 8-pin din power ac/dc 9v, 6w adapter (100/240 vac, ...

  • Page 27: 15.

    24 15. Warranty this meter is supplied with a warranty against significant deviations in material and workmanship for a period of three years from date of purchase whereas probe with a six month warranty. If repair or adjustment is necessary and has not been the result of abuse or misuse within the ...

  • Page 28: 16.

    25 16. Return of items authorization must be obtained from our customer service department or authorized distributor before returning items for any reason. A “return material authorization” (rma) form is available through our authorized distributor. Please include data regarding the reason the items...