Oakton pH 10 Series Operating Instructions Manual

Summary of pH 10 Series

  • Page 1

    Operating instructions oakton ® ph 10 series ph/mv/°c meter printed in singapore 8/00 rev 1 ph • mv • ° c meter hold mode on off cal meas ° c ph meas ph atc ready ph 10 series con.

  • Page 2

    Table of contents 1. Introduction ..............................................................................Page 4 getting started • maintenance, troubleshooting 2. Display and keypad functions.................................................Page 5 3. Preparation ...................................

  • Page 3

    2. Display and keypad functions the water-resistant membrane keypad makes the instrument easy to use. Each button, when pressed, has a corresponding graphic indicator on the large liquid crystal display. The lcd has a primary and secondary display. The primary display shows the readings for ph and m...

  • Page 4

    Page 7 3. Preparation inserting the batteries (included) 1. If the electrode cable is wound around the the cable winder, unwind it. 2. Use a phillips screwdriver to remove the two screws holding the battery cover. See figure b. 3. Lift meter stand to expose battery cover. 4. Remove battery cover. 5....

  • Page 5

    Page 9 4. Ph calibration this instrument is capable of up to 3-point calibration to ensure accuracy across the entire range of the meter. You can perform 1-, 2,-, or 3-point calibration with standard ph buffers 4.01, 7.00 and 10.01. We recommend that you perform at least a 2-point calibration using ...

  • Page 6

    Page 11 5. Ph measurement during measurement, you can hold the electrode in your hand or insert it in the electrode holder for one-hand operation. Note: be sure to remove the protective rubber cap of the electrode before measurement. Measurement with automatic temperature compensation (atc) automati...

  • Page 7

    Page 13 page 12 6. Mv measurement the ph 10 series meter allows you to read the absolute mv output of your ph, orp or ion selective electrode (ise). For orp or ise electrodes contact your oakton distributor. 1. Connect your orp or ion selective electrode as shown in section 3 2 rinse the electrode w...

  • Page 8

    Page 15 page 14 8. Temperature calibration 1-point temperature calibration 1. Connect atc probe (or temperature connector of the “all-in-one” electrode) to the phone jack. The atc annunciator will appear at the right- hand side of the lcd. 2. Press the mode key to select temperature mode (temp). 3. ...

  • Page 9

    Page 16 c. Temperature low point calibration the primary display will show "0.0" and the secondary display will show "r2.0". See figure u. 1. Insert the temperature sensor or electrode with built-in temperature sensor into the low temperature bath. 2. Press the ▼ and ▲ keys to set the primary displa...

  • Page 10

    Electrode reactivation: for glass bodied electrodes only! Warning: this procedure should only be attempted by qualified persons proficient with the safe handling of dangerous chemicals. Provide proper containers, fume hoods, ventilation, and waste disposal. Safety goggles and protective clothing mus...

  • Page 11

    11. Troubleshooting problem cause solution power on but a) batteries not in place. A) insert batteries. No display. B) batteries not in correct b) re-insert batteries in polarity (+ and –). Correct polarity. C) weak batteries. C) replace batteries or attach optional ac adapter. Unstable reading. A) ...

  • Page 12

    14. Warranty oakton warrants this product to be free from significant deviations in material and workmanship for a period of one year from date of purchase. If repair or adjustment is necessary and has not been the result of abuse or misuse within the one year period, please return—freight prepaid—a...

  • Page 13