Oakton Temp-340 Instruction Manual

Manual is about: Thermistor Datalogging Thermometer

Summary of Temp-340

  • Page 1

    Instruction manual oakton ® temp-340 thermistor datalogging thermometer instruction manual temp-340 thermistor datalogging thermometer part of thermo fisher scientific 68x536601 r0 mar 2010

  • Page 3: Table Of Contents

    Table of contents 1. Introduction ....................................... 1 2. Safety precautions.......................... 2 3. Specifications .................................... 3 4. Battery installation and replacement........................................ 6 5. Inserting and removing optional ...

  • Page 4: 1. Introduction

    1. Introduction this versatile hand-held instrument provides highly accurate temperature measurements. The instrument is designed for easy operation and includes the following features: • menu driven setup and operation • datalogging for up to 2000 points • usb output • operator selection of celsius...

  • Page 5: 2. Safety

    - 2 - 2. Safety precautions warning: 1. This instrument is designed to accept low level signals supplied by standard thermistors. Under no circumstances should the input voltage exceed the specified 50v rms. 2. To prevent ignition of a hazardous atmosphere, batteries must only be changed in an area ...

  • Page 6: 3. Specifications

    - 3 - 3. Specifications thermistor thermometers type temperature range -40 to 150°c 400 series (-40 to 302°f) out of range display: open resolution 0.01 or 0.1 °f/°c; auto-ranging to 0.1° above +99.99° accuracy from –40.00 to 99.99°f (–40.00 to 99.99°c): ±0.06°f (±0.03°c), from 100.0 to 257.0°f (100...

  • Page 7: Auto Off (Adjustable Time)

    - 4 - auto off (adjustable time) enable/disable option available stability criteria yes, upon stability of 5 seconds display update rate 0.6 sec per update. Input one bayonet connector. Input protection 50v rms storage –40°c to 65°c (–40°f to 149°f) humidity 10% to 90% (non-condensing) battery life ...

  • Page 8: Ingress Protection:

    - 5 - ingress protection: meets iec-529 ip-54 for dust and water resistant enclosures (probe attached) ce compliance en61326-1/a1: 1998 (eu emc directive).

  • Page 9: 4. Battery

    4. Battery installation and replacement the total battery life without backlighting is about 400 hours. Remaining battery power is indicated but the battery life indicator. - 6 - indicator voltage cell + 3 bars more than 4.1 v cell + 2 bars more than 3.6 v cell + 1 bars more than 3.0 v cell + empty ...

  • Page 10

    - 7 - selected settings are stored in memory and will remain in memory even after power is turned off, or while batteries are being replaced. 1. Before changing battery, turn instrument off and disconnect thermistor. 2. Loosen screw and lift battery cover off back of case. 3. Remove the three aa bat...

  • Page 11: 5. Inserting

    5. Inserting and removing optional rubber armour 1. To insert thermometer into the optional rubber armor, slide in from the top of meter before pushing the bottom edges of meter down to set it into position. Lift up the stand at the back of meter for bench top applications if necessary. 2. To remove...

  • Page 12: 6. Assembling

    6. Assembling optional handsfree accesories you can use the optional magnets and strap in the handsfree kit accessories for hands free operations. - 9 -.

  • Page 13: 7. Connecting

    7. Connecting a thermistor use the correct 400 series thermistor for your instrument. Using an incorrect probe type will result in erroneous readings. Insert the bayonet plug into the mating connector on the top of the instrument. If no probe is connected the display will read “open”. Thermistors ar...

  • Page 14: 8. Key

    8. Key functions note: function keys change in setup mode to provide advanced operation flexibility. Text above key will indicate function. F1 step through min, max and avg readings. F2 toggle between f and c display f3 toggle between menu and measure mode hold freeze display on/off light turns mete...

  • Page 15: 9. Display

    9. Display overview the dot matrix display features a large primary display, smaller secondary displays for channel info or min/max/ave, and helpful annuciators for added measurement data - 12 -.

  • Page 16

    - 13 - 1 power supply indicator (battery/main adaptor/usb) 2 date (format of mmm – dd) 3 time (hour:min 4 time format (am/pm/hrs) 5 measurement mode 6 hold function indicator 7 stable indicator 8 data logging indicator 9 pc data logging indicator 10 main reading display 11 current reading unit indic...

  • Page 17: 10. Measurement Mode

    10. Measurement mode on initial start-up the meter will display the measured value for in the primary display. Pressing the f2 key will toggle reading between f and c display. Pressing the f1 key initiates and toggles through minimum, maximum, and average reading modes. Pressing f3 enter accesses se...

  • Page 18: 11. Hold Functon

    - 15 - 11. Hold functon press the hold key to retain the reading on the display. Press the hold key again for normal operation. 12. Min, max, and ave function press the f1 key to toggle between the minimum, maximum, and average readings. The minimum and maximum reading function is ideal for monitori...

  • Page 19: 14. Setup Mode

    14. Setup mode to access the setup mode from measurement mode press the f3 key (menu). Press ▲▼ keys on the meter key pad to scroll through options. To enter a setup screen press select f1 key. To return to measurement mode press meas f3 key. Following menu options are listed 1. General setup 2. Use...

  • Page 20: 15.  General Setup Screen

    15. General setup screen the first page of the general setup screens let you set resolution, auto-off time, and password. Press f1 to indicate you want to change the setting of the current parameter or recall▲ or log▼ to move to the next parameter. Press recall▲ or log▼ to change the options. Press ...

  • Page 21

    This screen below is used to reset/change password. In the event if uses forget his/her password, 5586 can be used to eset to a new value r on the second page you can set time and date. - 18 -.

  • Page 22: 16. Calibration Screen

    16. Calibration screen the thermometer is factory calibrated and does not require calibration before use. The calibration function allows single point calibration of the thermometer, at any temperature point to compensate for thermistor off-set error. It is not necessary to perform a field calibrati...

  • Page 23

    There are two calibration options: offset - adjusts at a single point. Offset calibration can be performed at any temperature in the offset range of ±5.00 °c or 9.00 °f slope – adjusts at two points. The two calibration points must be at least 20.00 °c (36.00 °f) apart. The 2nd calibration point sho...

  • Page 24: 17. Alarms Screen

    17. Alarms screen there are two kind of alarm setting is available under the alarm setting options measurement alarm disable or enable the alarm by pressing recall▲ or log▼and f1 to accept. Increase or decrease individual limit by pressing recall▲ or log▼. Meter in alarm mode - 21 -.

  • Page 25

    Countdown alarm you can enable/disable the countdown alarm and set the countdown time from 5 sec to 1 hour. After setting (enabling) the countdown alarm, the measurement screen should look like this: - 22 -.

  • Page 26

    Press f2 key to start the countdown: press f2 key in the middle of a countdown will stop the process: you will need to restart the countdown by selecting the “start” key (f2). Alternatively, you can also choose “menu” to go into the timer menu and disabling the alarm. - 23 -.

  • Page 27

    The buzzer will sound for 30 seconds, or until the reset (f2) is pressed, at the end of the countdown. Flashing announciator indicates that the countdown has expired to reset the timer, select “restart” (f2) it. You display should look like this: to repeat the functions, select f2 note: the °c/ °f f...

  • Page 28

    - 25 - “alarm setting page 2” and disable the count down alarm function. When activated, the countdown timer temporarily over-rides the auto-shutoff until the countdown is completed or manually stopped. If the meter is manually or auto shutoff, the countdown alarm is automatically reset to “disable”...

  • Page 29: 18. Data Logging Screen

    18. Data logging screen press recall▲ or log▼to choose the logging methods as auto or manual. If it is auto logging, using recall▲ or log▼to set time interval. Its range is from 0min to 60min. Data transfer from meter to computer - 26 - once the usb connection is establish with.

  • Page 30: On Prior To Connecting Usb

    Pc, press the start button to download data from meter to pc using hyperterminal. ***note: meter must be turned on prior to connecting usb cable or computer may not recognize the instrument*** 19. Calibration report screen the calibration report will show the time and date along with results of the ...

  • Page 31: 20.  Clear / Reset Screen

    20. Clear / reset screen press f1 to choose which data you want to clear or reset. For calibration, logged data and reset all, you will have to enter the password to proceed. (default password is 9900) - 28 -.

  • Page 32: 21.  Maintenance

    - 29 - 21. Maintenance properly used, the thermometer should maintain calibration indefinitely and not require service other than occasional cleaning of the housing and changing of the batteries. 22. Cleaning warning: to prevent ignition of a hazardous atmosphere by electrostatic discharge, clean wi...

  • Page 33: 24. Trouble Shooting

    24. Trouble shooting the following chart lists the most probable faults. There are no internal adjustments or user-replaceable parts. Fault action no display check condition of batteries. Check that batteries are inserted properly. Display shows open no thermistor connected in the connector display ...

  • Page 34: 25.  Accessories

    25. Accessories replacement meters and meter accessories item part number temp 340 thermometer 35426-50 91426-50 rubber armour with stand 35427-80 hands-free kit (two magnets and a strap) 35427-85 general purpose probe (immersion into liquids) 08491-06 flexible general purpose probe (vinyl sheath) 0...

  • Page 35: 26. Warranty

    - 32 - 26. Warranty the manufacturer warrants this product to be free from significant deviations from published specifications for a period of three years. If repair or adjustment is necessary within the warranty period, the problem will be corrected at no charge if it is not due to misuse or abuse...

  • Page 36: Repair Services

    - 33 - 28. Innocal® calibration and repair services optimum performance of your temperature-measuring instrument is not a timeless condition. To ensure quality measurements, have your instrument calibrated regularly. Trust innocal® to satisfy your calibration and equipment repair needs. With over a ...

  • Page 37: Reliable Support

    - 34 - reliable support trust in our free diagnostic support and troubleshooting advice. Our factory- trained metrologists and technicians are armed with years of experience and extensive technical data. Convenient reminders it’s so easy to keep your instruments functioning properly. Based on your r...

  • Page 38: Enough?

    - 35 - as iso 9000, certification is becoming increasingly important. Traceability is not a timeless condition. It must be verified and maintained over the life of the calibration to ensure the highest accuracy possible. When you have your calibration done by innocal, we will send you an automatic r...

  • Page 39: 866-Innocal (866-466-6225)

    - 36 - calibration test points against nist- traceable standards meter only probe only system (meter + probe)* four test points across range of instrument. 0, 100, 165, 230 c (-4, 32, 446, 770f) 17000-06 17001-06 17002-06 innocal—the choice of quality 866-innocal (866-466-6225) innocalsolutions.Com.

  • Page 41: Technical Assistance

    Technical assistance if you have any questions about the use of this product, contact the manufacturer or authorized seller. For more information on oakton instruments products, please contact your nearest distributor or visit our web site listed below: oakton instruments 625 e bunker court vernon h...