OHAUS Defender 2200 Instruction Manual

Manual is about: Bench Scale

Summary of Defender 2200

  • Page 1

    Defender ™ 2200 bench scale instruction manual.

  • Page 3

    Defender 2200 bench scale en-1 table of contents 1. Introduction ............................................................................................................... 3 1.1 safety precautions......................................................................................................

  • Page 4

    Defender 2200 bench scale en-2 table of contents continued 3.8 print menu.............................................................................................................................................22 3.8.1 reset ..........................................................................

  • Page 5

    Defender 2200 bench scale en-3 1. Introduction this manual contains installation, operation and maintenance instructions for the ohaus defender tm 2200 bench scales. Please read this manual completely before installation and operation. 1.1 safety precautions for safe and dependable operation of this...

  • Page 6

    Defender 2200 bench scale en-4 1.2 overview of parts and controls 1 table 1-1. Indicator parts 1 data label 2 front housing 3 adjusting knob (2) 4 control panel 5 mounting bracket 6 screw (4) 7 rear housing 8 data label 9 power receptacle 10 strain relief for load cell cable 11 rs 232 connector figu...

  • Page 7

    Defender 2200 bench scale en-5 1.2 overview of parts and controls (cont.) figure 1-2. Main pc board table 1-2. Main pc board item description 1 line power input j1 2 lft on / off switch 3 rs232 connector j3 4 sense jumper w2 5 load cell terminal block j7 6 sense jumper w1.

  • Page 8

    Defender 2200 bench scale en-6 1.2 overview of parts and controls (cont.) lb oz kg figure 1-3. Controls and indicators t22mc table 1-3. Control panel.. No. Designation 1 capacity label window 2 pound indicator 3 ounce 4 kilogram, gram indicator 5 tare menu button 6 calibration mode indicator 7 funct...

  • Page 9

    Defender 2200 bench scale en-7 1.3 control functions table 1-2. Functions of display controls button primary function (short press) on/zero turns the indicator on. Print sends the current value to the com port if autoprint is set to off. Function initiates an application mode. Tare performs a tare o...

  • Page 10

    Defender 2200 bench scale en-8 2. Installation 2.1 unpacking unpack and inspect the product to make sure that all components have been included. The package includes the following: • 2200 series indicator • column assembly • ac adapter • mounting bracket • knobs (2) • base and weighing pan • warrant...

  • Page 11

    Defender 2200 bench scale en-9 2.3.2 scale base to the indicator pass the load cell cable through the strain relief and attach it to terminal block j7. Pin connection wire color j7-1 +excitation green j7-2 +sense black j7-3 +signal red j7-4 ground white j7-5 –signal blue j7-6 –sense brown j7-7 –exci...

  • Page 12

    Defender 2200 bench scale en-10 2.5 selecting the location to ensure accuracy, proper performance and safety, locate and operate the base on a stable, level surface. Avoid locations with rapid temperature changes or excessive dust, air currents, vibrations, electromagnetic fields or heat. Level the ...

  • Page 13

    Defender 2200 bench scale en-11 3.2 menu navigation to enter the menu mode press and hold the menu button until menu appears on the display. The first upper level menu appears on the display. Summary of button navigation functions in menu mode: yes allows entry into the displayed menu. Accepts the d...

  • Page 14

    Defender 2200 bench scale en-12 3.3.1 span calibration span calibration uses two points to adjust the scale. The first point is the zero value where there is no weight on the scale. The second point is the span value where a calibration mass is placed on the scale. When span is displayed, press the ...

  • Page 15

    Defender 2200 bench scale en-13 3.3.3 geographical adjustment factor the geographcial adjustment factor (geo) is used to compensate for variations in gravity. Settings from 0 to 31 are available. Refer to table 3-2 to determine the geo factor that corresponds to your location. Note: changing the geo...

  • Page 16

    Defender 2200 bench scale en-14 table 3-2. Geographical adjustment values elevation above sea level in meters 00 0 325 325 650 650 975 0 975 1300 1300 1625 1625 1950 1950 2275 2275 2600 2600 2925 2925 3250 3250 3575 elevation above sea level in feet geographical latitude away from the equator (north...

  • Page 17

    Defender 2200 bench scale en-15 3.4 setup menu when the indicator is used for the first time, enter this menu to set the capacity and graduation. Reset legal for trade calibration unit capacity graduation power on unit zero range retain zero data end setup no, yes off, on kg¸ lb 5…20000 0.0005…20 au...

  • Page 18

    Defender 2200 bench scale en-16 table 3-3. Setup & calibration values capacity graduation size with lft off graduation size with lft on span calibration points linearity calibration points 5 0.0005, 0.001, 0.002, 0.005 0.001, 0.002, 0.005 5 2, 5 10 0.0005, 0.001, 0.002, 0.005, 0.01 0.002, 0.005, 0.0...

  • Page 19

    Defender 2200 bench scale en-17 3.4.5 graduation set the scale readability. 0.0005, 0.001, 0.002, 0.005, 0.01, 0.02, 0.05, 0.1, 0.2, 0.5, 1, 2, 5, 10, 20. Note: not all settings are available for each capacity. Refer to the setup table 3.3 for available settings. 3.4.6 power on unit set the unit of ...

  • Page 20

    Defender 2200 bench scale en-18 3.5 readout menu enter this menu to customize display functionality. Reset stable filter level auto-zero tracking sleep (t22me) light (t22me) backlight (t22mc) auto off timer expand end readout no, yes 0.5d, 1d, 2d, 3d lo, med, hi off, 0.5d, 1d, 3d off, on lo, med, hi...

  • Page 21

    Defender 2200 bench scale en-19 3.5.5 sleep set the sleep functionality.(display powers off after 60 seconds of inactivity) off = disabled. On = enabled. 3.5.6 light set the brightness of led display. Lo = low intensity med = medium intensity hi = high intensity 3.5.7 auto off timer set the automati...

  • Page 22

    Defender 2200 bench scale en-20 3.6 mode menu enter this menu to activate the desired application modes. Reset count hold check totalize end mode no, yes off, on off, on off, on off, on exit mode menu 3.6.1 reset set the mode menu to the factory defaults. No = not reset. Yes = reset. 3.6.2 parts cou...

  • Page 23

    Defender 2200 bench scale en-21 3.7 unit menu enter this menu to activate the desired units. Default settings are bold. Reset kilogram pound gram ounce pound ounce end unit no, yes off, on off, on off, on off, on off, on exit unit menu 3.7.1 reset set the unit menu to the factory defaults. No = not ...

  • Page 24

    Defender 2200 bench scale en-22 3.8 print menu enter this menu to define printing parameters. Default settings are bold. Reset baud rate parity stop bit handshake stable only auto print content end print no, yes 300, 600, 1200, 2400, 4800, 9600, 19200 7 even, 7 odd, 7 none, 8 none 1, 2 off, xon/xoff...

  • Page 25

    Defender 2200 bench scale en-23 3.8.5 handshake set the flow control method. Off = no handshaking. On-off = xon/xoff software handshaking. 3.8.6 print stable data only set the print critera. Off = values are printed immediately. On = values are only printed when the stability criteria are met. 3.8.7...

  • Page 26

    Defender 2200 bench scale en-24 3.9 lock menu use this menu to prevent unauthorized changes to menu settings. Default settings are bold. Reset lock calibration menu lock setup menu lock readout menu lock mode menu lock unit menu lock print menu end lock menu no, yes off, on off, on off, on off, on o...

  • Page 27

    Defender 2200 bench scale en-25 3.9.7 lock print set the status. Off = print menu is not locked. On = print menu is locked and hidden 3.9.8 end lock advance to the next menu. 3.9 end menu exit to weighing. 3.10 security switch a security switch is located on the main pcb board. When the switch is se...

  • Page 28

    Defender 2200 bench scale en-26 4.4 changing units of measure press and hold the print units button until the desired measuring unit appears. Only measuring units enabled in the unit menu will be displayed (refer to section 3.7). 4.5 printing data printing the displayed data to a printer or sending ...

  • Page 29

    Defender 2200 bench scale en-27 establishing the average piece weight (apw) the display shows put 5 pcs. Establishing a new apw press the no button to increment the sample size. Choices are 5, 10, 20, 50, 100 and 200. To establish the apw, place the specified quantity of samples on the scale and pre...

  • Page 30

    Defender 2200 bench scale en-28 4.6.3 display-hold use this mode to capture and store the first stable display value. To enter the display hold mode, press and hold the mode button until hold is displayed. When the mode button is released, the hold indicator will be turned on. Place the item on the ...

  • Page 31

    Defender 2200 bench scale en-29 editing the over setting the display shows set.Hi. Press the yes button to edit the over setting. Settings: –99999 to 999999 refer to menu navigation section 3.2 to enter settings. Begin check weighing the appropriate under, accept or over led lights to indicate check...

  • Page 32

    Defender 2200 bench scale en-30 4.6.5 totalize weighing use this mode to store the total of a series of weight measurements. Note: only positive numbers will be totaled. Accumulation (acc) when the mode button is released, clr.Acc accum is displayed. Start totalize press the yes button to clear the ...

  • Page 33

    Defender 2200 bench scale en-31 5 serial communication the setup of rs232 operating parameters are fully explained in section 3.8. The physical hardware connection is explained in in section 2.2. The interface enables display data to be sent to a computer or printer. A computer can be used to contro...

  • Page 34

    Defender 2200 bench scale en-32 5.2 output format the default serial output format is shown below. Field: polarity space weight space unit stability legend cr lf length: 1 1 7 1 5 1 3 1 1 definitions: polarity, “–” sign if negative, blank if positive. Weight, up to 6 numbers and 1 decimal, right jus...

  • Page 35

    Defender 2200 bench scale en-33 6. Legal for trade when the indicator is used in trade or a legally controlled application it must be set up, verified and sealed in accordance with local weights and measures regulations. It is the responsibility of the purchaser to ensure that all pertinent legal re...

  • Page 36

    Defender 2200 bench scale en-34 6.3 sealing 6.3.1 physical seals for jurisdictions that use the physical sealing method, the local weights and measures official or authorized service agent must apply a security seal to prevent tampering with the settings. Refer to the illustrations below for sealing...

  • Page 37

    Defender 2200 bench scale en-35 6.3.2 audit trail seal for jurisdictions that use the audit trail sealing method, the local weights and measures official or authorized service agent must record the current configuration and calibration event counter values at the time of sealing. These values will b...

  • Page 38

    Defender 2200 bench scale en-36 7 maintenance caution: disconnect the unit from the power supply before cleaning. 7.1 indicator cleaning • the housing may be cleaned with a cloth dampened with a mild detergent if necessary. • do not use solvents, chemicals, alcohol, ammonia or abrasives to clean the...

  • Page 39

    Defender 2200 bench scale en-37 7.3 service information if the troubleshooting section does not resolve your problem, contact an authorized ohaus service agent. For service assistance in the united states, call toll-free 1-800-526-0659 between 8:00 am and 5:00 pm eastern standard time. An ohaus prod...

  • Page 40

    Defender 2200 bench scale en-38 8. Technical data 8.1 specifications materials housing rear: carbon-steel housing front: abs plastic keypad: polyester display window: polycarbonate ambient conditions the technical data is valid under the following ambient conditions: ambient temperature: –10°c to 40...

  • Page 41

    Defender 2200 bench scale en-39 8.2 accessories table 8-2. Accessories description part number interface cable/pc 25-pin 80500524 interface cable/pc 9-pin 80500525 column kit, 35 cm painted steel 80251743 column kit, 70 cm painted steel 80251744 column kit, 35 cm stainless steel 80251745 column kit,...

  • Page 42

    Defender 2200 bench scale en-40 8.3.2 defender 2200 base base shown with defender 2200 indicator and column figure 8-2. Defender 2200 base dimension drawing table 8.3. Dimensions a b c d e base pan depth pan width height of pan to surface of table* height of pan depth including level indicator d30br...

  • Page 43

    Defender 2200 bench scale en-41 8.4 compliance compliance to the following standards is indicated by the corresponding mark on the product. Marking standard this product conforms to the emc directive 2004/108/ec, the low voltage directive 2006/95/ec and the non-automatic weighing instruments directi...

  • Page 44

    Defender 2200 bench scale en-42 fcc note this equipment has been tested and found to comply with the limits for a class b digital device, pursuant to part 15 of the fcc rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a ...

  • Page 46

    Ohaus corporation 7 campus drive suite 310 parsippany, nj 07054 usa tel: (973) 377-9000 fax: (973) 944-7177 with offices worldwide www.Ohaus.Com *83033841* p/n 83033841 © 2011 ohaus corporation, all rights reserved printed in china.