OAKWORKS 300 Series User Manual

Manual is about: Procedure Chair

Summary of 300 Series

  • Page 1

    Oakworks ® 300 series procedure chair user manual www.Oakworksmed.Com · 717.235.6807 made in the usa with us & imported parts.

  • Page 2

    ? copyright 2013 oakworks ® , inc. This manual was originally drafted and approved in the english lan- guage. Printed in usa all rights are reserved. No part of this document may be photocopied, reproduced or translated to another language without prior written consent of oakworks ® , inc. Oakworks ...

  • Page 3

    Table of contents product use description .....................................................................................................1 important safety instructions symbol identification ....................................................................................................1 s...

  • Page 4

    Product use description / important safety instructions 1 caution read and save these instructions the back rest and leg rest sections are not designed to support the entire weight of the patient. Do not sit on the back rest or leg rest sections. Medical electrical equipment needs special precaution...

  • Page 5

    2 important safety instructions the arm rests are not designed to be used as patient restraints. Use appropriate patient restraints. The chair should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, the chair should be observed to verify norma...

  • Page 6

    3 product description 300 series procedure chair for electrical specs, see specifications section. Standard specifications motion 1 powered 19”-35” (48-89 cm.) height range motion 3 powered 0-80º back rest motion 2 powered +15º/-5º trendelenburg/ reverse trendelenburg motion 4 powered 0-90º leg rest...

  • Page 7

    4 the procedure chairs come completely assembled and ready to use. Plug the cord into a functioning out- let that is rated for the table. (see grounding below) arrange the power cord and control cords so that they will not create a tripping hazard and where the controls are located to your liking an...

  • Page 8

    5 directions for use caution hand control operations do not sit on the fowler (backrest) or leg rest sections. Do not leave the patient unat- tended. When lowering the chair or using the trendelenburg functions, make sure there is nothing that can impede motion (like stools, cabinets, accessory part...

  • Page 9

    6 directions for use caution foot control operations do not sit on the fowler (backrest) or leg rest sections. Do not leave the patient unat- tended. When lowering the chair or using the trendelenburg functions, make sure there is nothing that can impede motion (like stools, cabinets, accessory part...

  • Page 10

    Directions for use the chair can be rotated with a patient in it. Locate the black push buttons at the top of the back rest section. Press in and hold either one of the buttons to disengage the base locking mechanism. While holding in the push button, rotate the chair to the desired position up to 2...

  • Page 11

    Directions for use 8 caution head rest use do not extend the dowels more than 4” (10 cm.) from the back rest section. Adjustable large the adjustable head rest angle is adjusted using a locking knob. The head rest distance from the table is adjusted by using a knob. To adjust the position of the hea...

  • Page 12

    Slide out the procedure tray by pulling on the center slot of the tray holder until the tray stops. After a procedure is performed, remove the procedure tray with both hands at the side notches from the tray holder. Discard fluids, clean and place back into tray holder. Cleaning and disinfecting ins...

  • Page 13

    10 directions for use to load the paper, locate the two paper holder support arms located on the underside of back rest. Insert rod through paper roll and place rod into the holes in the support arms. Paper roll holder use pull paper and thread through the straps across the upholstered top to keep i...

  • Page 14

    11 directions for use procedure chair face rest crescent the procedure chair face rest crescent is designed to be used for prone positioning. An adjustable head rest is required for this accessory. T-rail adapter the t-rail adapter can be attached at designated locations on one or both sides of the ...

  • Page 15

    12 cleaning & disinfection / inspections / warranty cleaning & disinfection before cleaning with any liquid cleaner be sure to unplug the power cord from the outlet. Warning use a 10% sodium hypochlorite (bleach) solution or recommended disinfectants on all surfaces. Clean all sides of each upholste...

  • Page 16

    13 model number & serial number model & serial numbers model number description p c 3 0 g 3 x x x x x x 3 m t t x x x x procedure chair base top - 3 section back rest table width minimum table height table top length padding: comfort plus fabric: terratouch spaces not used options model number seria...

  • Page 17

    14 specifications product specifications weight 280-325 lbs. (127-148 kg.) shipping weight 372-417 lbs. (169-189 kg.) lifting capacity 550 lbs. (250 kg.) environmental conditions conditions temperature humidity atmospheric pressure normal use 50° (10°c) to 104° (40°c) 20% to 60% rh 98 to 105 kpa sto...

  • Page 18

    15 specifications guidance and manufacturer’s declaration - electromagnetic emissions the chair is intended for use in the electromagnetic environment specified below. The customer or the user of the chair should assure that it is used in such an environment. Emissions test compliance electromagneti...

  • Page 19

    16 specifications guidance and manufacturer’s declaration - electromagnetic immunity the chair is intended for use in the electromagnetic environment specified below. The customer or the user of the chair should assure that it is used in such an environment. Immunity test iec 60601 test level compli...

  • Page 20

    17 specifications guidance and manufacturer’s declaration - electromagnetic immunity the chair is intended for use in the electromagnetic environment specified below. The customer or the user of the chair should assure that it is used in such an environment. Immunity test iec 60601 test level compli...

  • Page 21

    18 this page is intentionally left blank.

  • Page 22

    19 this page is intentionally left blank.

  • Page 23

    20 this page is intentionally left blank.

  • Page 24

    Manual part number mmmnst0037-en revision level: b english, printed in usa contact information: o akworks ® inc. 923 east wellspring road new freedom, pa 17349 phone: 717-235-6807 fax: 717-235-6798 www.Oakworksmed.Com made in the usa with us & imported parts 3034177 conforms to: ansi/aami std es6060...