OfiTE 131-50 Instruction Manual

Summary of 131-50

  • Page 1

    Electrical stability meter part no. 131-50 instruction manual updated 12/10/2015 ver. 3.0 ofi testing equipment, inc. 11302 steeplecrest dr. · houston, texas · 77065 · u.S.A. Tele: 832.320.7300 · fax: 713.880.9886 · www.Ofite.Com © copyright ofite 2012.

  • Page 2: Table of

    Ofite, 11302 steeplecrest dr., houston, tx 77065 usa / tel: 832-320-7300 / fax: 713-880-9886 / www.Ofite.Com 1 intro..................................................................................................2 description ...........................................................................

  • Page 3: Intro

    Ofite, 11302 steeplecrest dr., houston, tx 77065 usa / tel: 832-320-7300 / fax: 713-880-9886 / www.Ofite.Com 2 the ofite electrical stability meter (es) is a sine wave instrument. It is manufactured in accordance with the american petroleum institute (api) “recommended practice standard procedure fo...

  • Page 4: Components

    Ofite, 11302 steeplecrest dr., houston, tx 77065 usa / tel: 832-320-7300 / fax: 713-880-9886 / www.Ofite.Com 3 components #131-01 probe with cable #131-51 high/low calibration standard #147-02 9-volt alkaline battery optional: #110-10 marsh funnel viscometer or 12 mesh screen #130-38 thermocup, 115 ...

  • Page 5: Specifications

    Ofite, 11302 steeplecrest dr., houston, tx 77065 usa / tel: 832-320-7300 / fax: 713-880-9886 / www.Ofite.Com 4 specifications meter: wave form: sine, ac frequency: 340 ± 10 hz output units: peak volts ramp rate: 150 ± 10 volts per second, automatic operation minimum output range: 3 - 2,000 volts (pe...

  • Page 6: Procedure

    Ofite, 11302 steeplecrest dr., houston, tx 77065 usa / tel: 832-320-7300 / fax: 713-880-9886 / www.Ofite.Com 5 procedure 1. Inspect the electrode probe and cable for any evidence of damage. En- sure that the entire electrode gap is free of deposits and that the connec- tor to the instrument is clean...

  • Page 7: Calibration -

    Ofite, 11302 steeplecrest dr., houston, tx 77065 usa / tel: 832-320-7300 / fax: 713-880-9886 / www.Ofite.Com 6 calibration - performance test 1. Inspect the electrode probe and cable for evidence of damage. Make sure the entire electrode gap is free of deposits and the connector to the instrument is...

  • Page 8: Warranty and

    Ofite, 11302 steeplecrest dr., houston, tx 77065 usa / tel: 832-320-7300 / fax: 713-880-9886 / www.Ofite.Com 7 warranty and return policy warranty: ofi testing equipment, inc. (ofite) warrants that the products shall be free from liens and defects in title, and shall conform in all respects to the t...