OfiTE 130-85 Instruction Manual

Summary of 130-85

  • Page 1

    Analog resistivity meter part no. 130-85 instruction manual updated 8/5/2011 ver. 1.4 ofi testing equipment, inc. 11302 steeplecrest dr. · houston, texas · 77065 · u.S.A. Tele: 832.320.7300 · fax: 713.880.9886 · www.Ofite.Com © copyright ofite 2011.

  • Page 2: Table of

    Table of contents intro....................................................................................................2 components......................................................................................2 procedure.......................................................................

  • Page 3: Components

    Components the ofite resistivity meter is a portable measuring device designed to give a quick, reliable measurement of the resistivity of a small sample expressed in ohm-meters. The transistorized meter accurately measures the resistivity of fluids, slurries and semisolids having resistivities from...

  • Page 4: Procedure

    1. Use the suction bulb to pull the sample into the lucite cell. Empty and refill the cell several times to thoroughly wet the cylinder walls. 2. Connect the cell assembly to the two terminal posts on the meter. Be sure the sample fills the area between the two metal posts in the cell. If the sample...

  • Page 5: Maintenance

    Maintenance 1. Always clean the cell immediately after making a measurement. Purge the cell with distilled water until it is clear. If additional cleaning is nec- essary, use a pipe cleaner and mild soap. Use no other solvents. Avoid scratching the surface of the cell. 2. When it is no longer possib...

  • Page 6: Nomograph

    Nomograph ofite, 11302 steeplecrest dr., houston, tx 77065 usa / tel: 832-320-7300 / fax: 713-880-9886 / www.Ofite.Com 5.