OEM TUF-2000F User Manual

Manual is about: Integrated Ultrasonic Flow meter

Summary of TUF-2000F

  • Page 1

    Tuf-2000f user manual 1. Introduction 1.1 preface welcome to use new generation transit-time ultrasonic flow meter, please read the user manual carefully before using. 1.2 features ◆ linearity: 0.5%, repeatability: 0.2%, accuracy:±1% ◆ several type transducers for selection, measuring pipe size is f...

  • Page 2

    Tuf-2000f user manual 2 d is the pipe diameter tup is the time for the beam from upstream transducer to the downstream one tdown is the time for the beam from downstream transducer to the upstream one Δ t=tup –tdown 1.4 optional transducer ts ◆ -1 clamp-on type transducer, pipe size from dn15-100mm ...

  • Page 3

    Tuf-2000f user manual 3 analogue output: 4-20ma or 0-20ma current output. Impedance 0 1k . Accuracy 0.1%. Oct output: frequency signal (1~9999hz) relay output: over 20 source signal (no signal, reverse flow etc.) output rs485 serial port three analogue input input three-wire pt100 resistor input (op...

  • Page 4: 1. Menu Window Details

    Tuf-2000f user manual 1. Menu window details the flow meter user interface comprises about 100 independent windows. That are numbered by m00, m01,m02……. M99,m+0,m+1 etc. You do not need to remember all the menu windows. Just remember the most command used windows and the proximate window numbers of ...

  • Page 5

    Tuf-2000f user manual 5 if the positive totalizer is turned off, the positive totalizer value shown on the screen is the value prior to its turn off m03 display flow rate and neg(negative) totalizer if the negative totalizer is turned off, the negative totalizer value shown on the screen is the valu...

  • Page 6

    Tuf-2000f user manual 6 (3) gasoline (4) fuel oil (5) crude oil (6) propane at -45c (7) butane at 0c (8)other liquids(need to enter sound speed in m21 and viscosity in m22) (9) diesel oil (10)caster oil (11)peanut oil (12) #90 gasoline (13) #93 gasoline (14) alcohol (15) hot water at 125c m21 window...

  • Page 7

    Tuf-2000f user manual 7 (2) function to store the current parameters into the flash memory, so that these parameters will be solidified and will be loaded as the default parameters every time when power is turned on. M27 entry to store to or restore from the internal flash memory, as many as 9 diffe...

  • Page 8

    Tuf-2000f user manual 8 display is used as the display device. M3. Setup for local segmental lcd display. Enter 0 or 1 for the non-auto-scan mode; enter 2~39 for the auto-scan mode. In the auto-scan mode the display will automatically scan displaying from 00 to the entered number of the local segmen...

  • Page 9

    Tuf-2000f user manual 9 (0) 4-20ma output mode (setup the output range from 4-20ma) (1) 0-20ma output mode (setup the output range from 4-20ma, this mode can only be used with version-15 flow meter) (2) serial port controls 0-20ma (3) 4-20ma corresponding fluid sound speed (4) 20-4-20ma mode (5) 0-4...

  • Page 10

    Tuf-2000f user manual 10 for version-15 flow meter, you need to indicate on your orders that you need the frequency function; otherwise you will get a flow meter which has no frequency output circuits. M68 window to setup the minimum flow rate value which corresponds to the lower frequency limit of ...

  • Page 11

    Tuf-2000f user manual 11 20. Timed alarm #2 21.Batch totalizer full 22. Periodically m51 timer 23. Oct not using the oct circuit does not source voltage at its output. It must be connected with an external power and pull-up resistant for some occasions. When the oct circuit is close, it will draw cu...

  • Page 12

    Tuf-2000f user manual 12 which stand for the total flow is over 100 cubic meters during the period of every day from 20:00 to 06:00, setups is like m51 start time =20:00:00 m51 interval =10:00:00 m51 log times =9999 (means always) m80 select item #9 m81 input 100 (unit is defined in m30,m31,m32) m81...

  • Page 13

    Tuf-2000f user manual 13 m91 displays the time ratio between the measured total transit time and the calculated time. If the pipe parameters are entered correctly and the transducers are properly installed, the ratio value should be in the range of 100±3%. Otherwise the entered parameters and the tr...

  • Page 14

    Tuf-2000f user manual 14 this function can used to produce an alarm or output when fluid material changes. M+7 displays total flow for this month(only for the time past) m+8 displays total flow for this year(only for the time past) m+9 display the not-working total time in seconds. The total failure...

  • Page 15

    Tuf-2000f user manual 15 (5) liquid type (for non-standard liquids, the sound speed of the liquid is also needed) (6) transducer type adapted to the flow meter. Generally the standard m1 clamp-on transducers will be the selected option. (7) transducer mounting methods (the v-method or z-method is th...

  • Page 16

    Tuf-2000f user manual an optimum location would be defined as a straight pipe length full of liquid that is to be measured. The piping can be in vertical or horizontal position. The following table shows 16 examples of optimum locations. L up l dn l up l dn l up l up l dn l dn p i p i n g c o n f i ...

  • Page 17

    Tuf-2000f user manual 4.3 transducers installation the transducers used by the tuf-2000 series ultrasonic flow meter are made of piezoelectric crystals both for transmitting and receiving ultrasonic signals through the wall of liquid piping system. The measurement is realized by measuring the travel...

  • Page 18

    Tuf-2000f user manual 4.3.4 z-method installation z-method is commonly used when the pipe diameter is above 200mm. 4.3.5 w-method installation w-method is usually used on plastic pipes with a diameter from 15mm to 50mm. 18.

  • Page 19

    Tuf-2000f user manual 19 4.3.6 n-method installation rarely used method. 4.4 installation checkup through the checkup of the installation, one can: check the receiving signal strength, the signal quality q value, the traveling time difference of the signals, the estimated liquid speed, the measured ...

  • Page 20: 5. Communication Protocols

    Tuf-2000f user manual 20 this ratio would be used to check the transducer installation. If the pipe parameters are entered correctly and the transducers are installed properly, the value for this ratio should be in the range of 100±3. If this range is exceeded, the user should check: (1) if the pipe...

  • Page 21

    Tuf-2000f user manual 21 for example, reading the registers from reg0001 to reg0010 in the unit #1 (ultrasonic flow meter) under the modbus-rtu format, the command could be as following 01 03 00 00 00 0a c5 cd ( hex ) unit function start reg numbers of regs check-sum while under the modbus-ascii for...

  • Page 22

    Tuf-2000f user manual 22 0019-0020 2 positive energy decimal fraction real4 0021-0022 2 negative energy accumulator long 0023-0024 2 negative energy decimal fraction real4 0025-0026 2 net accumulator long 0027-0028 2 net decimal fraction real4 0029-0030 2 net energy accumulator long 0031-0032 2 net ...

  • Page 23

    Tuf-2000f user manual 23 0062 1 times for the beeper integer writable 。 max 255 0062 1 pulses left for oct integer writable 。 max 65535 0072 1 error code bit 16bits, see note 4 0077-0078 2 pt100 resistance of inlet real4 in unit ohm 0079-0080 2 pt100 resistance of outlet real4 in unit ohm 0081-0082 ...

  • Page 24

    Tuf-2000f user manual 24 0115-0116 2 positive accumulator real4 in cubic meter , float 0117-0118 2 negative accumulator real4 in cubic meter , float 0119-0120 2 net energy accumulator real4 in gj , float 0121-0122 2 positive energy accumulator real4 in gj , float 0123-0124 2 negative energy accumula...

  • Page 25

    Tuf-2000f user manual 25 0056 0221-0222 2 inner pipe diameter real4 in millimeter 0229-0230 2 upstream delay real4 in microsecond 0231-0232 2 downstream delay real4 in microsecond 0233-0234 2 calculated travel time real4 in microsecond 0257-0288 32 lcd buffer bcd 0289 1 lcd buffer pointer integer 03...

  • Page 26

    Tuf-2000f user manual 26 3 imperial gallon (igl) 4 american million gallon (mgl) 5 cubic feet (cf) 6 american oil barrel (1 barrel =42gallon) (ob) 7 imperial oil barrel (ib) while the energy flow rate =(n+nf )×10n-4 ( unit decided by reg 1441 ) n=(0~10) is the energy multiplier which is in reg1440 (...

  • Page 27

    Tuf-2000f user manual 27 bit13 reserved bit14 internal timer over flow bit15 analog input over range please try to override these energy-related bits first when in flow-only measurement, (5) unit code for flow rate 0 cubic meter/second 1 cubic meter /minute 2 cubic meter /hour 3 cubic meter /day 4 l...

  • Page 28

    Tuf-2000f user manual 28 n/a 0162 1 data pointer integer range:0~63 2817 1 day and error code bcd day in high byte 2818 1 month and year bcd year in high byte 2819-2820 2 total working time long 2821-2822 2 net total flow for the day real4 0 2823-2824 2 net total energy for the day real4 2825 1 day ...

  • Page 29

    Tuf-2000f user manual 29 3339-3340 2 total working time long 3341-3342 2 net total flow for the month real4 1 3343-3344 2 net total energy for the month real4 。。。。 。。。。。。。。。。 。。。。。。 。。。。。。。。。。。。。。 。。。。。。 。。。。。。。。。。。。。。。。。。。。 31 3577-3584 8 data block no. 31 (3) there is no direct data for the year, ...

  • Page 30

    Tuf-2000f user manual 30 3597-3598 2 time duration when off long in seconds 3599-3600 2 corrected lost flow when off real4 in cubic meters 3601 1 power-on second and minute bcd second in low byte, minute in high 3602 1 power-on hour and day bcd hour in low byte, day in high 3603 1 power-on month and...

  • Page 31

    Tuf-2000f user manual 31 din(cr) return net totalizer ±ddddddde±d(cr) die(cr) return net thermal energy totalizer ±ddddddde±d(cr) die+(cr) return positive energy totalizer ±ddddddde±d(cr) die-(cr) return negtive energy totalizer ±ddddddde±d(cr) dit(cr) return net total flow for today ±ddddddde±d(cr)...

  • Page 32

    Tuf-2000f user manual 32 aoa (cr) output current ‘a’ ma at the ao output terminal. Aoa (cr)(lf)note 6 ba1(cr) return the resistance for t1 ±d.Dddddde±dd(cr)(lf) ba2(cr) return the resistance for t2 ±d.Dddddde±dd(cr)(lf) ba3(cr) return current value of ai3 (0~20ma) ±d.Dddddde±dd(cr)(lf) ba4(cr) retur...

  • Page 33

    Tuf-2000f user manual 33 2 . d stand for digit 0~9, the number before ‘e’ is an integer. 3 . working status code, 1~6 letters, refer to error code related chapter. 4 . ‘s’ is ‘on’,’off’ or ‘ud’ for example ‘tr:on,rl:on’ means the oct and relay are closed ‘tr:ud,rl:ud” means the oct and relay are not...

  • Page 34

    Tuf-2000f user manual 34 for example if the address number 88 flow meter is going to be addressed, the command should like: nxdv(cr), the decimal value of x should be 88. The prefix w is strongly recommended for new users. (3) the ‘w’ prefix usage: w + character string address + basic command the va...

  • Page 35

    Tuf-2000f user manual flow meters made by huizhong have more than 10 different communications protocols. For the easier replacement of a water meter, most of these protocols are realized in tuf-2000f flow meters. Here only one of them, the default for compatible protocols crl-61d (d these protocols ...

  • Page 36

    Tuf-2000f user manual 36 (2) reading the recorded meter data (command 49) format : 2ah xxh 49h answer: 26h xxh 49h ll ( bcd 码) zzh the difference between the command 4a and command 49 is that the late command reads out the data which are recorded in the meter by the time which is defined by command ...

  • Page 37

    Tuf-2000f user manual 37 05h stand for there exist hardware failure, repair may needed. 5.4 key value table the key values are used in a network application. By use of the key value and a command ‘m’, we can operate the flow meter through the network on a computer or other kind of terminals. For exa...